A disposable homemade screen printed electrode for famotidine analysis using a computer-designed molecularly imprinted polymer based on the MWCNT-Fe3O4 nanocomposite in simulated real samples

被引:1
作者
Badameh, Anousheh [1 ]
Nezhadali, Azizollah [1 ]
机构
[1] Payame Noor Univ, Dept Chem, BOX 19395-4697, Tehran 19569, Iran
关键词
VOLTAMMETRIC DETERMINATION; SENSOR;
D O I
10.1039/d4ay01122h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A home-made screen printed electrode (SPE) was designed with a magnetic multi-walled carbon nanotube composite (MWCNT-Fe3O4) and a molecularly imprinted polymer (MIP) for sensitive and selective electrochemical analysis of famotidine (FAM). The SPE was fabricated using non-commercial conductive inks such as carbon and silver inks. The electrodes were printed by a painting technique on polyvinyl chloride (PVC) sheets as a substrate. To optimize the composition of the carbon ink, a mixture design methodology was employed. A computational approach was used to select the functional monomer. The imprinted layer was synthesized by electropolymerization of FAM (template) and pyrrole (PY) (monomer) using cyclic voltammetry (CV) on the working electrode surface modified with the MWCNT-Fe3O4 nanocomposite. Differential pulse voltammetry (DPV) was applied to characterize the template removal and rebinding processes. The Plackett-Burman design (PBD) and central composite design (CCD) investigated the effect of parameters on the MIP/MWCNT-Fe3O4/SPE performance. The sensor exhibited a linear dynamic range of 1-1000 mu M (R-2 = 0.9919) with a limit of detection (LOD) of 0.027 mu M (3s(b), n = 3). It demonstrated good repeatability and reproducibility, with relative standard deviations (RSD) of 3.6 and 4.2, respectively. This disposable and cost-effective sensor successfully detected FAM in simulated real samples and correlated well with high-performance liquid chromatography (HPLC) results.
引用
收藏
页码:7534 / 7545
页数:12
相关论文
共 51 条
  • [11] Carneiro M. C. C. G., 2018, MICROCHEM J, V138, P35
  • [12] Voltammetric determination of famotidine on a disposable pencil graphite electrode
    David, Iulia Gabriela
    Popa, Dana Elena
    Calin, Andreea-Alexandra
    Buleandra, Mihaela
    Iorgulescu, Emilia-Elena
    [J]. TURKISH JOURNAL OF CHEMISTRY, 2016, 40 (01) : 125 - 135
  • [13] Development of a new electrochemical imprinted sensor based on poly-pyrrole, sol-gel and multiwall carbon nanotubes for determination of tramadol
    Deiminiat, Behjat
    Rounaghi, Gholam Hossein
    Arbab-Zavar, Mohammad Hossein
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2017, 238 : 651 - 659
  • [14] Application of Plackett-Burman Design for Spectrochemical Determination of the Last-Resort Antibiotic, Tigecycline, in Pure Form and in Pharmaceuticals: Investigation of Thermodynamics and Kinetics
    El-Shafie, Ahmed S.
    Yousef, Aseel
    El-Azazy, Marwa
    [J]. PHARMACEUTICALS, 2022, 15 (07)
  • [15] Famotidine Use Is Associated With Improved Clinical Outcomes in Hospitalized COVID-19 Patients: A Propensity Score Matched Retrospective Cohort Study
    Freedberg, Daniel E.
    Conigliaro, Joseph
    Wang, Timothy C.
    Tracey, Kevin J.
    Callahan, Michael V.
    Abrams, Julian A.
    [J]. GASTROENTEROLOGY, 2020, 159 (03) : 1129 - +
  • [16] Electrochemical platform for the detection of adenosine using a sandwich-structured molecularly imprinted polymer-based sensor
    Gliga, Laura-Elena
    Iacob, Bogdan-Cezar
    Chesches, Bianca
    Florea, Adrian
    Barbu-Tudoran, Lucian
    Bodoki, Ede
    Oprean, Radu
    [J]. ELECTROCHIMICA ACTA, 2020, 354
  • [17] Goleij M, 2017, IRAN J CHEM CHEM ENG, V36, P129
  • [18] Graphite disk Lanthanum(III)-selective electrode based on Kryptofix-22DD
    Habibi, Nona
    Rounaghi, Gholam Hossein
    Ebrahimi, Mahmoud
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2019, 99 (14) : 1459 - 1470
  • [19] Application of Nanocrystalline Graphite-like Pyrolytic Carbon Film Electrode in the Electroanalytical Determination of Famotidine
    Hadi, Mojtaba
    Ehsani, Ali
    Honarmand, Ebrahim
    [J]. ELECTROANALYSIS, 2017, 29 (03) : 756 - 764
  • [20] Hartman Paolillo Cendon L., 2012, MINIATURIZED AGAGCL