Development of 20 kHz swept source optical coherence tomography system

被引:3
|
作者
Wu, Tong [1 ]
Ding, Zhihua [1 ]
机构
[1] State Key Laboratory of Modern Optical Instrumentation, Zhejiang University
来源
Zhongguo Jiguang/Chinese Journal of Lasers | 2009年 / 36卷 / 02期
关键词
Medical optics and biotechnology; Optical coherence tomography; Resolution; Swept source; Wave number caliberation;
D O I
10.3788/CJL20093602.0503
中图分类号
学科分类号
摘要
The self-developed swept source optical coherence tomography (SS-OCT) system is reported. Based on a high speed scanning laser source, the system realizes high speed A-scan rate of 20 kHz from 500 Hz A-scan rate of time-domain OCT. To realize linear calibration of wave number space, a pre-calibration method based on the SS-OCT system is used. Because of the non-Gaussian spectrum output of the swept source, a method based on window function is used for spectrum reshaping in the spectrum domain. To eliminate the direct current(DC) and auto-correlation term in the interference spectrum, the common-rejection of balanced detection, and a software method of subtract the mean value were used. The axial and lateral resolution of the SS-OCT system achieves 14 μm and 12 μm, respectively. The ranging depth of the SS-OCT achieves 3.9 mm. High speed optical coherence tomography images of finger-pad organism using the SS-OCT system were obtained.
引用
收藏
页码:503 / 508
页数:5
相关论文
共 23 条
  • [1] Huang D., Swanson E.A., Lin C.P., Optical coherence tomography, Science, 254, 5035, pp. 1178-1181, (1991)
  • [2] Yu X., Ding Z., Chen Y., Development of fiber-based optical coherence tomographic imaging system, Acta Optica Sinica, 26, 2, pp. 235-238, (2006)
  • [3] Meng J., Ding Z., Zhou L., Axial superresolution in optical coherence tomography, Acta Photonica Sinica, 37, 3, pp. 533-536, (2008)
  • [4] Yang Y., Ding Z., Yu X., Dispersion compensation in OCT system by rapid scanning optical delay line, Acta Photonica Sinica, 37, 1, pp. 21-24, (2008)
  • [5] Wang L., Ding Z., Shi G., Fiber-based optical coherence tomography imaging system with rapid scanning optical delay line as phase modulator, Chinese J. Lasers, 35, 3, pp. 472-476, (2008)
  • [6] Yun S., Tearney G., de Boer J., High-speed optical frequency-domain imaging, Opt. Express, 11, 22, pp. 2953-2963, (2003)
  • [7] Fercher F., Hitzenberger C.K., Kamp G., Measurement of intraocular distances by backscattering spectral interferometry, Opt. Commun., 117, pp. 443-448, (1995)
  • [8] Hausler G., Lindner M.W., Coherence radar and spectral radar -new tools for dermatological diagnosis, J. Biomed. Opt., 3, pp. 21-31, (1998)
  • [9] Chinn S.R., Swanson E.A., Fujimoto J.G., Optical coherence tomography using a frequency-tunable optical source, Opt. Lett., 22, 5, pp. 340-342, (1997)
  • [10] Leitgeb R., Hitzenberger C., Fercher A., Performance of fourier domain vs. time domain optical coherence tomography, Opt. Express, 11, 8, pp. 889-894, (2003)