摘要
Introduction: A ventricular assist device (VAD) is an electromechanical pump used to treat heart failures. For designing the physiological control system for a VAD, one needs a mathematical model and its related parameters. This paper presents a characterization procedure for determining the model parameter values of the electrical, mechanical, and hydraulic subsystems of a pediatric Rotary Blood Pump (pRBP). Methods: An in vitro test setup consisting of a pRBP prototype, a motor driver module, an acrylic reservoir, mechanical resistance and tubings, pressure and fluid flow sensors, and data acquisition, processing, and visualization system. The proposed procedure requires a set of experimental tests, and a parameter estimation algorithm for determining the model parameters values. Results: The operating limits of the pRBP were identified from the steady-state data. The relationship between the pressure head, flow rate, and the rotational speed of the pRBP was found from the static tests. For the electrical and mechanical subsystems, the dc motor model has a viscous friction coefficient that varies nonlinearly with the flow. For the hydraulic subsystem, the pressure head is assumed to be a sum of terms related to the resistance, the inertance, the friction coefficient, and the pump speed. Conclusion: The proposed methodology was successfully applied to the characterization of the pRBP. The combined use of static and dynamic tests provided a precise lumped parameter model for representing the pRBP dynamics. The agreement, regarding mean squared deviation, between experimental and simulated results demonstrates the correctness and feasibility of the characterization procedure. © 2018, Brazilian Society of Biomedical Engineering. All rights reserved.
引用
收藏
页码:299 / 309
页数:10
相关论文
共 50 条
  • [41] Estimation of Maximum Ventricular Elastance Under Assistance With a Rotary Blood Pump
    Sugai, Telma K.
    Tanaka, Akira
    Yoshizawa, Makoto
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Nitta, Shin-ichi
    Baba, Atsushi
    ARTIFICIAL ORGANS, 2010, 34 (05) : 442 - 446
  • [42] Lumped Parameter Model of the Systemic Circulation and a Rotary Blood Pump Interaction
    Rubtsova, Ekaterina
    Malinina, Alena
    Galyastov, Andrey
    PROCEEDINGS OF THE 2021 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (ELCONRUS), 2021, : 2855 - 2859
  • [43] Modeling and control of an implantable rotary blood pump for heart failure patients
    AlOmari, Abdul-Hakeem H.
    Savkin, Andrey V.
    Ayre, Peter J.
    Lim, Einly
    Lovell, Nigel H.
    2010 AMERICAN CONTROL CONFERENCE, 2010, : 3574 - 3579
  • [44] Baroreflex Sensitivity of an Arterial Wall During Rotary Blood Pump Assistance
    Yambe, Tomoyuki
    Imachi, Kou
    Shiraishi, Yasuyuki
    Yamaguchi, Tasuku
    Shibata, Mune-ichi
    Kameyama, Takeshi
    Yoshizawa, Makoto
    Sugita, Norihiro
    ARTIFICIAL ORGANS, 2009, 33 (09) : 767 - 769
  • [45] Estimating Flow Rate Using the Motor Torque in a Rotary Blood Pump
    Hijikata, Wataru
    Rao, Jun
    Abe, Shodai
    Takatani, Setsuo
    Shinshi, Tadahiko
    SENSORS AND MATERIALS, 2015, 27 (04) : 297 - 308
  • [46] Cannula Tip With Integrated Volume Sensor for Rotary Blood Pump Control: Early-Stage Development
    Cysyk, Joshua
    Newswanger, Ray
    Popjes, Eric
    Pae, Walter
    Jhun, Choon-Sik
    Izer, Jenelle
    Weiss, William
    Rosenberg, Gerson
    ASAIO JOURNAL, 2019, 65 (04) : 318 - 323
  • [47] Investigation of Blood Hemolysis Study in Rotary Blood Pump between Continuous Flow and Pulsatile Flow
    Rungsirikunnan, Chitsanupong
    Mondee, Kamonwan
    Gesprasert, Goragoch
    Naiyanetr, Phornphop
    2018 11TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2018), 2018,
  • [48] Development of a novel shrouded impeller pediatric blood pump
    Khan, Talha Irfan
    Zad, Haris Sheh
    Lazoglu, Ismail
    Yalcin, Ozlem
    JOURNAL OF ARTIFICIAL ORGANS, 2018, 21 (02) : 142 - 149
  • [49] Evaluation of hydraulic radial forces on the impeller by the volute in a centrifugal rotary blood pump
    Boehning F.
    Timms D.L.
    Amaral F.
    Oliveira L.
    Graefe R.
    Hsu P.-L.
    Schmitz-Rode T.
    Steinseifer U.
    Artificial Organs, 2011, 35 (08) : 818 - 825
  • [50] Computational Parametric Study of the Axial and Radial Clearances in a Centrifugal Rotary Blood Pump
    Rezaienia, Mohammad Amin
    Paul, Gordon
    Avital, Eldad
    Rothman, Martin
    Korakianitis, Theodosios
    ASAIO JOURNAL, 2018, 64 (05) : 643 - 650