High resolving power electrospray drift tube ion mobility spectrometer with heated desolvation tube

被引:0
|
作者
Boillat, Marc-Aurele [1 ]
Hauser, Peter C. [1 ]
机构
[1] Univ Basel, Dept Chem, Klingelbergstr 80, CH-4056 Basel, Switzerland
关键词
Ion mobility spectrometry; Drift tube; High resolving power; Electrospray; Mass spectrometry; IONIZATION; MASS;
D O I
10.1016/j.aca.2024.343574
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background: The separation performance of drift tube ion mobility spectrometers is usually relatively weak with resolving powers well below 100. Improving this aspect requires, besides the optimization of other parameters, the challenging increase of the drift voltage as deduced from fundamental equations describing the broadening of a drifting ion swarm. We recently succeeded in constructing an improved high voltage instrument equipped with an electrospray source capable of analysing liquid samples with resolving powers above 200. This instrument was further perfected with a newly developed heated desolvation tube guiding the ions to the injection gate. The electrospray desolvation process was thus enhanced while the separation region was maintained at room temperature to keep diffusional band broadening to a minimum. A further modification was also made to the detector circuitry in order to reduce its noise. Results: To demonstrate the performance of the instrument several challenging ion mixtures, including two herbicides, a polymer mixture, and isomers of carnitine esters were successfully separated and detected. The resolving powers, calculated from the arrival time and peak width at half height, ranged from 123 to 228. The latter value represents 96 % of the theoretically achievable resolving power at the working drift voltage of 24.1 kV.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Ion Profiling in an Ambient Drift Tube-Ion Mobility Spectrometer Using a High Pixel Density Linear Array Detector IonCCD
    Davila, Stephen J.
    Hadjar, Omar
    Eiceman, Gary A.
    ANALYTICAL CHEMISTRY, 2013, 85 (14) : 6716 - 6722
  • [22] Evaluation of drift gas selection in complex sample analyses using a high performance drift tube ion mobility-QTOF mass spectrometer
    Kurulugama, Ruwan. T.
    Darland, Ed
    Kuhlmann, Frank
    Stafford, George
    Fjeldsted, John
    ANALYST, 2015, 14 (20) : 6834 - 6844
  • [23] Effect of electric field of drift tube on performance of a vacuum, ultraviolet ionization-ion mobility spectrometer
    Shi Yingguo
    Shao Shiyong
    Li Anlin
    Yao Lian
    Wang Bin
    Li Fang
    Wang Junde
    Li Haiyang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2006, 34 (09) : 1353 - 1356
  • [24] Development of a Modular, Open-Source, Reduced-Pressure, Drift Tube Ion Mobility Spectrometer
    Buzitis, Nathan W.
    Clowers, Brian H.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2024, 35 (04) : 804 - 813
  • [25] Ion Trapping for Ion Mobility Spectrometry Measurements in a Cyclical Drift Tube
    Glaskin, Rebecca S.
    Ewing, Michael A.
    Clemmer, David E.
    ANALYTICAL CHEMISTRY, 2013, 85 (15) : 7003 - 7008
  • [26] Separation of biologically relevant isomers on an Orbitrap mass spectrometer using high-resolution drift tube ion mobility and varied drift gas mixtures
    Kaszycki, Julia L.
    La Rotta, Aurelio
    Colsch, Benoit
    Fenaille, Francois
    Dauly, Claire
    Kamleh, Anas
    Wu, Ching
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2019, 33 : 3 - 10
  • [27] Electrospray Ion Mobility Spectrometer Based on Flexible Printed-Circuit Board Electrodes with Improved Resolving Power
    Boillat, Marc-Aurele
    Rakus, Julian M.
    Hauser, Peter C.
    ANALYTICAL CHEMISTRY, 2023, 95 (28) : 10777 - 10784
  • [28] Expanded theory for the resolving power of a linear ion mobility spectrometer
    Spangler, GE
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 220 (03) : 399 - 418
  • [29] Measurements of reduced mobility of standard compounds by high resolving power ion mobility spectrometer in remote laboratories
    D. M. Chernyshev
    I. S. Frolov
    A. S. Frolov
    M. S. Mukhanov
    A. A. Sysoev
    Journal of Analytical Chemistry, 2011, 66 : 1253 - 1257
  • [30] Measurements of reduced mobility of standard compounds by high resolving power ion mobility spectrometer in remote laboratories
    Chernyshev, D. M.
    Frolov, I. S.
    Frolov, A. S.
    Mukhanov, M. S.
    Sysoev, A. A.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 66 (13) : 1253 - 1257