The effect of mixing technique on fatigue of bone cement when stress concentrations are present

被引:1
作者
Hoey D. [1 ]
Taylor D. [1 ]
机构
[1] Trinity Centre for Bioengineering, Trinity College
关键词
Bone cement; Critical distance; Fatigue; Mixing technique; Stress concentration;
D O I
10.1504/IJNBM.2010.036106
中图分类号
学科分类号
摘要
Mixing techniques used to reduce porosity in bone cement demonstrate large improvements in fatigue behaviour in test specimens, but do not appear to reduce the rate of loosening of implants in vivo. We examined data showing the effect of notches on fatigue strength and found, surprisingly, that notches with low stress-concentration factors (Kt < 2.4) had no effect compared to unnotched specimens. Using the theory of critical distances (TCD) we found the characteristic fatigue strength of the porous material to be similar to the measured fatigue strength of pore-free material; the characteristic length (L) was similar to the average pore size. This allowed us to define a number of different cases, depending on the relative dimensions of the notch and the pores. We concluded that techniques which reduce porosity can be expected to improve fatigue life only in certain cases, which are frequently encountered in test specimens but only rarely occur in clinical situations. Copyright © 2010 Inderscience Enterprises Ltd.
引用
收藏
页码:36 / 48
页数:12
相关论文
共 13 条
  • [1] Breusch S., Malchau H., The Well-cemented Total Hip Arthroplasty, (2005)
  • [2] Culleton T.P., Prendergast P.J., Taylor D., Fatigue failure in the cement mantle of an artificial hip joint, Clinical Materials, 12, pp. 95-102, (1993)
  • [3] Dunne N.J., Orr J.F., Mushipe M.T., Eveleigh R.J., The relationship between porosity and fatigue characteristics of bone cements, Biomaterials, 24, pp. 239-245, (2003)
  • [4] Geiger M., Ritter M.A., Ginther J.A., Faris P.M., Meding J.B., The clinical significance of vacuum mixing bone cement, Clinical Orthopaedics and Related Research, 382, pp. 258-266, (2001)
  • [5] Hoey D., Taylor D., Fatigue in porous PMMA: The effect of stress concentrations, International Journal of Fatigue, 30, pp. 989-995, (2008)
  • [6] James S.P., Jasty M., Davies J., Piehler H., Harris W.H., A fractographic investigation of PMMA bone cement focusing on the relationship between porosity reduction and increased fatigue life, Journal of Biomedical Materials Research, 26, pp. 651-662, (1992)
  • [7] Ling R.S., Lee A.J.C., Porosity reduction in acrylic cement is clinically irrelevant, Clinical Orthopaedics and Related Research, 355, pp. 249-253, (1998)
  • [8] Malchau H., Herberts P., Ahnfelt L.E., Johnell O., Prognosis of Total Hip Replacement' Results from A National Register of Revised Failures 1979-1990 in Sweden - A 10 Year Follow Up of 92675 Total Hip Replacements, (1994)
  • [9] Sinnett-Jones P.E., Browne M., Moffat A.J., Jeffers J.R.T., Saffari N., Buffiere J.Y., Sinclair I., Crack initiation processes in acrylic bone cement, Journal of Biomedical Materials Research - Part A, 89, pp. 1088-1097, (2009)
  • [10] Susmel L., Taylor D., Fatigue design in the presence of stress concentrations, International Journal of Strain Analysis, 38, pp. 443-452, (2003)