A high-precision self-calibration method for laser tracker without common points

被引:0
|
作者
Qi Z. [1 ]
Zhu D. [2 ,3 ]
Luo T. [1 ,2 ]
Miao X. [1 ]
He X. [2 ]
机构
[1] Institute of Advanced Science Facilities, Shenzhen
[2] National Synchrotron Radiation Laboratory, University of Science and Technology of China, Hefei
[3] Laboratory for Ultrafast Transient Facility, Chongqing University, Chongqing
来源
Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering | 2024年 / 53卷 / 02期
基金
中国国家自然科学基金;
关键词
accuracy; laser tracker; nonlinear model; self-calibration; spherical fitting;
D O I
10.3788/IRLA20230607
中图分类号
学科分类号
摘要
Significance Because the distance measurement error of the laser tracker is much smaller than the angle measurement error, a high-precision coordinate measurement system composed of multiple laser trackers is widely used in large-scale space measurement. The system requires self-calibration before measurements, which is a process that determines the distance between centers of different laser trackers. Although the method based on common points can achieve higher accuracy, it has high requirements for the measurement environment and has a high workload. Although the method based on sphere fitting is simple to operate, its self-calibration accuracy is low. In general, it is difficult to balance measurement efficiency and measurement accuracy with current methods. Progress To overcome the shortcomings of current methods, we propose a self-calibration method based on triangular structure. Firstly, error analysis for the method based on circle fitting is conducted in two-dimensional space. For simplification, laser tracker A is assumed to measure the target ball on laser tracker B at a distance of 5 m. The angle measurement error of A will make the measurement error of points on the circle more than 30 μm through the amplification effect of length, and reduce the self-calibration accuracy further. Second, we notice that the laser tracker B also has angle observations, and propose a new method that uses angle observations from B coupled with distance observations from A, which can avoid the big error from laser tracker A. The function model based on the triangular structure is established, and the self-calibration results are obtained through iterative optimization. Compared with the method based on circle fitting, the advantages of the new method are analyzed quantitatively. Finally, our method is easily applicable to three-dimension measurement. Results We verify the superiority of this algorithm through simulated and actual measurements. In the two-dimensional simulated experiment, the true value of self-calibration is set to 5 m. In 100 repeated experiments, most of the absolute deviations based on sphere fitting were greater than 20 μm, while the absolute deviations of the proposed method were less than 10 μm. The root mean square error (RMSE) of the proposed algorithm in this article was 20.98% of RMSE for the sphere fitting method. Moreover, the number of points that the algorithm in this paper needs to measure was significantly less than the method based on spherical fitting. The measurement was implemented in the alignment laboratory of the National Synchrotron Radiation Laboratory (NSRL). Two Leica AT930 laser trackers were used and their true distance was 7 231.548 8 mm. We tested the proposed method and the method based on sphere fitting ten times. The mean absolute bias and standard deviation for the former method were only 6.21 μm and 2.44 μm respectively, while the bias and standard deviation for the latter one was 21.30 μm and 7.37 μm. The proposed method, which had better repeatability, showed the superiority of accuracy in real measurement. Conclusions and Prospects We analyze that the reason for the decrease in accuracy of the self-calibration method based on sphere center fitting is the low accuracy of the angle observation values of the aiming laser tracker. The self-calibration method based on a triangular structure is proposed. This method uses the angle on the target laser tracker to match the distance observation with high accuracy. The accuracy advantages of this algorithm are verified through quantitative analysis and two experiments. The algorithm in this paper can complete self-calibration between laser trackers without common points and improves measurement efficiency while ensuring accuracy. © 2024 Chinese Society of Astronautics. All rights reserved.
引用
收藏
相关论文
共 23 条
  • [1] Wang W, Li X, Luo T, Et al., Alignment and installation for resonant cavities of FELiChEM [J], Chinese Journal of Lasers, 49, 23, (2022)
  • [2] Luo T, Qi Z, Wang W, Et al., An improved adjustment method applied in the transformation of laser tracker [J/OL]
  • [3] Yang Z, Liu M, Zhou Y, Et al., Study on processing scheme and precision measurement method of payload adapter with irregular shape, Infrared and Laser Engineering, 48, (2019)
  • [4] Wang Z, Zhang A., Laser tracking measurement accuracy improvement method for mega gear, Infrared and Laser Engineering, 50, 11, (2021)
  • [5] Fan B, Li G, Li P, Et al., Adjustment of a laser interferometer 3D rank-defect free-network, Geomatics and Information Science of Wuhan University, 40, 2, pp. 222-226, (2015)
  • [6] Wang Z, He X, Wang W, Et al., High precision pre-alignment method based on laser tracker, Geomatics and Information Science of Wuhan University, 46, 4, pp. 555-560, (2021)
  • [7] Wu B, Xu Y, Yang F, Et al., 3D coordinate measuring system based on laser tracking absolute length measurement multilateral method, Infrared and Laser Engineering, 47, 8, (2018)
  • [8] Lin Y, Zhang G, Li Z, Et al., Self-calibration and simulation of the four-beam laser tracking interferometer system for 3D coordinate measurement, Chinese Journal of Scientific Instrument, 24, 2, pp. 205-210, (2003)
  • [9] Lin Y, Zhang G, Li Z, Et al., Optimal arrangement of four-beam laser tracking system for 3D coordinate measurement, Chinese Journal of Lasers, 29, 11, (2002)
  • [10] Hu J, Yu X, Ren X, Et al., Optimal layout of three-dimensional coordinate measurement system based on laser multi-lateration, Chinese Journal of Lasers, 41, 7, (2014)