The design principle of magnetic probe for current measurement through plasmas is introduced. A micro magne-tic probe device is manufactured to diagnose the load current in Z-pinch experiments on pulsed power facilities, according to the actual structure of wire-array load and the experimental layout. The sensitivity of the probe is calibrated through a load simulator that has the same size as the load used in experiments. Experimental results show that, when the pulsed power facility works at 1.2 MA peak current with 60 ns rise-time, the load current measured by this probe has 12% difference in magnitude and 5 ns difference in peak current time from that measured by accelerator current monitor.
引用
收藏
页码:2055 / 2058
页数:3
相关论文
共 7 条
[1]  
Bland S.N., Ampleford D.J., Bott S.C., Et al., Use of Faraday probing to estimate current distribution in wire array Z pinches, Rev Sci Instrum, 77, (2006)
[2]  
Kock L., Bender S., Costley A., Et al., Magnetic diagnostics for ITER, Diagnostics for Experimental Thermonuclear Fusion Reactors, 2, pp. 307-318, (1998)
[3]  
Xi W., Wu S., Shen B., Et al., Design of magnetic probe coils in the EAST Tokamak, Nuclear Fusion and Plasma Physics, 28, 1, pp. 73-76, (2008)
[4]  
Huddlestone R.H., Leonard S.L., Plasma Diagnostic Techniques, (1965)
[5]  
Lochte-Holtgreven W., Plasma Diagnostics, (1968)
[6]  
Xiang Z., Yu C., High Temperature Plasma Diagnostic Techniques: The First Part, (1982)
[7]  
Volkov G.S., Grabovskii E.V., Zaitsev V.I., Et al., Plasma diagnostics at the Angala-5-1 facility, Instruments and Experimental Techniques, 47, 2, pp. 201-208, (2004)