Technology of near infrared spectroscopy analysis system design based on fixed grating and charge-coupled device

被引:0
作者
Zhang, Yinqiao [1 ]
Zhao, Bo [1 ]
Wang, Hui [1 ]
Mao, Wenhua [1 ]
Zhang, Xiaochao [1 ]
机构
[1] Chinese Academy of Agricultural Mechanization Sciences
来源
Nongye Gongcheng Xuebao/Transactions of the Chinese Society of Agricultural Engineering | 2009年 / 25卷 / 09期
关键词
Agricultural products; CCD temperature control; Design; Light system; Near infrared spectroscopy; Quality detection;
D O I
10.3969/j.issn.1002-6819.2009.09.031
中图分类号
学科分类号
摘要
Near infrared(NIR) spectroscopy analysis system was designed based on a fixed grating and a linear array charge-coupled device (CCD). The main content includes CCD drive circuit, spectrum signal acquisition circuit, fixed grating optical system, light source control system, CCD temperature control system and sample turntable control system. It was verified that the NIR spectral analysis method was feasible in testing the quality of grain by using the MPA Fourier spectrometer such as moisture and protein components of wheat. A variety of pretreatment methods were used to optimize the spectral data, and the optimal treatment means was the first derivative with multiplicative scatter correction. The CA-06 instrument performance and accuracy were verified, and wheat moisture and protein models were established.
引用
收藏
页码:176 / 181
页数:5
相关论文
共 15 条
  • [1] Zhang H., Research of AOTF based NIR ionstruments, (2005)
  • [2] Wang H., Improvement design for fast monitor of grain quality based on NIR, (2007)
  • [3] Wang Z., Li B., Wang W., A spectral data system for collecting and processing based on ilx511, Instrument Users, 6, 12, pp. 58-59, (2005)
  • [4] Wang Z., Zhang J., Bao G., Application of PLD using in digital signal processing, Optoelectronic Technology Applications, 20, 6, pp. 52-55, (2005)
  • [5] Wang L., Wen Z., Xiang X., Design and simulation of the NIR Micro-Spectrometer optical system, Spectroscopy and Spectral Analysis, 29, 6, pp. 1721-1725, (2009)
  • [6] Gao F., Gao M., Implementation of embedded proportioning weighing recorder based on MiniARM, Instrument Technique and Sensor, 12, pp. 41-43, (2008)
  • [7] Peng Y., Zhao Y., High precision temperature acquiring system, Instrument Technique and Sensor, 5, pp. 45-46, (2007)
  • [8] Zhang L., Li B., The application of RS-232 autoshutdown transceiver MAX3221, Journal of Changzhou Institute of Technology, 20, 4, pp. 27-30, (2007)
  • [9] Liang X., Yao Q., High stability and highly integrated laser diode driver, Semiconductor Optoelectronics, 28, 3, pp. 427-429, (2007)
  • [10] Dou L., Zheng Y., Huo Y., Application of 1-wire BUS in tempelature monitoring device in agriculture, Journal of Agricultural Mechanization Research, 31, 7, pp. 210-212, (2009)