Effects of intermittent overloads on fatigue of PMMA bone cement

被引:0
作者
Evans S.L. [1 ]
机构
[1] School of Engineering, Cardiff University, Cardiff CF24 3AA, Parade
关键词
Bone cement; Crazing; Fatigue; Fatigue crack growth; Fatigue crack propagation; Overloads; Plasticity induced closure; PMMA; Polymethylmethacrylate; Retardation; Variable amplitude loading;
D O I
10.1504/IJNBM.2010.036108
中图分类号
学科分类号
摘要
PMMA bone cements are widely used in orthopaedics, but questions remain as to their long term durability. Although fatigue of bone cements has been widely studied, the effects of variable amplitude loading are not fully understood. This paper presents a novel method using a crossover experimental design, in which a crack is propagated under δK control and a series of different overload patterns are applied. This method allowed accurate and repeatable measurements of the effects of overloads on crack growth. It was found that single overloads every ten to 100 cycles caused a pronounced increase in the crack growth rate compared to Paris Law predictions; there was no evidence of retardation as would be expected in most materials. Overloads may fracture intermittently propagating crazes and help the crack past micro-structural obstacles, allowing disproportionately faster crack growth. The observed increase in crack growth rate over a wide range of different overload intervals and for single and block overloads could be accounted for by an eight-fold increase in crack growth rate for approximately fifty cycles after each overload. Copyright © 2010 Inderscience Enterprises Ltd.
引用
收藏
页码:65 / 77
页数:12
相关论文
共 36 条
[21]  
Jasty M., Maloney W.J., Bragdon C.R., O'Connor D.O., Haire T., Harris W.H., The initiation of failure in cemented femoral components of hip arthroplasties, J Bone Joint Surg, 73, 4, pp. 551-558, (1991)
[22]  
Lewis G., Fatigue testing and performance of acrylic bone-cement materials: State-of-the-art review, J BiomedMat Res, 66 B, 1, pp. 457-486, (2003)
[23]  
Lewis G., Janna S., Carroll M., Effect of test frequency on the in vitro fatigue life of acrylic bone cement, Biomaterials, 24, pp. 1111-1117, (2003)
[24]  
Ling R.S.M., Lee A.J.C., Porosity reduction in acrylic bone cement is clinically irrelevant, Clinical Orthopaedics and Related Research, 355, pp. 249-253, (1998)
[25]  
Malchau H., Herberts P., Ahnfelt L.E., Johnell O., Prognosis of total hip replacement, results from a national register of revised failures 1979-1990 in Sweden- a 10 year follow-up of 92,675 total hip replacements', Scientific Exhibition presented at the, 61st American Academy of Orthopaedic Surgeons, (1994)
[26]  
Manero J.M., Ginebra M.P., Gil F.J., Planell J.A., Delgado J.A., Morejon L., Artola A., Gurruchage M., Goni I., Propagation of fatigue cracks in acrylic bone cements containing different radiopaque agents, Proc Instn Mech Engrs Part H Journal of Engineering in Medicine, 218, pp. 167-172, (2004)
[27]  
Nguyen N.C., Maloney W.J., Dauskardt R.H., Reliability of PMMA bone cement fixation: Fracture and fatigue crack-growth behaviour, J Mat Sci Materials in Medicine, 8, 8, pp. 473-483, (1997)
[28]  
Nicholas M., Development of experimental rubber toughened bone cements, PhD Thesis, (2007)
[29]  
Orr J.F., Dunne N.J., Quinn J.C., Shrinkage stresses in bone cement, Biomaterials, 24, 17, pp. 2933-2940, (2003)
[30]  
Piazza E., Baleani M., Cristofolini L., Toni A., Effect of barium sulfate and sodium fluoride on the fatigue strength and on the fatigue crack propagation rate of PMMA bone cement, 12th Conference of the European Society of Biomaterials, (2000)