摘要
Background: A growing body of clinical evidence has demonstrated that intraosseous minimally invasive basivertebral nerve (BVN) ablation results in significant and durable improvements in vertebrogenic back pain. Thus, it is important to develop, refine and validate new and additional devices to accomplish this procedure. Methods: Using reconstructions of 31 patient computed tomography (CT) scans of the lumbosacral spine (L1-S1), the primary objective was to simulate the intravertebral placement of a novel multitined expandable electrode in bipolar configuration at the targeted ablation site and determine if the proper trajectories could be achieved in order for the device tips to be in the correct position for lesion formation at the BVN plexus. Successful device deployment required that the distance between tips was between 10 mm and 20 mm. Results: The mean distances between device tips ranged from 11.35 mm (L5) to 11.87 mm (L3), and there were no statistically significance differences across the six vertebral levels (F = 0.72, p = 0.61). The percentage of successful intraosseous device placements within the tip distance acceptable range (>= 10 mm to <= 20 mm) was 90% (162 of 180), with no tip-to-tip distances > 20 mm. There was a notable association between decreasing vertebral level and mean degree of angulation between contralateral devices ranging from 50.90 degrees at L1 to 91.51 degrees at S1, and the difference between across the six vertebral levels was significant (F = 89.5, p < 0.01). Conclusion: Feasibility evidence is provided from real world CT imaging data that validates using the multitined electrode for proper intraosseous placement within the vertebral body to effectively ablate the BVN plexus.
引用
收藏
页码:323 / 337
页数:15
相关论文
共 23 条
  • [1] Influence of Surgical Staples on Radiofrequency Ablation Using Multitined Expandable Electrodes
    Yusuke Sakuhara
    Tadashi Shimizu
    Daisuke Abo
    Yu Hasegawa
    Fumi Kato
    Yoshihisa Kodama
    Hiroki Shirato
    CardioVascular and Interventional Radiology, 2007, 30 : 1201 - 1205
  • [2] Radiofrequency Ablation of Renal Tumors with an Expandable Multitined Electrode: Results, Complications, and Pilot Evaluation of Cooled Pyeloperfusion for Collecting System Protection
    Olivier Rouvière
    Lionel Badet
    François Joseph Murat
    Jean Marie Maréchal
    Marc Colombel
    Xavier Martin
    Denis Lyonnet
    Albert Gelet
    CardioVascular and Interventional Radiology, 2008, 31 : 595 - 603
  • [3] Radiofrequency ablation of renal tumors with an expandable multitined electrode:: Results, complications, and pilot evaluation of cooled pyeloperfusion for collecting system protection
    Rouviere, Olivier
    Badet, Lionel
    Murat, Francois Joseph
    Marechal, Jean Marie
    Colombel, Marc
    Martin, Xavier
    Lyonnet, Denis
    Gelet, Albert
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 31 (03) : 595 - 603
  • [4] An Individually Controlled Multitined Expandable Electrode Using Active Cannula-Based Shape Morphing for On-Demand Conformal Radiofrequency Ablation Lesions
    Chai, Zhiping
    Lyu, Liangxiong
    Pu, Menghao
    Chen, Xianwen
    Zhu, Jiaqi
    Liang, Huageng
    Ding, Han
    Wu, Zhigang
    ADVANCED INTELLIGENT SYSTEMS, 2022, 4 (07)
  • [5] Radiofrequency ablation (RFA) of T1a renal cancer with externally cooled multitined expandable electrodes
    De Filippo, Massimo
    Ziglioli, Francesco
    Russo, Umberto
    Pagano, Paolo
    Brunese, Luca
    Bertelli, Elena
    Pagnini, Francesco
    Maestroni, Umberto
    RADIOLOGIA MEDICA, 2020, 125 (08): : 790 - 797
  • [6] Low back pain-related healthcare utilization following intraosseous basivertebral nerve radiofrequency ablation: a pooled analysis from three prospective clinical trials
    Mccormick, Zachary L.
    Curtis, Timothy
    Cooper, Amanda
    Wheatley, Margo
    Smuck, Matthew
    PAIN MEDICINE, 2024, 25 (01) : 20 - 32
  • [7] The Relationship Between Patient Demographic and Clinical Characteristics and Successful Treatment Outcomes After Basivertebral Nerve Radiofrequency Ablation: A Pooled Cohort Study of Three Prospective Clinical Trials
    Boody, Barrett S.
    Sperry, Beau P.
    Harper, Katrina
    Macadaeg, Kevin
    McCormick, Zachary L.
    PAIN MEDICINE, 2022, 23 (SUPPL 2) : S2 - S13
  • [8] Infusion of Hypertonic Saline into the Lung Parenchyma during Radiofrequency Ablation of the Lungs with Multitined Expandable Electrodes: Results Using a Porcine Model
    Iishi, Tatsuhiko
    Hiraki, Takao
    Mimura, Hidefumi
    Gobara, Hideo
    Kurose, Taichi
    Fujiwara, Hiroyasu
    Sakurai, Jun
    Yanai, Hiroyuki
    Yoshino, Tadashi
    Kanazawa, Susumu
    ACTA MEDICA OKAYAMA, 2009, 63 (03) : 137 - 144
  • [9] Fluoroscopic-guided radiofrequency ablation of the basivertebral nerve: application and analysis with multiple imaging modalities in an ovine model.
    Bergeron, JA
    Eskey, CJ
    Attawia, M
    Patel, SJ
    Ryan, TP
    Pellegrino, R
    Sutton, J
    Crombie, J
    Paul, BT
    Hoopes, PJ
    Thermal Treatment of Tissue: Energy Delivery and Assessment III, 2005, 5698 : 156 - 167
  • [10] Treatment of Renal Tumors by Percutaneous Ultrasound-Guided Radiofrequency Ablation Using a Multitined Electrode: Effectiveness and Complications
    del Cura, Jose L.
    Zabala, Rosa
    Iriarte, Jose I.
    Unda, Miguel
    EUROPEAN UROLOGY, 2010, 57 (03) : 459 - 465