Stochastic model for predicting the temporal structure of the plan delivery in a synchrotron-based pencil beam scanning proton therapy system

被引:2
|
作者
Burguete, J. [1 ]
Garcia-Cardosa, M. [1 ]
Antolin, E. [2 ]
Aguilar, B. [2 ]
Azcona, J. D. [2 ]
机构
[1] Univ Navarra, Dept Phys & Appl Math, Irunlarrea 1, Pamplona 31008, Spain
[2] Clin Univ Navarra, Serv Radiat Phys & Radiat Protect, Marquesado Santa Marta 1, Madrid 28027, Spain
关键词
Proton therapy; Synchrotron; Stochastic model; Pencil beam scanning; Temporal structure of the beam; SPILL RIPPLE; TIME; MOTION; LUNG;
D O I
10.1016/j.radphyschem.2024.112276
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Accurately predicting dose delivery is crucial for achieving fully personalized treatments in external beam radiation therapy. However, this task remains challenging in some current technologies. In the case of Proton Therapy, for example, current systems employ complex strategies where a pencil beam is scanned in the tumor for treatment delivery. Some parameters in these treatments fluctuate and cannot be fully controlled. Therefore, a stochastic model that accounts for temporal uncertainties can be the best approach to describe these behaviors, particularly when the time-dependent beam interacts with other processes such as moving tumors or organs at risk. This paper aims to provide medical physicists with a tool for accurately predicting the temporal structure of beam delivery. To achieve this, we followed a two-step process. First, we characterized the probability distributions for all relevant times in dose delivery. Second, we developed a model based on the measured data. This model serves as a starting point to improve treatment planning performance by providing a range of expected times for dose delivery. While the process was carried out using a compact synchrotron at our university, it can be easily adapted to other technologies.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Commissioning dose computation model for proton source in pencil beam scanning therapy by convolution neural networks
    Liu, Yaoying
    Shang, Xuying
    Zhao, Wei
    Li, Nan
    Qu, Baolin
    Zou, Yue
    Le, Xiaoyun
    Zhang, Gaolong
    Xu, Shouping
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (15)
  • [22] Development and long-term stability of a comprehensive daily QA program for a modern pencil beam scanning (PBS) proton therapy delivery system
    Rana, Suresh
    Bennouna, Jaafar
    Samuel, E. James Jebaseelan
    Gutierrez, Onso N.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (04): : 29 - 44
  • [23] Integration and dosimetric validation of a dynamic collimation system for pencil beam scanning proton therapy
    Nelson, Nicholas P.
    Culberson, Wesley S.
    Hyer, Daniel E.
    Geoghegan, Theodore J.
    Patwardhan, Kaustubh A.
    Smith, Blake R.
    Flynn, Ryan T.
    Gutierrez, Alonso N.
    Boland, Thibault
    Hill, Patrick M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (06)
  • [24] Pencil beam scanning proton therapy for mediastinal lymphomas in deep inspiration breath-hold: a retrospective assessment of plan robustness
    Hoerberger, Filip
    Andersson, Karin M.
    Enmark, Marika
    Kristensen, Ingrid
    Flejmer, Anna
    Edvardsson, Anneli
    ACTA ONCOLOGICA, 2024, 63 : 62 - 69
  • [25] The use of a mini-ridge filter with cyclotron-based pencil beam scanning proton therapy
    O'Grady, Fionnbarr
    Janson, Martin
    Rao, Avani D.
    Chawla, Ashish K.
    Stephenson, Lisa
    Fan, Jiajin
    Wang, Peng
    MEDICAL PHYSICS, 2023, 50 (04) : 1999 - 2008
  • [26] Clinical Implementation of Proton Therapy Using Pencil-Beam Scanning Delivery Combined With Static Apertures
    Baeumer, Christian
    Plaude, Sandija
    Khalil, Dalia Ahmad
    Geismar, Dirk
    Kramer, Paul-Heinz
    Kroninger, Kevin
    Nitsch, Christian
    Wulff, Joerg
    Timmermann, Beate
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [27] Implication of spot position error on plan quality and patient safety in pencil-beam-scanning proton therapy
    Yu, Juan
    Beltran, Chris J.
    Herman, Michael G.
    MEDICAL PHYSICS, 2014, 41 (08) : 96 - 103
  • [28] Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam
    Son, Jaeman
    Lee, Se Byeong
    Lim, Youngkyung
    Park, Sung Yong
    Cho, Kwanho
    Yoon, Myonggeun
    Shin, Dongho
    SENSORS, 2018, 18 (01):
  • [29] Gantry-based pencil beam scanning proton therapy for uveal melanoma: IMPT versus proton arc therapy
    Qi, Hang
    Hu, Lei
    Huang, Sheng
    Lee, Yen-Po
    Yu, Francis
    Chen, Qing
    Yang, Yunjie
    Kang, Minglei
    Zhai, Huifang
    Vermeulen, Milo
    Shim, Andy
    Park, Peter
    Ding, Xuanfeng
    Zhou, Jun
    Abramson, David H.
    Francis, Jasmine H.
    Simone, Charles B.
    Barker, Christopher A.
    Lin, Haibo
    RADIATION ONCOLOGY, 2025, 20 (01)
  • [30] Commissioning of the 4-D treatment delivery system for organ motion management in synchrotron-based scanning ion beams
    Ciocca, Mario
    Mirandola, Alfredo
    Molinelli, Silvia
    Russo, Stefania
    Mastella, Edoardo
    Vai, Alessandro
    Mairani, Andrea
    Magro, Giuseppe
    Pella, Andrea
    Donetti, Marco
    Valvo, Francesca
    Fossati, Piero
    Baroni, Guido
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (12): : 1667 - 1671