Detection of Ammonia and Deuterated Hydrocarbons in Exhaust Gas by Infrared Absorption Spectroscopy during Wall Conditioning

被引:0
|
作者
Tanaka M. [1 ,2 ]
Kato H. [1 ]
Suzuki N. [1 ]
Chimura H. [1 ]
机构
[1] National Institute for Fusion Science, National Institutes of Natural Sciences, 322-6 Oroshi-cho, Toki, 509-5292, Gifu
[2] The Graduate University for Advanced Studies, SOKENDAI, 322-6 Oroshi-cho, Toki, 509-5292, Gifu
基金
日本学术振兴会;
关键词
ammonia; deuterated hydrocarbons; exhaust gas monitoring; FTIR; infrared absorption spectroscopy; large fusion test device;
D O I
10.1585/PFR.16.2405062
中图分类号
学科分类号
摘要
To detect ammonia and deuterated hydrocarbons in exhaust gas from the Large Helical Device (LHD), infrared absorption spectrometry, FTIR with a long optical path gas cell, was applied. Ammonia (NH3) and deuterated hydrocarbons (CxHyDz) could be detected during the first operations of wall baking at 368 K and the D2 glow discharge conducted after vacuum vessel closure. The concentration of ammonia increased with increasing baking temperature, and deuterated ammonia was not detected. Thus, the ammonia, which likely originated from sweat of workers produced during vacuum vessel maintenance activities, was released from the vacuum vessel wall. Hydrocarbons were likely produced by chemical sputtering of carbon tiles and were deuterated by a hydrogen isotope exchange reaction due to D2 glow discharge, while H2O was released from the vacuum vessel during wall baking. It was confirmed that ammonia and various types of deuterated hydrocarbons could be measured discriminately by an FTIR spectroscopy system using a long optical path gas cell. © 2021 The Japan Society of Plasma Science and Nuclear Fusion Research
引用
收藏
页码:2405062 / 1
相关论文
共 50 条
  • [1] Detection of Ammonia and Deuterated Hydrocarbons in Exhaust Gas by Infrared Absorption Spectroscopy duringWall Conditioning
    Tanaka, Masahiro
    Kato, Hiromi
    Suzuki, Naoyuki
    Chimura, Hiroki
    PLASMA AND FUSION RESEARCH, 2021, 16
  • [2] Analysis of Hydrocarbons in the Exhaust Gas of a Fusion Test Device Using Infrared Absorption Spectroscopy
    Tanaka, Masahiro
    Kato, Hiromi
    Suzuki, Naoyuki
    PLASMA AND FUSION RESEARCH, 2020, 15
  • [3] Automotive exhaust gas analysis by infrared tunable diode laser absorption spectroscopy
    Louvet, Y
    deBorniol, E
    Henry, D
    Mohamed, A
    Rosier, B
    Jones, W
    COMBUSTION DIAGNOSTICS, 1997, 3108 : 54 - 62
  • [4] Gas concentration detection method based on infrared absorption spectroscopy technology
    Liu, Yong-Ping
    Wang, Xia
    Li, Shuai-Shuai
    Hu, Ling
    Guangzi Xuebao/Acta Photonica Sinica, 2015, 44 (01):
  • [5] EXHAUST GAS ANALYSIS BY GAS CHROMATOGRAPHY COMBINED WITH INFRARED DETECTION
    HEATON, WB
    WENTWORTH, JT
    ANALYTICAL CHEMISTRY, 1959, 31 (03) : 349 - 357
  • [6] ULTRALOW GAS CONCENTRATION INFRARED ABSORPTION SPECTROSCOPY
    KREUZER, LB
    JOURNAL OF APPLIED PHYSICS, 1971, 42 (07) : 2934 - &
  • [7] Fingerprints of polycyclic aromatic hydrocarbons (PAHs) in infrared absorption spectroscopy
    Tommasini, Matteo
    Lucotti, Andrea
    Alfe, Michela
    Ciajolo, Anna
    Zerbi, Giuseppe
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2016, 152 : 134 - 148
  • [8] Detection of ethene and other hydrocarbons in gas turbine engine exhaust using non-intrusive FTIR spectroscopy
    Arrigone, GM
    Welch, MA
    Hilton, M
    Miller, MN
    Wilson, CW
    REMOTE SENSING OF CLOUDS AND THE ATMOSPHERE VII, 2003, 4882 : 384 - 393
  • [9] Detection of new hydrocarbons in Uranus' atmosphere by infrared spectroscopy
    Burgdorf, Martin
    Orton, Glenn
    van Cleve, Jeffrey
    Meadows, Victoria
    Houck, James
    ICARUS, 2006, 184 (02) : 634 - 637
  • [10] Automotive exhaust gas flow control for an ammonia water absorption refrigeration system
    Rego, A. T.
    Hanriot, S. M.
    Oliveira, A. F.
    Brito, P.
    Rego, T. F. U.
    APPLIED THERMAL ENGINEERING, 2014, 64 (1-2) : 101 - 107