X-ray sources for in situ wavelength calibration of x-ray imaging crystal spectrometers

被引:0
|
作者
Shah, K. [1 ]
Delgado-Aparicio, L. [1 ]
Kraus, B. F. [1 ]
Ono, M. [1 ]
Gao, L. [1 ]
Umbach, B. [1 ]
Perkins, L. [1 ]
Pablant, N. [1 ]
Hill, K. W. [1 ]
Bitter, M. [1 ]
Teall, S. [2 ]
Drake, R. [2 ]
Schmidt, G. [2 ]
机构
[1] Princeton Plasma Phys Lab, Princeton, NJ 08543 USA
[2] Proto Mfg Inc, Taylor, MI 48180 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 09期
关键词
Lattice constants - Linear accelerators - Tokamak devices - X ray detectors - X ray laboratories - X ray spectrographs - X ray spectrometers - X ray spectroscopy;
D O I
10.1063/5.0219583
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
X-ray sources for a range of wavelengths are being considered for in situ calibration of X-ray Imaging Crystal Spectrometers (XICSs) and for monitoring line shifts due to changes in the crystal temperature, which can vary during experimental operation over a day [A. Ince-Cushman et al., Rev. Sci. Instrum. 79, 10E302 (2008), L. Delgado-Aparicio et al., Plasma Phys. Control. Fusion 55, 125011 (2013)]. Such crystal temperature dependent shifts, if not accounted for, could be erroneously interpreted as Doppler shifts leading to errors in plasma flow-velocity measurements. The x-ray sources encompass characteristic x-ray lines falling within the wavelength range of 0.9-4.0 & Aring;, relevant for the XICSs on present and future fusion devices. Several technological challenges associated with the development of x-ray sources for in situ calibration are identified and are being addressed in the design of multiple x-ray tubes, which will be installed inside the spectrometer housing of the XICS for the JT-60SA tokamak. These x-ray sources will be especially useful for in situ calibration between plasma discharges. In this paper, laboratory experiments are described that were conducted with a Cu x-ray source, a heated quartz (102) crystal, and a pixelated Pilatus detector to measure the temperature dependent shifts of the Cu K alpha 1 and K alpha(2) lines at 1.5405 and 1.5443 & Aring;, respectively, and to evaluate the 2d-lattice constant for the Bragg reflecting crystal planes as a function of temperature, which, in the case of in situ wavelength calibration, would have to be used for numerical analysis of the x-ray spectra from the plasma.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Imaging x-ray crystal spectrometers for KSTAR
    Lee, SG
    Bak, JG
    Bitter, M
    Moon, MK
    Nam, UW
    Jin, KC
    Kong, KN
    Seon, KI
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03): : 1997 - 2000
  • [2] CRYSTAL WAVELENGTH CALIBRATORS FOR SYNCHROTRON X-RAY SPECTROMETERS
    HART, M
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1979, 12 (10): : 911 - 912
  • [3] X-ray imaging spectrometers for Astro-E: Ground calibration in soft X-ray range
    Hayashida, K
    Kitamoto, S
    Miyata, E
    Tsunemi, H
    Yoshita, E
    Kohmura, T
    Mori, K
    Katayama, K
    Katayama, H
    Shouho, M
    Ohta, M
    Dotani, T
    Ozaki, M
    Koyama, K
    Awaki, H
    Tsuru, TG
    Ricker, G
    Bautz, MW
    Foster, R
    Kissel, S
    X-RAY OPTICS, INSTRUMENTS, AND MISSIONS III, 2000, 4012 : 123 - 136
  • [4] In situ wavelength calibration system for the X-ray Imaging Crystal Spectrometer (XICS) on W7-X
    Kring, J.
    Pablant, N.
    Langenberg, A.
    Rice, J.
    Delgado-Aparicio, L.
    Maurer, D.
    Traverso, P.
    Bitter, M.
    Hill, K.
    Reinke, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (10):
  • [5] CURVED CRYSTAL X-RAY SPECTROMETERS
    KEMP, JW
    JONES, J
    ZEITZ, L
    ANALYTICAL CHEMISTRY, 1955, 27 (02) : 313 - 313
  • [6] Development of wavelength calibration techniques for high-resolution x-ray imaging crystal spectrometers on the EAST tokamak
    Lyu, B.
    Chen, J.
    Hu, R. J.
    Delgado-Aparicio, L. F.
    Wang, F. D.
    Bitter, M.
    Hill, K. W.
    Pablant, N.
    Lee, S. G.
    Ye, M. Y.
    Shi, Y. J.
    Wan, B. N.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (10):
  • [7] PRECISE CALIBRATION OF SEQUENTIAL X-RAY SPECTROMETERS
    SPERLING, Z
    X-RAY SPECTROMETRY, 1988, 17 (04) : 155 - 160
  • [8] Research and development of x-ray imaging crystal spectrometers for KSTAR
    Lee, SG
    Bak, JG
    Bitter, M
    Hill, K
    Nam, UW
    Kim, YJ
    Moon, MK
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3693 - 3695
  • [9] Using tritium and x-ray tubes as x-ray calibration sources for the AXAF CCD imaging spectrometer
    Jones, SE
    Bautz, M
    Kissel, S
    Pivovaroff, M
    EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY VII, 1996, 2808 : 158 - 169
  • [10] Development of x-ray microcalorimeter imaging spectrometers for the X-ray Surveyor mission concept
    Bandler, Simon R.
    Adams, Joseph S.
    Chervenak, James A.
    Datesman, Aaron M.
    Eckart, Megan E.
    Finkbeiner, Fred M.
    Kelley, Richard L.
    Kilbourne, Caroline A.
    Betancourt-Martineza, Gabriel
    Miniussi, Antoine R.
    Porter, Frederick S.
    Sadleir, John E.
    Sakai, Kazuhiro
    Smith, Stephen J.
    Stevenson, Thomas R.
    Wakeham, Nicholas A.
    Wassell, Edward J.
    Yoon, Wonsik
    Becker, Dan
    Bennett, Douglas
    Doriese, Wilham B.
    Fowler, Joseph W.
    Gard, Johnathon D.
    Hilton, Gene C.
    Mates, Benjamin
    Morgan, Kelsey M.
    Reintsema, Carl D.
    Swetz, Daniel
    Ullom, Joel N.
    Chaudhuri, Saptarshi
    Irwin, Kent D.
    Lee, Sang-Jun
    Vikhlinin, Alexey
    SPACE TELESCOPES AND INSTRUMENTATION 2016: ULTRAVIOLET TO GAMMA RAY, 2016, 9905