Heart rate sensing during exercise by means of photoplethysmography

被引:0
|
作者
Shimazaki T. [1 ]
Hara S. [1 ]
Okuhata H. [2 ]
Nakamura H. [3 ]
Kawabata T. [4 ]
机构
[1] Graduate School of Engineering, Osaka City University
[2] Healthcare Corporation, Aihara Second Hospital
[3] Graduate School of Health and Well-being, Kansai University
关键词
Adaptive filter; Canceler; Heart rate; Motion artifact; Photoplethysmography;
D O I
10.11239/JSMBE.54.225
中图分类号
学科分类号
摘要
Monitoring heart rate (HR) during exercise can be used not only for early detection of a change in physical condition but also as an indication of the exercise training condition. Photoplethysmography (PPG) is a simple andnon-invasive methodfor HR sensing, but its use during exercise is difficult, since the PPG output is contaminatedwith motion artifact (MA). Especially for vigorous exercise, MA has the same frequency component as that of bloodvolume pulses (BVPs) in the PPG output, andno linear MA cancellation technique such as bandpass filtering (BPF) can work effectively. This paper proposes a PPG-basedMA canceling HR sensor. It is equipped with two sensors; one is a normal PPG sensor with a light emitting diode (LED)photo detector (PD) that contacts the skin to detect BVPs andMA, andthe other is an MA sensor with a LEDPD that does not contact the skin to detect only MA. Applying the outputs from the two sensors to an adaptive filter, the BVPs can be obtainedeasily. We evaluatedthe performance of the proposedHR sensor in ten subjects who performeda series of resting, walking, running andjumping. The results revealedthat the proposedHR sensor, together with an outlier rejection method, achieved root mean square errors (RMSEs) of 7.1 bpm, 6.4 bpm and6.1 bpm for walking, running andjumping, respectively, when using a Holter monitor as the reference. These results imply a reduction of RMSE by a maximum of 83% comparedto no MA cancellation. © 2016, Japan Soc. of Med. Electronics and Biol. Engineering. All rights reserved.
引用
收藏
页码:225 / 235
页数:10
相关论文
共 50 条
  • [41] Heart rate variability reproducibility during exercise
    McNarry, Melitta A.
    Lewis, Michael J.
    PHYSIOLOGICAL MEASUREMENT, 2012, 33 (07) : 1123 - 1133
  • [42] Changes of heart rate variability during exercise
    Mongin, D.
    Chabert, C.
    Gomez Extremera, M.
    Hue, O.
    Courvoisier, D. S.
    Carpena, P.
    Bernaola Galvan, P. A.
    2020 11TH CONFERENCE OF THE EUROPEAN STUDY GROUP ON CARDIOVASCULAR OSCILLATIONS (ESGCO): COMPUTATION AND MODELLING IN PHYSIOLOGY NEW CHALLENGES AND OPPORTUNITIES, 2020,
  • [43] Photoplethysmography and Heart Rate Variability for the Diagnosis of Preeclampsia
    Euliano, Tammy Y.
    Michalopoulos, Kostas
    Singh, Savyasachi
    Gregg, Anthony R.
    Del Rio, Mariem
    Vasilopoulos, Terrie
    Johnson, Amber M.
    Onkala, Allison
    Darmanjian, Shalom
    Euliano, Neil R.
    Ho, Monique
    ANESTHESIA AND ANALGESIA, 2018, 126 (03): : 913 - 919
  • [44] Remote photoplethysmography for heart rate measurement: A review
    Xiao, Hanguang
    Liu, Tianqi
    Sun, Yisha
    Li, Yulin
    Zhao, Shiyi
    Avolio, Alberto
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2024, 88
  • [45] Heart Rate Turbulence Analysis Based on Photoplethysmography
    Gil, Eduardo
    Laguna, Pablo
    Pablo Martinez, Juan
    Barquero-Perez, Oscar
    Garcia-Alberola, Arcadi
    Sornmo, Leif
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (11) : 3149 - 3155
  • [46] Kinetics of catecholamines and potassium, and heart rate during exercise testing in obese subjects - Heart rate regulation in obesity during exercise
    Salvadori, A
    Fanari, P
    Giacomotti, E
    Palmulli, P
    Bolla, G
    Tovaglieri, I
    Luzi, L
    Longhini, E
    EUROPEAN JOURNAL OF NUTRITION, 2003, 42 (04) : 181 - 187
  • [47] Heart rate response during cardiopulmonary exercise in the denervated heart
    Ono, Ryohei
    Iwahana, Togo
    Kato, Hirotoshi
    Kobayashi, Yoshio
    CLINICAL CASE REPORTS, 2023, 11 (01):
  • [48] CONTROL OF HEART-RATE IN THE TRANSPLANTED HEART DURING EXERCISE
    SCHELL, J
    MEYER, M
    MARCONI, C
    RIEU, M
    CERRETELLI, P
    CABROL, C
    ATEMWEGS-UND LUNGENKRANKHEITEN, 1990, 16 (09) : 437 - 440
  • [49] Towards Photoplethysmography-Based Estimation of Instantaneous Heart Rate During Physical Activity
    Jarchi, Delaram
    Casson, Alexander J.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2017, 64 (09) : 2042 - 2053
  • [50] The relationship between heart rate and oxygen consumption in heart transplant recipients during a cardiopulmonary exercise test Heart rate dynamic during exercise test
    Carvalho, Vitor Oliveira
    Bocchi, Edimar Alcides
    Pascoalino, Lucas Nobilo
    Guimaraes, Guilherme Veiga
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 145 (01) : 158 - 160