Motion Simulation And Finite Element Analysis Of Knee Prosthesis With Implant

被引:0
|
作者
Fang, Zhibin [1 ]
Zhang, Shaobin [1 ]
Cheng, Jiamei [2 ]
Li, Shaoming [2 ]
机构
[1] Hebei Agr Univ, Dept Sports Work, Hebei 071000, Hebei, Peoples R China
[2] Hebei Agr Univ, Coll Sci & Technol, Huanghua 061100, Hebei, Peoples R China
来源
JOURNAL OF APPLIED SCIENCE AND ENGINEERING | 2025年 / 28卷 / 04期
关键词
knee prosthesis; knee joint; finite element analysis; ABAQUS software; stress distribution; KINEMATICS; REPLACEMENT; MENISCUS; CONTACT;
D O I
10.6180/jase.202504_28(4).0001
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In this study, we investigated the mechanical behavior of a knee replacement prosthesis (TKR) manufactured by the Zimmer company. To facilitate our analysis, we initially utilized a coordinate measuring device, specifically a contact 3D scanner, to prepare a cloud-of-point model of the prosthesis. This scanning process allowed us to accurately capture the geometry and dimensions of the TKR, providing a detailed representation of its physical structure. By utilizing this advanced scanning technology, we ensured that our subsequent simulations and analyses were based on precise and reliable data, enabling a thorough examination of the mechanical performance of the knee replacement prosthesis. ABAQUS software was then used to analyze the threedimensional model and nonlinear static analysis was performed on the model. This simulation examined the mechanical performance of the prosthesis for different weight ranges, and the distribution of stress, strain, and displacement within the prosthesis was analyzed. The results show that the maximum stress created in the investigated prosthesis increases from 16MPa to 64MPa per weight of 55 kg to 75 kg. Although, with a 26% increase in the weight of the individual using a knee prosthesis, the maximum stress created in the prosthesis increases by 76%. This type of prosthesis is suitable for the maximum weight category of 80 kg, as it has a reliability coefficient of 3. In light of these results, it is clear that weight categories must be taken into account when considering a particular prosthesis. Otherwise, the prosthesis may be destroyed due to the application of larger forces during various everyday situations and result in serious knee injuries.
引用
收藏
页码:667 / 679
页数:13
相关论文
共 50 条
  • [31] Finite element analysis of a transmandibular implant
    Nomura, Tsutomu
    Powers, Michael P.
    Katz, J. Lawrence
    Saito, Chikara
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 80B (02) : 370 - 376
  • [32] Finite Element Analysis of a TMJ Implant
    Kashi, A.
    Chowdhury, A. Roy
    Saha, S.
    JOURNAL OF DENTAL RESEARCH, 2010, 89 (03) : 241 - 245
  • [33] Spherical center and rotating platform hinged knee prosthesis: Finite-element model establishment, verification and contact analysis
    Zhang, Jing-Yu
    Wang, Jian
    Tian, Dong-Mu
    Jiang, Da-Ping
    Li, Jian-Jun
    Hu, Yong-Cheng
    KNEE, 2020, 27 (03) : 731 - 739
  • [34] The Effects of Different Femoral Component Materials on Bone and Implant Response in Total Knee Arthroplasty: A Finite Element Analysis
    Galas, Allegra
    Banci, Lorenzo
    Innocenti, Bernardo
    MATERIALS, 2023, 16 (16)
  • [35] Mechanical alignment tolerance of a cruciate-retaining knee prosthesis under gait loading-A finite element analysis
    Luan, Yichao
    Wang, Huizhi
    Fang, Chaohua
    Zhang, Min
    Li, Junwei
    Zhang, Ningze
    Liu, Bolun
    Su, Jian
    Cheng, Cheng-Kung
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [36] Finite element analysis of zygomatic implant positioning protocols with conventional implant configurations
    Balkan, Burcu
    Atalay, Belir
    BIOTECHNOLOGY & BIOTECHNOLOGICAL EQUIPMENT, 2024, 38 (01)
  • [37] Finite element analysis of mini implant biomechanics on peri-implant bone
    Hisam, Muhammad Juzaili
    Zaman, Izzuddin
    Shari, Safian
    Kurniawan, Denni
    Nor, Fethma M.
    DIGITAL MANUFACTURING TRANSFORMING INDUSTRY TOWARDS SUSTAINABLE GROWTH, 2019, 30 : 308 - 314
  • [38] Influence of post-cam design of posterior stabilized knee prosthesis on tibiofemoral motion during high knee flexion
    Lin, Kun-Jhih
    Huang, Chang-Hung
    Liu, Yu-Liang
    Chen, Wen-Chuan
    Chang, Tsung-Wei
    Yang, Chan-Tsung
    Lai, Yu-Shu
    Cheng, Cheng-Kung
    CLINICAL BIOMECHANICS, 2011, 26 (08) : 847 - 852
  • [39] Optimized Orthodontic Palatal Miniscrew Implant Insertion Angulation: A Finite Element Analysis
    Mir, Arash Poorsattar Bejeh
    Ravadgar, Mehdi
    Mir, Morvarid Poorsattar Bejeh
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (01) : E1 - E9
  • [40] FEM Analysis of One Element Prosthesis on Dental Implant
    Roatesi, Iulia
    Roatesi, Simona
    Rotaru, Constantin