Comparative study on the use of different metal electrodes in low-pressure glow discharge plasma sterilization

被引:0
作者
Daseco, Joanna Abigael [1 ]
Pabeliña, Karel G. [1 ]
Siringan, Ma. Auxilia T. [2 ]
Ramos, Henry J. [1 ]
机构
[1] Plasma Physics Laboratory, National Institute of Physics, University of the Philippines Diliman, Quezon City
[2] Microbiological Research and Services Laboratory, Natural Science and Research Institute, University of the Philippines Diliman, Quezon City
关键词
Bacillus subtilis; Electrodes; Hydrogen peroxide plasma; Plasma sterilization;
D O I
10.1615/PlasmaMed.2014011720
中图分类号
学科分类号
摘要
The sterilizing efficacy of low-pressure direct current glow discharge hydrogen peroxide (H2O2) plasma generated by a planar parallel plate plasma source using the plasma enhanced chemical vapor deposition (PECVD) facility was tested. This study compares the effect of using different metals (copper, stainless steel, and aluminum) as electrodes in plasma sterilization of stainless steel dishes. Test samples were exposed to H2O2 plasma under different sets of discharge currents and exposure times. Bacillus subtilis was used as the test organism and microbial analysis was made by means of the standard plate count method of serial dilution and pour plating. Evaluation of microbial death was done using survival curves, percent reduction, and decimal reduction value. Results showed that sterilization using copper electrodes exhibited the highest decimal reduction value (D-value) and percent reduction among the three electrodes. However, statistical analysis using multivariate analysis of variance (MANOVA) at 0.05 significance level assessed that the type of electrode material is not a significant factor in the H2O2 plasma sterilization of Bacillus subtilis cells. © 2014 by Begell House, Inc.
引用
收藏
页码:1 / 10
页数:9
相关论文
共 19 条
[1]  
Moisan M., Barbeau J., Crevier M., Pelletier J., Philip N., Saoudi B., Plasma sterilization: Methods and mechanisms, Pure Appl Chem., 74, 3, pp. 349-358, (2002)
[2]  
Yardimci O., Setlow P., Plasma sterilization: Opportunities and microbial assessment strategies in medical device manufacturing, IEEE Trans Plasma Sci., 38, 4, pp. 973-981, (2010)
[3]  
Soloshenko I.A., Tsiolko V.V., Khomich V.A., Bazhenov V.Y., Ryabtsev A.V., Schedrin I.A., Mikhno I.L., Features of sterilization using low-pressure DC-discharge hydrogen peroxide plasma, IEEE Trans Plasma Sci., 30, 4, pp. 1440-1444, (2002)
[4]  
Moman R.M., Najmadeen H., The bactericidal efficacy of cold atmospheric plasma technology on some bacterial strains, Egypt Acad J Biol Sci., 2, 2, pp. 43-47, (2010)
[5]  
Schutze A., Jeong J., Babayan S., Park J., Selwyn S., Hicks R., The atmospheric-pressure plasma jet: A review and comparison to other plasma sources, IEEE Trans Plasma Sci., 26, 6, pp. 1685-1694, (1998)
[6]  
Moreira A.J., Et al., Sterilization of oxygen plasma, Appl. Surf. Sci., 235, pp. 151-155, (2004)
[7]  
Soloshenko I.A., Tsiolko V.V., Khomich V.A., Bazhenov V.Y., Ryabtsev A.V., Schedrin I.A., Mikhno I.L., Sterilization of medical products in low-pressure glow discharges, Plasma Phys. Rep., 26, 9, pp. 792-800, (2000)
[8]  
Moreau S., Et al., Using the flowing afterglow of a plasma to inactivate Bacillus subtilis spores: Influence of the operating conditions, J Appl Phys., 88, 2, pp. 1166-1174, (2000)
[9]  
Hayashi N., Tsutsui S., Yokoohji T., Satoh S., Sterilization characteristics for medical equipments using low-pressure oxygen/water plasma, Proceedings of XXVIIth ICPIG, (2005)
[10]  
Laroussi M., Mendis D.A., Rosenberg M., Plasma interaction with microbes, New J Phys, 5, pp. 41.1-41.10, (2003)