3D printed polymeric stent design: Mechanical testing and computational modeling

被引:1
作者
Canalejo-Codina, Francesc [1 ,2 ,3 ]
Cano-Morenilla, Mariola [4 ,5 ]
Martorell, Jordi [2 ]
Balcells, Mercedes [2 ,3 ]
Pegueroles, Marta [4 ,5 ]
Garcia-Granada, Andres A. [1 ]
机构
[1] Univ Ramon Llull, IQS Sch Engn, Ind Prod Engn Grp GEPI, Via Augusta 390, Barcelona 08017, Spain
[2] Univ Ramon Llull, IQS Sch Engn, Vasc Engn & Appl Biomed Grp GEVAB, Barcelona, Spain
[3] MIT, Inst Med Engn & Sci, Cambridge, MA USA
[4] Univ Politecn Cataluna, Ctr Res Multiscale Sci & Engn CCEM, Dept Mat Sci & Engn, Biomat Biomech & Tissue Engn Grp, Barcelona, Spain
[5] Univ Politecn Cataluna, Res Ctr Biomed Engn CREB, EEBE, Barcelona, Spain
关键词
Additive manufacturing; Bioresorbable stent; Poly(L-lactide-co; epsilon -caprolactone); Mechanical performance; Crush resistance; Finite element analysis; ELUTING STENTS; RADIAL FORCE; DEGRADATION; RESTENOSIS; EXPANSION; IMPLANTS; STRESSES; BEHAVIOR; PLA;
D O I
10.1016/j.matdes.2024.113395
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Polymer-based bioresorbable scaffolds (BRS) aim to reduce the long-term issues associated with metal stents. Yet, first-generation BRS designs experienced a significantly higher rate of clinical failures compared to permanent implants. This prompted the development of alternative scaffolds, such as the poly(L-lactide-co-epsilon-caprolactone) (PLCL) solvent-casted stent, whose mechanical performance has yet to be addressed. This study examines the mechanical behavior of this novel scaffold across a wide range of parallel and radial compression diameters. The analysis highlights the scaffold's varying responses under different loading conditions and provides insights into interpreting simulation model parameters to accurately reflect experimental results. Stents demonstrated a parallel crush resistance of 0.11 N/mm at maximum compression, whereas the radial forces were significantly higher, reaching up to 1.80 N/mm. Additionally, the parallel test keeps the stent in the elastic regime, with almost no regions exceeding 50 MPa of stress, while the radial test causes significant structural deformation, with localized plastic strain reaching up to 30 %. Results showed that underestimating yield strain in computational models leads to discrepancies with experimental results, being 5 % the most accurate value for matching computational and experimental results for PLCL solvent-casted stents. This comprehensive approach is vital for optimizing BRS design and predicting clinical performance.
引用
收藏
页数:11
相关论文
共 76 条
  • [1] New Drug-Eluting Stents An Overview on Biodegradable and Polymer-Free Next-Generation Stent Systems
    Abizaid, Alexandre
    Costa, J. Ribamar, Jr.
    [J]. CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (04) : 384 - 393
  • [2] A Review on the Modeling of the Elastic Modulus and Yield Stress of Polymers and Polymer Nanocomposites: Effect of Temperature, Loading Rate and Porosity
    Alasfar, Reema H.
    Ahzi, Said
    Barth, Nicolas
    Kochkodan, Viktor
    Khraisheh, Marwan
    Koc, Muammer
    [J]. POLYMERS, 2022, 14 (03)
  • [3] American Society for Testing and Materials, 2021, ASTM D882-18: Standard Test Method for Tensile Properties of Thin Plastic Sheeting'
  • [4] Auxetic coronary stent endoprosthesis: fabrication and structural analysis
    Amin, Faisal
    Ali, Murtaza Najabat
    Ansari, Umar
    Mir, Mariam
    Minhas, Muhammad Asim
    Shahid, Wakeel
    [J]. JOURNAL OF APPLIED BIOMATERIALS & FUNCTIONAL MATERIALS, 2015, 13 (02): : E127 - E135
  • [5] Fabrication and in vitro deployment of a laser-activated shape memory polymer vascular stent
    Baer, Geraldine M.
    Small, Ward, IV
    Wilson, Thomas S.
    Benett, William J.
    Matthews, Dennis L.
    Hartman, Jonathan
    Maitland, Duncan J.
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2007, 6 (1)
  • [6] Impact of strain rate on mechanical properties of polylatic acid fabricated by fusion deposition modeling
    Balasubramanian, Muthuselvan
    Saravanan, R.
    Shanmugam, Vigneshwaran
    [J]. POLYMERS FOR ADVANCED TECHNOLOGIES, 2024, 35 (03)
  • [7] Design and Shape Optimization of a Biodegradable Polymeric Stent for Curved Arteries Using FEM
    Baradaran, Yasaman
    Baghani, Mostafa
    Kazempour, Morteza
    Hosseini, Seyed Kianoosh
    Karimpour, Morad
    Baniassadi, Majid
    [J]. FRONTIERS IN MECHANICAL ENGINEERING-SWITZERLAND, 2021, 7
  • [8] Mechanical properties of the drug-eluting bioresorbable magnesium scaffold compared with polymeric scaffolds and a permanent metallic drug-eluting stent
    Barkholt, Trine O.
    Webber, Bruce
    Holm, Niels R.
    Ormiston, John A.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 96 (07) : E674 - E682
  • [9] An Overview of Mechanical Properties and Material Modeling of Polylactide (PLA) for Medical Applications
    Bergstroem, Joergen S.
    Hayman, Danika
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2016, 44 (02) : 330 - 340
  • [10] Cardiovascular stents: overview, evolution, and next generation
    Borhani S.
    Hassanajili S.
    Ahmadi Tafti S.H.
    Rabbani S.
    [J]. Progress in Biomaterials, 2018, 7 (3) : 175 - 205