A Clinically Feasible Electrode Array for 3D Intraoperative Electrical Impedance Tomography-Based Surgical Margin Assessment in Robot-Assisted Radical Prostatectomy

被引:3
|
作者
Kossmann, Shannon E. [1 ]
Murphy, Ethan K. [1 ]
Doussan, Allaire F. [1 ]
Lloyd, Sophie [1 ]
Halter, Ryan J. [1 ,2 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Dartmouth Coll, Geisel Sch Med, Hanover, NH 03755 USA
基金
美国国家卫生研究院;
关键词
Electrodes; Voltage measurement; Electrical impedance tomography; Arrays; Impedance; Conductivity; Biomedical measurement; Electrical impedance tomography (EIT); prostate cancer; surgical margin assessment; BREAST; SPECTROSCOPY; SYSTEM; EIT;
D O I
10.1109/TBME.2024.3408720
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: A novel small form factor circular electrode array was designed specifically for electrical impedance tomography (EIT) based assessment of surgical margins during robot assisted radical prostatectomy (RARP). Methods: The electrode array consists of 33 gold-plated electrodes arranged within a 9.5 mm diameter circular footprint on the end of a surgical probe that can be introduced through a standard 12 mm laparoscopic port used during RARP. The electrode array contains 8 larger, low-contact impedance outer electrodes dedicated for current drive and an internal grid of 25 smaller electrodes for simultaneous voltage measurement. Separating electrode geometry by function is designed to improve current delivery, speed, and resolution while reducing hardware requirements. Results: Simulations demonstrated that 1 mm diameter hemispherical prostate cancer inclusions could be localized within regions of adipose and benign prostate tissue; 1.5 mm diameter inclusions were required for localization within muscle tissue. A 2.38 mm diameter aluminum rod in 0.2 S/m saline could be localized throughout the imaging domain with a position error of less than 2.5 mm for depths from the electrode array surface of up to 1.7 mm. Ex vivo tissue experiments with a bovine model demonstrate visual congruence of muscle and adipose tissue locations between the sample and reconstructed images. Conclusion: Simulation and experimental results indicate good detection and location of inclusions. Significance: These results suggest the proposed electrode array design can provide sufficient accuracy in the detection and localization of prostate cancer against clinically relevant background tissues for use during RARP.
引用
收藏
页码:3134 / 3145
页数:12
相关论文
共 32 条
  • [31] 3D printed patient-specific prostate cancer models to guide nerve-sparing robot-assisted radical prostatectomy: a systematic review
    Coles-Black, Jasamine
    Ong, Sean
    Teh, Jiasian
    Kearns, Paul
    Ischia, Joseph
    Bolton, Damien
    Lawrentschuk, Nathan
    JOURNAL OF ROBOTIC SURGERY, 2023, 17 (01) : 1 - 10
  • [32] IDENTIFYING CAPSULAR INVOLVEMENT DURING DISSECTION OF PERIPROSTATIC TISSUE USING 3D ELASTIC AUGMENTED-REALITY DURING ROBOT-ASSISTED RADICAL PROSTATECTOMY: A FEASIBILITY STUDY
    Porpiglia, Francesco
    Checcucci, Enrico
    Amparore, Daniele
    Manfredi, Matteo
    Massa, Federica
    Tota, Daniele
    Garrou, Diletta
    Piana, Alberto
    Pecoraro, Angela
    Aimar, Roberta
    Bollito, Enrico
    Fiori, Cristian
    JOURNAL OF UROLOGY, 2019, 201 (04): : E962 - E962