Deep learning-based automatic contour quality assurance for auto-segmented abdominal MR-Linac contours

被引:1
|
作者
Zarenia, Mohammad [1 ,2 ]
Zhang, Ying [1 ,3 ]
Sarosiek, Christina [1 ]
Conlin, Renae [1 ]
Amjad, Asma [1 ]
Paulson, Eric [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
[2] MedStar Georgetown Univ Hosp, Dept Radiat Med, Washington, DC 53226 USA
[3] Univ Texas Southwestern Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 21期
关键词
adaptive radio therapy; MR-Linac; deep learning; contour quality assurance; abdominal MRI; RADIOTHERAPY;
D O I
10.1088/1361-6560/ad87a6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Deep-learning auto-segmentation (DLAS) aims to streamline contouring in clinical settings. Nevertheless, achieving clinical acceptance of DLAS remains a hurdle in abdominal MRI, hindering the implementation of efficient clinical workflows for MR-guided online adaptive radiotherapy (MRgOART). Integrating automated contour quality assurance (ACQA) with automatic contour correction (ACC) techniques could optimize the performance of ACC by concentrating on inaccurate contours. Furthermore, ACQA can facilitate the contour selection process from various DLAS tools and/or deformable contour propagation from a prior treatment session. Here, we present the performance of novel DL-based 3D ACQA models for evaluating DLAS contours acquired during MRgOART. Approach. The ACQA model, based on a 3D convolutional neural network (CNN), was trained using pancreas and duodenum contours obtained from a research DLAS tool on abdominal MRIs acquired from a 1.5 T MR-Linac. The training dataset contained abdominal MR images, DL contours, and their corresponding quality ratings, from 103 datasets. The quality of DLAS contours was determined using an in-house contour classification tool, which categorizes contours as acceptable or edit-required based on the expected editing effort. The performance of the 3D ACQA model was evaluated using an independent dataset of 34 abdominal MRIs, utilizing confusion matrices for true and predicted classes. Main results. The ACQA predicted 'acceptable' and 'edit-required' contours at 72.2% (91/126) and 83.6% (726/868) accuracy for pancreas, and at 71.2% (79/111) and 89.6% (772/862) for duodenum contours, respectively. The model successfully identified false positive (extra) and false negative (missing) DLAS contours at 93.75% (15/16) and %99.7 (438/439) accuracy for pancreas, and at 95% (57/60) and 98.9% (91/99) for duodenum, respectively. Significance. We developed 3D-ACQA models capable of quickly evaluating the quality of DLAS pancreas and duodenum contours on abdominal MRI. These models can be integrated into clinical workflow, facilitating efficient and consistent contour evaluation process in MRgOART for abdominal malignancies.
引用
收藏
页数:8
相关论文
共 38 条
  • [1] Robust deep learning-based forward dose calculations for VMAT on the 1.5T MR-linac
    Tsekas, G.
    Bol, G. H.
    Raaymakers, B. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (22):
  • [2] A Deep Learning-based correction to EPID dosimetry for attenuation and scatter in the Unity MR-Linac system
    Olaciregui-Ruiz, Igor
    Torres-Xirau, Iban
    Teuwen, Jonas
    van der Heide, Uulke A.
    Mans, Anton
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 71 : 124 - 131
  • [3] Deep learning application for abdominal organs segmentation on 0.35 T MR-Linac images
    Zhou, You
    Lalande, Alain
    Chevalier, Cedric
    Baude, Jeremy
    Aubignac, Leone
    Boudet, Julien
    Bessieres, Igor
    FRONTIERS IN ONCOLOGY, 2024, 13
  • [4] Deep learning based modelling of radiation dose deposition in a 1.5 T MR-Linac
    Schneider, Moritz
    Gutwein, Simon
    Nyazi, Maximilian
    Thorwarth, Daniela
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3674 - S3676
  • [5] Deep learning based automatic contour refinement for inaccurate auto-segmentation in MR-guided adaptive radiotherapy
    Ding, Jie
    Zhang, Ying
    Amjad, Asma
    Sarosiek, Christina
    Dang, Nguyen Phuong
    Zarenia, Mohammad
    Li, X. Allen
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (05):
  • [6] Is it necessary to perform measurement-based patient-specific quality assurance for online adaptive radiotherapy with Elekta Unity MR-Linac?
    Xu, Yuan
    Xia, Wenlong
    Ren, Wenting
    Ma, Min
    Men, Kuo
    Dai, Jianrong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (02):
  • [7] Stability of conventional and machine learning-based tumor auto-segmentation techniques using undersampled dynamic radial bSSFP acquisitions on a 0.35 T hybrid MR-linac system
    Friedrich, Florian
    Hoerner-Rieber, Juliane
    Renkamp, C. Katharina
    Kluter, Sebastian
    Bachert, Peter
    Ladd, Mark E.
    Knowles, Benjamin R.
    MEDICAL PHYSICS, 2021, 48 (02) : 587 - 596
  • [8] A deep learning-based prediction model for gamma evaluation in patient-specific quality assurance
    Tomori, Seiji
    Kadoya, Noriyuki
    Takayama, Yoshiki
    Kajikawa, Tomohiro
    Shima, Katsumi
    Narazaki, Kakutarou
    Jingu, Keiichi
    MEDICAL PHYSICS, 2018, 45 (09) : 4055 - 4065
  • [9] Deep Learning-Based Automatic Segmentation of the Proximal Femur from MR Images
    Zeng, Guodong
    Zheng, Guoyan
    INTELLIGENT ORTHOPAEDICS: ARTIFICIAL INTELLIGENCE AND SMART IMAGE-GUIDED TECHNOLOGY FOR ORTHOPAEDICS, 2018, 1093 : 73 - 79
  • [10] Deep Learning Based on EfficientNet for Multiorgan Segmentation of Thoracic Structures on a 0.35 T MR-Linac Radiation Therapy System
    Chekroun, Mohammed
    Mourchid, Youssef
    Bessieres, Igor
    Lalande, Alain
    ALGORITHMS, 2023, 16 (12)