Effect of platform design of dental implant abutment on loosening and fatigue performance

被引:0
|
作者
Sun, Fei [1 ]
Xu, Li-Bing [1 ]
Lai, Song-Xian [1 ]
Xu, Hai [2 ]
Li, Xin-Chang [1 ,2 ]
Lin, Zeng [1 ]
机构
[1] Northeastern Univ, Sch Mech Engn & Automat, Shenyang 110004, Peoples R China
[2] WEGO JERICOM Biomat Co Ltd, Weihai 264210, Peoples R China
基金
中国博士后科学基金;
关键词
Abutment; Structural design; Loosening; Fatigue; FEA; IN-VITRO; TORQUE; CONNECTIONS; SCREW; STRENGTH; SURVIVAL; HEXAGON; CYCLES;
D O I
10.1016/j.engfailanal.2024.109134
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Under the influence of daily occlusal forces, mechanical complications such as loose connections and fatigue damage are significant factors contributing to dental implant failure. The structural design of implant components is crucial in enhancing the long-term durability of implant systems. This research explores the impact of the platform structure of the abutment on connection loosening and fatigue properties. Four abutments with varying platform structures were designed and produced. The study involved testing and comparing screw loosening behavior before and after loading, as well as evaluating static load strength and fatigue characteristics. Observations were made on abutment surface wear and fatigue sections. A three-dimensional model was utilized to confirm damage location using the finite element method. Results indicate that the abutment's platform structure enhances anti-loosening performance and static failure load, while reducing fatigue life. Additionally, the position of fatigue fracture in the abutment is influenced by the load magnitude. The finite element analysis (FEA) findings align with the results of the static load tests.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Influence of Framework Material and Abutment Configuration on Fatigue Performance in Dental Implant Systems: A Finite Element Analysis
    Erdogdu, Meryem
    Demirel, Mehmet Goekberkkaan
    Mohammadi, Reza
    Guentekin, Neslihan
    Chaleshtori, Masoud Ghanbarzadeh
    MEDICINA-LITHUANIA, 2024, 60 (09):
  • [2] Implant–abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants
    F. J. Gil
    M. Herrero-Climent
    P. Lázaro
    J. V. Rios
    Journal of Materials Science: Materials in Medicine, 2014, 25 : 1825 - 1830
  • [3] Implant Abutment Screw Loosening: A Review of Effective Factors
    Seddigh, Mohammad Ali
    Mostafavi, Azam Sadat
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2019, 13 (08) : ZE6 - ZE9
  • [4] Effect of central screw taper angles on the loosening performance and fatigue characteristics of dental implants
    Sun, Fei
    Lv, Li-Tao
    Xiang, Ding-Ding
    Ba, De-Chun
    Lin, Zeng
    Song, Gui-Qiu
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 129
  • [5] Load Fatigue Performance Evaluation on Two Internal Tapered Abutment-Implant Connection Implants Under Different Screw Tightening Torques
    Jeng, Ming-Dih
    Liu, Po-Yi
    Kuo, Jia-Hum
    Lin, Chun-Li
    JOURNAL OF ORAL IMPLANTOLOGY, 2017, 43 (02) : 107 - 114
  • [6] Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants
    Gil, F. J.
    Herrero-Climent, M.
    Lazaro, P.
    Rios, J. V.
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2014, 25 (07) : 1825 - 1830
  • [7] Abutment screw loosening of endosseous dental implant body/abutment joint by cyclic torsional loading test at the initial stage
    Katsuta, Yasuhiro
    Watanabe, Fumihiko
    DENTAL MATERIALS JOURNAL, 2015, 34 (06) : 896 - 902
  • [8] In Vitro Analysis of Post-fatigue Reverse-Torque Values at the Dental Abutment/Implant Interface for a Unitarian Abutment Design
    Cashman, Paul M.
    Schneider, Robert L.
    Schneider, Galen B.
    Stanford, Clark M.
    Clancy, James M.
    Qian, Fang
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2011, 20 (07): : 503 - 509
  • [9] Effect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis
    Kim, Seok-Gyu
    Chung, Chae-Heon
    Son, Mee-Kyoung
    JOURNAL OF ADVANCED PROSTHODONTICS, 2015, 7 (03) : 207 - 213
  • [10] Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading
    El-Sheikh, Mai Ahmed Yousry
    Mostafa, Tamer Mohamed Nasr
    El-Sheikh, Mohamed Maamoun
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2018, 4