Evaluation of 3D Printed Anthropomorphic Head Phantom for Calibration of TLD in Eye Lens Dosimeter

被引:0
|
作者
Endarko E. [1 ]
Pitaloka A.W. [1 ]
Inayah A.A. [1 ]
Ramadhan F.R. [1 ]
Kardianto K. [2 ]
Rahayuningsih B. [2 ]
机构
[1] Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS-Sukolilo, East Java, Surabaya
[2] Balai Pengamanan Fasilitas Kesehatan Gubeng, East Java, Surabaya
关键词
Calibration; Dosimetry; Phantom; Radiation; Thermoluminescent Dosimetry;
D O I
10.31661/jbpe.v0i0.2109-1401
中图分类号
学科分类号
摘要
Background: Calibration of Thermo Luminescent Dosimetry (TLD) in eye lens dosimeter requires a standard phantom. The use of anthropomorphic phantoms in calibration needs evaluation. Objective: This study aimed to analyze the angular response of the TLD on the fabricated 3D anthropomorphic head phantom and Computerized Imaging Reference Systems (CIRS)-Computed Tomography (CT) dose phantom as a standard phantom irradiated with Cs-137 and to compare the absorbed dose and linear attenuation for both phantoms. Hp (3) analysis, conversion coefficient (hpK (3)), and calibration factor (CF) are also investigated. Material and Methods: In this experimental study, the fabricated 3D printed anthropomorphic head phantom was analyzed using polylactic acid (PLA) with the skull and then filled with the artificial brain and cerebrospinal fluid (CSF) as a test phantom. TLD-700H and TLD Reader Harshaw 6600 plus were used to analyze the angular response of Cs-137 radiation and to determine the absorbed dose and linear attenuation coefficient of test and standard phantoms. Results: The effect of the angle of radiation source towards TLD reading at the anthropomorphic head phantom has a similar value to the standard phantom with a calibration factor ranging from 0.82 to 1. The absorbed dose measurement and the linear attenuation coefficient of the anthropomorphic head phantom with the standard phantom have different values of 2.52 and 3.78%, respectively. Conclusion: The fabricated 3D printed anthropomorphic head phantom has good potential as an alternative to standard phantoms for TLD calibration in eye lens do-simeter. © 2023, Shiraz University of Medical Sciences. All rights reserved.
引用
收藏
页码:193 / 202
页数:9
相关论文
共 50 条
  • [21] Abdo-Man: a 3D-printed anthropomorphic phantom for validating quantitative SIRT
    Gear, Jonathan I.
    Cummings, Craig
    Craig, Allison J.
    Divoli, Antigoni
    Long, Clive D. C.
    Tapner, Michael
    Flux, Glenn D.
    EJNMMI PHYSICS, 2016, 3 (01):
  • [22] Optimising total knee replacement imaging: a novel 3D printed PET/CT anthropomorphic phantom for metal artefact simulation
    Assiri, Rajeh
    Knapp, Karen
    Fulford, Jon
    Chen, Junning
    EJNMMI PHYSICS, 2024, 11 (01)
  • [23] Optimising total knee replacement imaging: a novel 3D printed PET/CT anthropomorphic phantom for metal artefact simulation
    Rajeh Assiri
    Karen Knapp
    Jon Fulford
    Junning Chen
    EJNMMI Physics, 11
  • [24] 3D printed non-uniform anthropomorphic phantoms for quantitative SPECT
    Jessen, Lovisa
    Gustafsson, Johan
    Ljungberg, Michael
    Curkic-Kapidzic, Selma
    Imsirovic, Muris
    Sjogreen-Gleisner, Katarina
    EJNMMI PHYSICS, 2024, 11 (01)
  • [25] Preclinical dose verification using a 3D printed mouse phantom for radiobiology experiments
    Esplen, Nolan
    Therriault-Proulx, Francois
    Beaulieu, Luc
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2019, 46 (11) : 5294 - 5303
  • [26] 3D printed non-uniform anthropomorphic phantoms for quantitative SPECT
    Lovisa Jessen
    Johan Gustafsson
    Michael Ljungberg
    Selma Curkic-Kapidzic
    Muris Imsirovic
    Katarina Sjögreen-Gleisner
    EJNMMI Physics, 11
  • [27] Realistic Head Phantom for Evaluation of Brain Stroke Localization Methods Using 3D Printer
    Lee, Juneseok
    Bang, Jihoon
    Choi, Jaehoon
    JOURNAL OF ELECTROMAGNETIC ENGINEERING AND SCIENCE, 2016, 16 (04) : 254 - 258
  • [28] A customized anthropomorphic 3D-printed phantom to reproducibility assessment in computed tomography: an oncological case study
    Cavaliere, Carlo
    Baldi, Dario
    Brancato, Valentina
    Aiello, Marco
    Salvatore, Marco
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [29] Development of a realistic 3D printed eye lens dosemeter using CAD integrated with Monte Carlo simulation
    Santos, M. F.
    Cassola, V.
    Kramer, R.
    Costa, J. V.
    Andrade, M. E. A.
    Asfora, V. K.
    Khoury, H. J.
    Barros, V. S. M.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (01)
  • [30] A customizable anthropomorphic phantom for dosimetric verification of 3D-printed lung, tissue, and bone density materials
    Tino, Rance Bolislis
    Yeo, Adam Unjin
    Brandt, Milan
    Leary, Martin
    Kron, Tomas
    MEDICAL PHYSICS, 2022, 49 (01) : 52 - 69