High-sensitive double incidence multi-pass cell for trace gas detection based on TDLAS

被引:5
|
作者
Shi, Yongpeng [1 ,2 ]
Hu, Zhen [1 ,2 ]
Niu, Mingsheng [1 ,2 ]
Li, Taohua [1 ,2 ]
Li, Hui [1 ,2 ]
Liu, Huiyuan [1 ,2 ]
Li, Xiangxian [3 ]
机构
[1] Qufu Normal Univ, Sch Phys & Phys Engn, Qufu 273165, Peoples R China
[2] Qufu Normal Univ, Laser Res Inst, Qufu 273165, Peoples R China
[3] Chinese Acad Sci, Anhui Inst Opt & Fine Mech, Hefei Inst Phys Sci, Hefei 230031, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
Gas detection system; TDLAS; Dual-incidence multi-pass cell; Laser frequency correction; Wavelength modulation spectroscopy; Atmospheric monitoring; EMISSIONS; METHANE; SENSOR; H2O; N2O;
D O I
10.1016/j.snb.2024.135829
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A high-precision gas detection system with a dual-incidence multi-pass cell (D-MPC) based on TDLAS was developed, which can simultaneously measure quadruple component gases. Two Distributed Feedback (DBF) Laser Diodes with central wavelength of 1654 nm and 2004 nm were utilized to measure methane (CH4) and carbon dioxide (CO2) for evaluating the performance of the system. High sensitivity and low frequency drift were accomplished by utilizing the second harmonic signals (WMS-2f) detection and peak-locking technology. A LabVIEW-based control and acquisition program, which could be used for direct absorption and wavelength modulation, was developed. Kalman adaptive filtering, Least Mean Squares (LMS) adaptive filtering and S-G filtering were employed for denoising in direct absorption spectroscopy (DAS), respectively, and optimal denoising technology of S-G filtering was selected. According to Allan deviation analysis, the optimal detection precision for CH4 and CO2 was 12 ppb at an integration time of 36 s and 270 ppb at an integration time of 43 s, respectively. A two-day continuous monitoring of atmospheric CO2 and CH4 was implemented to verify the system stability and can be applied to other trace gases monitoring.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] High-sensitive multi-pass imaging interferometry of a gas/plasma jet
    Sarkisov, G. S.
    McChesney, P. D.
    Stein, S.
    Pronko, S.
    Bluem, H.
    Lednum, G.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (02):
  • [2] High sensitive detection of carbon monoxide based on novel multi-pass cell
    Xia H.
    Dong F.
    Tu G.
    Wu B.
    Zhang Z.
    Wang Y.
    Guangxue Xuebao/Acta Optica Sinica, 2010, 30 (09): : 2596 - 2601
  • [3] Ultra-high sensitive photoacoustic gas detector based on differential multi-pass cell
    Zhao, Xinyu
    Chen, Ke
    Cui, Dongyu
    Guo, Min
    Li, Chenxi
    Qi, Hongchao
    Zhang, Guangyin
    Gong, Zhenfeng
    Zhou, Zhi
    Peng, Wei
    SENSORS AND ACTUATORS B-CHEMICAL, 2022, 368
  • [4] Detection of escaping NH3 based on TDLAS with multi-pass cells
    Zhang, Zeng-Fu
    Zou, De-Bao
    Chen, Wen-Liang
    Zhao, Hui-Juan
    Xu, Ke-Xin
    Zhang, Z.-F. (zfzhang@tju.edu.cn), 1600, Board of Optronics Lasers, No. 47 Yang-Liu-Qing Ying-Jian Road, Tian-Jin City, 300380, China (24): : 2367 - 2371
  • [5] A photocoustic spectroscopy system for gas detection based on the multi-pass cell
    Han, Luo
    Chen, Xinglong
    Xia, Hua
    Pang, Tao
    Zhang, Zhirong
    Wu, Bian
    Sun, Pengshuai
    Cui, Xiaojuan
    Li, Zhe
    Wang, Yu
    Dong, Fengzhong
    ADVANCED SENSOR SYSTEMS AND APPLICATIONS VII, 2016, 10025
  • [6] The design and simulation of a novel ring multi-pass optical cell for detection of environmental trace gas
    Cao, Ya-nan
    Cheng, Gang
    Tian, Xing
    Wang, Gun-Shi
    Cao, Yuan
    Sun, Chun-Yan
    Zhang, Yi-Long
    Cheng, Gao-Xin
    Yang, Hong-Tao
    OPTIK, 2021, 227
  • [7] Highly sensitive acetylene detection based on a compact multi-pass gas cell and optimized wavelength modulation technique
    Sun, Haiyue
    Ma, Yufei
    He, Ying
    Qiao, Shunda
    Yang, Xiaotao
    Tittel, Frank K.
    INFRARED PHYSICS & TECHNOLOGY, 2019, 102
  • [8] Highly sensitive CH4-TDLAS sensor based on 3D-printed multi-pass cell
    Liu, Yahui
    Sun, Xiaorong
    Sun, Haiyue
    Qiao, Shunda
    He, Ying
    Ma, Yufei
    INFRARED PHYSICS & TECHNOLOGY, 2024, 141
  • [9] Performance characteristics of a compact cylindrical multi-pass cell for laser absorption based trace gas sensor
    Kato, Yoshihiro
    Sato, Junichi
    Nakamura, Takahiro
    Endo, Masamori
    Yamaguchi, Shigeru
    Nanri, Kenzo
    Fujioka, Tomoo
    2007 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1-4, 2007, : 1370 - +
  • [10] Sensitive measurements of trace gas of formaldehyde using a mid-infrared laser spectrometer with a compact multi-pass cell
    Tanaka, Kotaro
    Miyamura, Kai
    Akishima, Kazushi
    Tonokura, Kenichi
    Konno, Mitsuru
    INFRARED PHYSICS & TECHNOLOGY, 2016, 79 : 1 - 5