Quantitative Comparison of Color-Coded Parametric Imaging Technologies Based on Digital Subtraction and Digital Variance Angiography: A Retrospective Observational Study

被引:0
|
作者
Gog, Istvan [1 ,2 ]
Sotonyi, Peter [1 ]
Nemes, Balazs [3 ]
Kiss, Janos P. [2 ]
Szigeti, Krisztian [4 ]
Osvath, Szabolcs [2 ,4 ]
Gyano, Marcell [2 ,3 ]
机构
[1] Semmelweis Univ, Heart & Vasc Ctr, Dept Vac & Endovasc Surg, Varosmajor Utca 68, H-1122 Budapest, Hungary
[2] Kinepict Hlth Ltd, Szilagyi Erzsebet Fasor 31, H-1027 Budapest, Hungary
[3] Semmelweis Univ, Heart & Vasc Ctr, Dept Intervent Radiol, Varosmajor Utca 68, H-1122 Budapest, Hungary
[4] Semmelweis Univ, Dept Biophys & Radiat Biol, Tuzolto U 37-47, H-1094 Budapest, Hungary
关键词
color-coded parametric imaging; digital subtraction angiography; digital variance angiography; time-density curve; peripheral artery disease; critical limb ischemia; ARTERY; FLOW;
D O I
10.3390/jimaging10100260
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
The evaluation of hemodynamic conditions in critical limb-threatening ischemia (CLTI) patients is inevitable in endovascular interventions. In this study, the performance of color-coded digital subtraction angiography (ccDSA) and the recently developed color-coded digital variance angiography (ccDVA) was compared in the assessment of key time parameters in lower extremity interventions. The observational study included 19 CLTI patients who underwent peripheral vascular intervention at our institution in 2020. Pre- and post-dilatational images were retrospectively processed and analyzed by a commercially available ccDSA software (Kinepict Medical Imaging Tool 6.0.3; Kinepict Health Ltd., Budapest, Hungary) and by the recently developed ccDVA technology. Two protocols were applied using both a 4 and 7.5 frames per second acquisition rate. Time-to-peak (TTP) parameters were determined in four pre- and poststenotic regions of interest (ROI), and ccDVA values were compared to ccDSA read-outs. The ccDVA technology provided practically the same TTP values as ccDSA (r = 0.99, R-2 = 0.98, p < 0.0001). The correlation was extremely high independently of the applied protocol or the position of ROI; the r value was 0.99 (R-2 = 0.98, p < 0.0001) in all groups. A similar correlation was observed in the change in passage time (r = 0.98, R-2 = 0.96, p < 0.0001). The color-coded DVA technology can reproduce the same hemodynamic data as a commercially available DSA-based software; therefore, it has the potential to be an alternative decision-supporting tool in catheter labs.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Morphological study of branches of lateral femoral circumflex artery based on digital subtraction angiography
    Yang, Lin
    Cheng, Junnan
    Liu, Zhijin
    Ju, Jihui
    Liu, Shengzhe
    Li, Lei
    Zhang, Tao
    Hou, Ruixing
    JOURNAL OF PLASTIC RECONSTRUCTIVE AND AESTHETIC SURGERY, 2023, 80 : 18 - 24
  • [32] Intravascular ultrasound versus digital subtraction angiography: direct comparison of intraluminal diameter measurements in pediatric and adolescent imaging
    Gill, Anne E.
    Ciszak, Tadi
    Braun, Hayley
    Hawkins, C. Matthew
    PEDIATRIC RADIOLOGY, 2017, 47 (04) : 450 - 457
  • [33] Intravascular ultrasound versus digital subtraction angiography: direct comparison of intraluminal diameter measurements in pediatric and adolescent imaging
    Anne E. Gill
    Tadi Ciszak
    Hayley Braun
    C. Matthew Hawkins
    Pediatric Radiology, 2017, 47 : 450 - 457
  • [34] Severity of Liver Cirrhosis Evaluated by Digital Subtraction Angiography Using Quantitative Color-Coding Analysis Before Transarterial Embolization
    Giiang, Lung-Hui
    Liu, Chang-Hsien
    Yu, Chih-Yung
    Lin, Te-Pao
    Chen, Hsiang-Cheng
    Juan, Chun-Jung
    Chou, Yu-Ching
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2023, 43 (05) : 566 - 573
  • [35] Comparison of Magnetic Resonance Angiography and Digital Subtraction Angiography for the Assessment of Infrapopliteal Arterial Occlusive Lesions, Based on the TASC II Classification Criteria
    Baubeta Fridh, Erik
    Ludwigs, Karin
    Svalkvist, Angelica
    Andersson, Manne
    Nordanstig, Joakim
    Falkenberg, Marten
    Johnsson, Ase A.
    DIAGNOSTICS, 2020, 10 (11)
  • [36] Treatment Evaluation of Flow-Limiting Stenoses of the Superficial Femoral and Popliteal Artery by Parametric Color-Coding Analysis of Digital Subtraction Angiography Series
    Michael Kostrzewa
    Kerim Kara
    Lothar Pilz
    Hannelore Mueller-Muertz
    Nils Rathmann
    Stefan O. Schoenberg
    Steffen J. Diehl
    CardioVascular and Interventional Radiology, 2017, 40 : 1147 - 1154
  • [37] Comparison of extracranial artery stenosis and cerebral blood flow, assessed by quantitative magnetic resonance, using digital subtraction angiography as the reference standard
    Cai, Jingjing
    Wu, Dan
    Mo, Yongqian
    Wang, Anxin
    Hu, Shiyu
    Ren, Lijie
    MEDICINE, 2016, 95 (46)
  • [38] Basal vein in the posterior incisural space: An anatomical comparison based on autopsy, digital subtraction angiography and computed tomographic venography
    Chen, Dan
    Liu, Bin
    Jin, Chuanbing
    Cao, Yongsheng
    Han, Hui
    Deng, Xuefei
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (02) : 319 - 327
  • [39] Direct percutaneous puncture digital-subtraction-angiography-based classification and treatment selection for soft-tissue arteriovenous malformations of maxillofacial region: a retrospective study
    Xun, H.
    Li, K.
    Li, X.
    Liu, Y.
    Du, X.
    Qin, Z.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2019, 48 (02) : 181 - 186
  • [40] Digital subtraction angiography findings of stroke in young adult population: a multi-center record-based study
    Kumara Tini
    Eric Hartono Tedyanto
    Deddy Andaka
    Nyoman Angga Krishna Pramana
    I Putu Eka Widyadharma
    The Egyptian Journal of Neurology, Psychiatry and Neurosurgery, 59