Non-enzymatic continuous glucose monitoring system

被引:0
作者
Moore S. [1 ]
You B.H. [1 ]
Kim N. [2 ]
Park T. [3 ]
Song I.-H. [1 ]
机构
[1] Department of Engineering Technology, Texas State University, San Marcos
[2] Ingram School of Engineering, Texas State University, San Marcos
[3] School of Mechanical Engineering, Kyungnam University, Gyeongsangnam-do, Changwon-si
关键词
All Open Access; Bronze;
D O I
10.1049/MNL.2017.0894
中图分类号
学科分类号
摘要
A prominent field of study in biosensor development is the study of electric properties which provide valuable insight into the fluids contents. In this research, a non-enzymatic glucose sensor is fabricated, characterised, and employed. The developed device determines the effects of electrode surface area on sensing efficacy, the effects of glucose concentration on impedimetric response, and a real-time measurement of glucose concentration. Deviations throughout the entire glucose range are detected as an inverse of the impedance in the cell due to the inverse relationship of glucose concentration and charge transfer resistance. The continuous monitoring of glucose is demonstrated by a rapid device response over two iterations of glucose concentration in ascending order: 50 µM, 400 µM, and 3.2 mM. At a sustained frequency of 10 kHz, the result shows a stable impedimetric response of 1038, 752, and 688 kΩ, respectively. The validity of the device as a continuous glucose monitoring method is carried out by repeating the cycle and observing the response. © The Institution of Engineering and Technology 2018.
引用
收藏
页码:1079 / 1084
页数:5
相关论文
共 24 条
[1]  
Lafleur J.P., Jonsson A., Senkbeil S., Et A., Recent advances in lab-on-a-chip for biosensing applications, Biosens. Bioelectron., 76, pp. 213-233, (2016)
[2]  
Chaubey A., Malhotra B.D., Mediated biosensors, Biosens. Bioelectron., 17, pp. 441-456, (2002)
[3]  
Clark L.C.J., Lyons C., Electrode systems for continuous monitoring in cardiovascular surgery, Ann. New York Acad. Sci., 102, pp. 29-45, (1962)
[4]  
Shervedani R.K., Mehrjardi A.H., Zamiri N., A novel method for glucose determination based on electrochemical impedance spectroscopy using glucose oxidase self-assembled biosensor, Bioelectrochemistry, 69, 2, pp. 201-208, (2006)
[5]  
Park S., Boo H., Chung T.D., Electrochemical nonenzymatic glucose sensors, Anal. Chim. Acta, 556, 1, pp. 46-57, (2006)
[6]  
Lee S.J., Yoon H.S., Xuan X., Et A., A patch type non-enzymatic biosensor based on 3d sus micro-needle electrode array for minimally invasive continuous glucose monitoring, Sens. Actuators B, Chem., 222, pp. 1144-1151, (2016)
[7]  
Chen C., Zhao X.L., Li Z.H., Current and emerging technology for continuous glucose monitoring, Sensors (Basel), 17, (2017)
[8]  
Pletcher D., Electrocatalysis: Present and future, J. Appl. Electrochem., 14, pp. 403-415, (1984)
[9]  
Dong W., Luo J., He H., Et al., A reinforced composite structure composed of polydiacetylene assemblies deposited on polystyrene microspheres and its application to H5n1 virus detection, Int. J. Nanomed., 8, pp. 221-232, (2013)
[10]  
Lee Y.-J., Park D.-J., Park J.-Y., Et al., Fabrication and optimization of a nanoporous platinum electrode and a non-enzymatic glucose micro-sensor on silicon, Sensors, 8, pp. 6154-6164, (2008)