Feasibility of noninvasive measurement of deep brain temperature in newborn infants by multifrequency microwave radiometry

被引:68
作者
机构
[1] Research Institute of Electronics, Shizuoka University, Hamamatsu
关键词
Biomédical monitoring; Brain cooling; hypothermia; Microwave radiometry; Neural rescue; Noninvasive temperature measurement; Pediatrics;
D O I
10.1109/22.884206
中图分类号
学科分类号
摘要
Clinical studies of hypotherrnal neural rescue therapy for newborn infants who have suffered hypoxia-ischaemia are currently hindered by the difficulty in measuring deep brain temperature. This paper addresses: the specific requirements for this measurement problem, the design of a proposed radiometer system, a method for retrieving the temperature profile within the cooled head, and an estimation of the precision of the measurement of deep brain temperature using the technique. A five-frequency-band radiometer with a contact-type antenna operating within the range 1-4 GHz is proposed to obtain brightness temperatures corresponding to temperature profiles predicted by a realistic thermal model of the cooled baby head. The problems of retrieving the temperature profile from this set of brightness temperatures, and the estimation of its precision, are solved using a combination of model fitting and Monte Carlo techniques. The results of this paper show that the proposed technique is feasible, that it is expected to provide a good estimate of the temperature profile within the cooled baby-head, and that the estimated precision (2cr) of the temperature measured in the deep brain structures is better than 0.8 K, depending upon the estimation procedure used. © 2000 IEEE.
引用
收藏
页码:2141 / 2147
页数:6