Dose distribution analysis of rapid arc and intensity modulated radiotherapy plan in head and neck cancer

被引:0
|
作者
Ansari S. [1 ]
Satpathy S.K. [1 ]
机构
[1] Post MSc Diploma in Radiation Physics, Department of Radiation Oncology, Apollo Hospitals Bilaspur, (C.G)
关键词
Beam; Cancer; Dosimetry;
D O I
10.22038/ijmp.2018.31896.1378
中图分类号
学科分类号
摘要
Introduction: Globally, intensity-modulated radiation therapy (IMRT) is considered as highly precise and accurate method of radiotherapy planning. This technique amplifies spatial dose distribution conformity by modulating the intensity of radiation beams in each sub-volume. Additionally, it can reduces the dose to surrounding critical organs and deliver the planned dose to targets with the nominal risk of side effects. Material and Methods: In this study, 13 patients with head and neck cancer were randomly taken for analysis. The IMRT and Rapid Arc plans were generated for each case in the Eclipse treatment planning system, version 11.0. There were seven to nine beams deployed in IMRT plan, while Rapid Arc plans were performed using two arcs with opposite direction of rotation. Portal dosimetry plans were created and analyzed before executing the plan on the patient. Results: The mean of V95%(Target's volume covered by 95% isodose line)was 97.89% and 97.47% for Rapid Arc and IMRT plans, respectively. Moreover, mean standard deviations were found 1.93 and 1.70 in Rapid Arc and IMRT plans, respectively. The mean gamma index was97.55% and 98.43% in Rapid Arc and IMRT, respectively. Conclusion: IMRT technique was slightly better in the treatment of head and neck cancer compared to the Rapid Arc method. The only advantage of Rapid Arc was saving the treatment time by two to three times on an average compared with IMRT. It is prudent to use IMRT technique in head and neck cancer treatment. © 2019 Mashhad University of Medical Sciences.
引用
收藏
页码:139 / 144
页数:5
相关论文
共 50 条
  • [1] Suggested Template Plan for Forward Intensity Modulated Radiotherapy in Head and Neck Cancer Patients
    Ahmed, Moustafa Tawfik
    Baker, Naer Ahmed
    Elshahat, Khaled Mohammed
    Akl, Mohamed Farouk
    Shawata, Ahmed Shawky
    MIDDLE EAST JOURNAL OF CANCER, 2021, 12 (01) : 128 - 136
  • [2] Comparison of intensity-modulated radiotherapy and volumetric modulated arc therapy dose measurement for head and neck cancer using optical stimulated luminescence dosimeter
    Lai, Lu-Han
    Chuang, Keh-Shih
    Lin, Hsin-Hon
    Liu, Yi-Chi
    Kuo, Chiung-Wen
    Lin, Jao-Perng
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 266 - 270
  • [3] Intensity-modulated radiotherapy for head and neck surgeons
    Gutiontov, Stanley I.
    Shin, Edward J.
    Lok, Benjamin
    Lee, Nancy Y.
    Cabanillas, Ruben
    HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2016, 38 : E2368 - E2373
  • [4] Intensity-modulated radiotherapy or volumetric-modulated arc therapy in patients with head and neck cancer: Focus on salivary glands dosimetry
    Vallard, Alexis
    Guy, Jean-Baptiste
    Ndong, Sylvie Mengue
    Vial, Nicolas
    Rivoirard, Romain
    Auberdiac, Pierre
    Mery, Benoite
    Langrand-Escure, Julien
    Espenel, Sophie
    Moncharmont, Coralie
    Ben Mrad, Majed
    Diao, Peng
    Goyet, Dominique
    Magne, Nicolas
    HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2016, 38 (07): : 1028 - 1034
  • [5] Target volume delineation for head and neck cancer intensity-modulated radiotherapy
    Lapeyre, M.
    Toledano, I.
    Bourry, N.
    Bailly, C.
    Cachin, F.
    CANCER RADIOTHERAPIE, 2011, 15 (6-7): : 466 - 472
  • [6] Physical and biological pretreatment quality assurance of the head and neck cancer plan with the volumetric modulated arc therapy
    Park, So-Hyun
    Lee, Dong-Soo
    Lee, Yun-Hee
    Lee, Seu-Ran
    Kim, Min-Ju
    Suh, Tae-Suk
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (05) : 946 - 955
  • [7] Skin dose for head and neck cancer patients treated with intensity-modulated radiation therapy(IMRT)
    Fu, Hsiao-Ju
    Li, Chi-Wei
    Tsai, Wei-Ta
    Chang, Chih-Chia
    Tsang, Yuk-Wah
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 435 - 441
  • [8] Dosimetry of conformal dynamic arc radiotherapy and intensity modulated radiotherapy in unresectable cholangiocarcinoma
    Peerawong, Thanarpan
    Khorprasert, Chonlakiet
    Suriyapee, Sivalee
    Sanghangthum, Taweap
    Ayuthaya, Isra Israngkul Na
    Shotelersuk, Kanjana
    ASIAN BIOMEDICINE, 2010, 4 (01) : 131 - 139
  • [9] Radiobiological Evaluation of Intensity Modulated Radiotherapy Treatment for Locally Advanced Head and Neck Squamous Cell Carcinomas
    Reddy, B. Rekha
    Ravikumar, M.
    Pasha, C. R. Tanvir
    Kumar, M. R. Anil
    Pyakuryal, A.
    Varatharaj, C.
    Narayanasamy, Ganesh
    MEDICAL PHYSICS, 2014, 41 (06) : 287 - +
  • [10] A New Phantom for Individual Verification of the Dose Distribution in Precision Radiotherapy for Head-and-Neck Cancer
    Grehn, Melanie
    Stille, Maik
    Ziemann, Christian
    Cremers, Florian
    Rades, Dirk
    Buzug, Thorsten M.
    ANTICANCER RESEARCH, 2019, 39 (12) : 6931 - 6938