Lumped element electrical model based on three resistors for electrical impedance in radiofrequency cardiac ablation: Estimations from analytical calculations and clinical data

被引:14
作者
Berjano, Enrique [1 ]
d'Avila, Andre [2 ]
机构
[1] Biomedical Synergy, Electronic Engineering Department, Universitat Politècnica de València
[2] Helmsley Cardiac Electrophysiology Center, Mt Sinai Medical Center and School of Medicine, New York, NY
关键词
Cardiac ablation; Electrical impedance; Lumped element model; Percutaneous ablation; Radiofrequency ablation; Theoretical model; Three-resistor model;
D O I
10.2174/1874120720130603001
中图分类号
学科分类号
摘要
The electrical impedance measured during radiofrequency cardiac ablation (RFCA) is widely used in clinical studies to predict the heating evolution and hence the success of the procedure. We hypothesized that a model based on three resistors in series can mimic the total electrical impedance measured during RFCA. The three resistors or impedances are given by: impedance associated with the tissue around the active electrode (myocardium and circulating blood) (Z-A), that associated with the tissue around the dispersive electrode (Z-DE) and that associated with the rest of the body (Z-B). Our objective was to quantify the values associated with these three impedance types by an analytical method, after which the values obtained would be compared to those estimated from clinical data from previous studies. The results suggest that an RFCA using a 7 Fr 4-mm electrode would give a Z-A of around 75 ohms, a Z-DE around 20 ohms, and ZB would be 15±10 ohms (for body surface area variations between 1.5 and 2.5 m^2). Finally, adaptations of the proposed model were used to explain the results of previous clinical studies using a different electrode arrangement, such as in bipolar ablation of the ventricular septum. © Berjano and d'Avila.
引用
收藏
页码:62 / 70
页数:8
相关论文
共 25 条
[1]  
Nath S., Haines D.E., Biophysics and pathology of catheter energy delivery systems, Prog. Cardiovasc. Dis, 37, pp. 185-204, (1995)
[2]  
Berjano E.J., Theoretical modeling for radiofrequency ablation: State-of-the-art and challenges for the future, Biomed. Eng. Online, 5, (2006)
[3]  
Wittkampf F.H., Nakagawa H., RF catheter ablation: Lessons on lesions, Pacing Clin. Electrophysiol., 29, pp. 1285-1297, (2006)
[4]  
Neufeld G.R., Principles and hazards of electrosurgery including laparoscopy, Surg. Gynecol. Obstet, 147, pp. 705-710, (1978)
[5]  
Ragheb T., Riegle S., Geddes L.A., Amin V., The impedance of a spherical monopolar electrode, Ann. Biomed. Eng, 20, pp. 617-627, (1992)
[6]  
Panescu D., Whayne J.G., Fleischman S.D., Mirotznik M.S., Swanson D.K., Webster J.G., Three-dimensional finite element analysis of current density and temperature distributions during radio-frequency ablation, IEEE Trans. Biomed. Eng, 42, pp. 879-890, (1995)
[7]  
Foster K.R., Schwan H.P., Dielectric properties of tissues and biological materials: A critical review, Crit. Rev. Biomed. Eng, 17, pp. 25-104, (1989)
[8]  
Pearce J.A., Electrosurgery, (1986)
[9]  
Yamamoto T., Yamamoto Y., Electrical properties of the epidermal stratum corneum, Med. Biol. Eng., 14, pp. 151-158, (1976)
[10]  
Miklavcic D., Pavselj N., Hart F.X., Electric Properties of Tissues, Wiley Encyclopedia of Biomedical Engineering, pp. 1-14, (2006)