Clinical application of an optimized reference measurement procedure for serum digoxin using bracketing calibration method by ID-LC-MS/MS

被引:0
|
作者
Yan, Qiaofang [1 ,2 ]
Huang, Fei [1 ,2 ]
Wang, Chuyang [1 ,2 ]
Zhan, Min [1 ]
Zhang, Qiaoxuan [1 ]
Huang, Di [1 ]
Yan, Jun [1 ]
Lin, Haibiao [1 ]
Huang, Xianzhang [1 ]
Han, Liqiao [1 ]
机构
[1] Guangzhou Univ Chinese Med, Guangdong Prov Hosp Chinese Med, Dept Lab Med, Affiliated Hosp 2, Guangzhou 510120, Peoples R China
[2] Guangzhou Univ Chinese Med, Clin Med Coll 2, Guangzhou 510405, Peoples R China
关键词
Digoxin; Standardization; Reference measurement procedure; ID-LC-MS/MS; Trueness Verification Plan; Clinical laboratory; INTOXICATION; INTERFERENCE; CANRENONE;
D O I
10.1007/s00216-024-05587-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Digoxin, a cardiac glycoside, is widely used in the treatment of cardiovascular diseases. Due to its narrow therapeutic range, precise monitoring of its blood concentration is essential. A reference measurement procedure (RMP) is pivotal for ensuring result accuracy and comparability. The RMP for serum digoxin by ID-LC-MS/MS was optimized with sample pre-treatment and detection processes, and the bracketing calibration method was used, which facilitates more accurate measurement, especially for extreme concentrations. The performance of this optimized RMP was thoroughly evaluated. The limit of detection (LoD) was 0.05 ng/mL (0.06 nmol/L) and the lowest limit of quantification (LLoQ) was 0.10 ng/mL (0.13 nmol/L). The intra- and inter-assay imprecisions were 2.24%, 2.51%, 1.40% and 1.72%, 1.65%, 0.97% at 0.5, 2.0, 5.0 ng/mL, respectively. Recoveries were 99.63 to 101.42% and the linear response ranged from 0.1 to 10.0 ng/mL. The relative bias was 0.41% and 2.00% of our results compared with the median of all participating reference laboratories for IFCC-RELA (External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine) 2023A and 2023B. The uncertainty, calibration and measurement capability (CMC) of this method were also evaluated. The optimized RMP was applied in the Trueness Verification Plan of Southern China, which indicates significant differences among clinical systems, highlighting the need for standardization efforts. In addition, two commonly used clinical systems which employed immunoassay methods were compared with this optimized RMP, and 26 individual serum samples were analyzed. The good correlations indicate the feasibility of standardization for serum digoxin. The optimized RMP serves as an accurate reference baseline for routine methods, aiming to enhance the accuracy and precision of measurements in clinical laboratories.
引用
收藏
页码:6909 / 6918
页数:10
相关论文
共 50 条
  • [21] An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of phenobarbital in human serum and plasma
    Schierscher, Tobias
    Salzmann, Linda
    Singh, Neeraj
    Bachmann, Martina
    Kobel, Anja
    Wild, Janik
    Bauland, Friederike
    Geistanger, Andrea
    Risch, Lorenz
    Geletneky, Christian
    Seger, Christoph
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (07) : 1314 - 1326
  • [22] Candidate Reference Measurement Procedure for the Quantification of Total Serum Cortisol with LC-MS/MS
    Hawley, James M.
    Owen, Laura J.
    MacKenzie, Finlay
    Mussell, Chris
    Cowen, Simon
    Keevil, Brian G.
    CLINICAL CHEMISTRY, 2016, 62 (01) : 262 - 269
  • [23] Development and validation of an isotope dilution-liquid chromatography (ID-LC-MS/MS)-based candidate reference measurement procedure for total 3,3′,5-triiodothyronine in human serum
    Li, Wenya
    Gao, Bingling
    Yang, Tao
    Tang, Liping
    Song, Dan
    Li, Hongmei
    Sun, Keqi
    Xiao, Peng
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2025,
  • [24] Quantitation of plasma metanephrines using isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS): a candidate reference measurement procedure and its application to evaluating routine ID-LC/MS/MS methods
    Liu, Zhenni
    Liu, Qingxiang
    Deng, Yuhang
    Zhao, Haijian
    Zeng, Jie
    Zhang, Tianjiao
    Zhang, Jiangtao
    Wang, Jing
    Zhou, Weiyan
    Zhang, Chuanbao
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (30) : 7509 - 7520
  • [25] Quantitation of plasma metanephrines using isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS): a candidate reference measurement procedure and its application to evaluating routine ID-LC/MS/MS methods
    Zhenni Liu
    Qingxiang Liu
    Yuhang Deng
    Haijian Zhao
    Jie Zeng
    Tianjiao Zhang
    Jiangtao Zhang
    Jing Wang
    Weiyan Zhou
    Chuanbao Zhang
    Analytical and Bioanalytical Chemistry, 2021, 413 : 7509 - 7520
  • [26] Simultaneous determination of ergothioneine, selenoneine, and their methylated metabolites in human blood using ID-LC-MS/MS
    Achouba, Adel
    Dumas, Pierre
    Ayotte, Pierre
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2023, 415 (29-30) : 7259 - 7267
  • [27] Isotope-dilution-LC-MS/MS candidate reference measurement procedure for cefepime in human serum
    Schaeffler, Judith
    Vogeser, Michael
    Habler, Katharina
    JOURNAL OF MASS SPECTROMETRY AND ADVANCES IN THE CLINICAL LAB, 2024, 33 : 43 - 48
  • [28] Measurement Procedure for Glucose in Serum Using Liquid Chromatography Isotope Dilution - Tandem Mass Spectrometry (LC-ID-MS/MS)
    Kress, M.
    Meissner, D.
    Kramer, U.
    Spannagl, M.
    Kaiser, P.
    CLINICAL LABORATORY, 2014, 60 (09) : 1509 - 1515
  • [29] Comparative evaluation of peptide vs. protein-based calibration for quantification of cardiac troponin I using ID-LC-MS/MS
    Asicioglu, Meltem
    Swart, Claudia
    Saban, Evren
    Yurek, Emrah
    Karaguler, Nevin Gul
    Oztug, Merve
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025,
  • [30] Quantification of 25OHD in serum by ID-LC-MS/MS based on oriented immobilization of antibody on magnetic materials
    Keke Yi
    Jie Xie
    Ziyu Qu
    Yanling Lin
    Zejian Huang
    Tao Peng
    Yang Zhao
    Rui Zhai
    Xiaoyun Gong
    You Jiang
    Xinhua Dai
    Xiang Fang
    Microchimica Acta, 2023, 190