Effect of Respirophasic Displacement of the Inferior Vena Cava on Size Measurement in 2-D Ultrasound Imaging

被引:0
作者
Policastro, Piero [1 ]
Ermini, Leonardo [2 ]
Civera, Stefania [3 ]
Albani, Stefano [3 ,4 ]
Musumeci, Giuseppe [3 ]
Roatta, Silvestro [2 ]
Mesin, Luca [1 ]
机构
[1] Politecn Torino, Dept Elect & Telecommun, Math Biol & Physiol, Turin, Italy
[2] Univ Torino, Dept Neurosci, Turin, Italy
[3] Osped Ordine Mauriziano Torino, Div Cardiol, Turin, Italy
[4] Umberto Parini Reg Hosp, Div Cardiol, Aosta, Italy
关键词
Inferior vena cava; Ultrasound; Tracking;
D O I
10.1016/j.ultrasmedbio.2024.07.005
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective: Volume status assessment of a patient by ultrasound (US) imaging of the inferior vena cava (IVC) is important for the diagnosis and prognosis of various clinical conditions. In order to improve the clinical investigation of IVC, which is mainly based on unidirectional US (in M-mode), automated processing of 2-D US scans (in Bmode) has enabled tissue movement tracking on the visualized plane and can average this in various directions. However, IVC geometry outside of the visualized plane is not under control and could result in errors that have not yet been evaluated. Methods: We used a method that integrates information from long- and short-axis IVC views (simultaneously acquired in the X-plane) to assess challenges in IVC diameter estimations using 2-D US scans in eight healthy subjects. Results: Relative movements between the US probe and IVC induced the following problems when assessing IVC diameter via 2-D view: a median error (i.e., absolute difference with respect to diameter measured in the X-plane) of 17% using 2-D US scans in the long-axis view of the IVC affected by medio-lateral displacements (median: 4 mm); and a median error of 7% and 9% when measuring the IVC diameter from a short-axis view in the presence of pitch angle (median: 0.12 radians) and cranio-caudal movement (median: 15 mm), respectively. Conclusion: Relative movements in the IVC that are out of view of B-mode scans cannot be detected, which results in challenges in IVC diameter estimation.
引用
收藏
页码:1785 / 1792
页数:8
相关论文
共 42 条
  • [31] Reduction of the decorrelation effect due to tissue lateral displacement by 2-D Spatial Comprehensive Correlation in elastography
    Ding, CX
    Bai, J
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 2351 - 2354
  • [32] Cardiac 2-D Shear Wave Imaging Using a New Dedicated Clinical Ultrasound System: A Phantom Study
    Saloux, Eric
    Le Garrec, Morgane
    Menet, Nina
    Dillon, Ludovic
    Simard, Christophe
    Fraschini, Christophe
    Manrique, Alain
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2024, 50 (06) : 843 - 851
  • [33] 2-D Bayesian Displacement Estimation Improves Contrast-to-Noise/Resolution Trade-off in Shear Wave Elasticity Imaging
    Ersepke, Thomas
    Kranemann, Tim C.
    Schmitz, Georg
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1379 - 1382
  • [34] EVALUATION OF INDIRECT METHODS FOR MOTION COMPENSATION IN 2-D FOCAL LIVER LESION CONTRAST-ENHANCED ULTRASOUND (CEUS) IMAGING
    Bakas, Spyridon
    Doulgerakis-Kontoudis, Matthaios
    Hunter, Gordon J. A.
    Sidhu, Paul S.
    Makris, Dimitrios
    Chatzimichail, Katerina
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2019, 45 (06) : 1380 - 1396
  • [35] Full 2D displacement vector and strain tensor estimation for superficial tissue using beam-steered ultrasound imaging
    Hansen, H. H. G.
    Lopata, R. G. P.
    Idzenga, T.
    de Korte, C. L.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (11) : 3201 - 3218
  • [36] Imaging Transverse Isotropic Properties of Muscle by Monitoring Acoustic Radiation Force Induced Shear Waves Using a 2-D Matrix Ultrasound Array
    Wang, Michael
    Byram, Brett
    Palmeri, Mark
    Rouze, Ned
    Nightingale, Kathryn
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (09) : 1671 - 1684
  • [37] Accuracy of tumor size measurement: comparison of B-mode ultrasound, strain elastography, and 2D and 3D shear wave elastography with histopathological lesion size
    Farrokh, Andre
    Maass, Nicolai
    Treu, Linn
    Heilmann, Thorsten
    Schafer, Fritz K. W.
    ACTA RADIOLOGICA, 2019, 60 (04) : 451 - 458
  • [38] 2-D SHEAR WAVE ELASTOGRAPHY FOR FOCAL LESIONS IN LIVER PHANTOMS: EFFECTS OF BACKGROUND STIFFNESS, DEPTH AND SIZE OF FOCAL LESIONS ON STIFFNESS MEASUREMENT
    Hwang, Jeong Ah
    Jeong, Woo Kyoung
    Song, Kyoung Doo
    Kang, Kyung A.
    Lim, Hyo K.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2019, 45 (12) : 3261 - 3268
  • [39] Design and Test of A Fully Controllable 64x128 2-D CMUT Array Integrated with Reconfigurable Frontend ASICs for Volumetric Ultrasound Imaging
    Kim, Bae-Hyung
    Kim, Youngil
    Lee, Seunghun
    Cho, Kyungil
    Song, Jongkeun
    2012 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2012, : 77 - 80
  • [40] Methods of Determining the Size of the Adult Liver Using 2D Ultrasound: A Systematic Review of Articles Reporting Liver Measurement Techniques
    Childs, Jessie T.
    Esterman, Adrian J.
    Phillips, Maureen
    Thoirs, Kerry A.
    Turner, Richard C.
    JOURNAL OF DIAGNOSTIC MEDICAL SONOGRAPHY, 2014, 30 (06) : 296 - 306