Comparison of Pulse Sequences of Magnetic Resonance Imaging for Optimization of Timing and Image Quality

被引:0
作者
Amin N. [1 ]
Yousaf M. [2 ]
Javid M.A. [3 ]
Farooq A. [1 ]
机构
[1] Department of Physics, COMSATS University Islamabad, Lahore Campus
[2] Mathematics Department, COMSATS University Islamabad, Lahore Campus
[3] Department of Basic Sciences, UET, Taxila
关键词
Diagnosis; Imaging; Magnetic Resonance; Phantoms; Signal to noise ratio;
D O I
10.22038/ijmp.2019.38606.1501
中图分类号
学科分类号
摘要
Introduction: The present study aimed to three frequently used pulse sequences of magnetic resonance imaging (MRI) to assess the image quality of theses pulse sequences at short acquisition time. Material and Methods: For the purpose of study two tissue equivalent gels were prepared. One gel was made from Polysaccharide and Agarose, whereas second gel was obtained from Ferrous Benzoic Xylenol Orange (FBX) which is tissue equivalent material. 6MV photons were used to irradiate FBX gel from linear accelerator with 25 Gray dose. Imaging parameters are performed in repetition time (TR) for experimental variations. The quantitative analysis included contrast-to-noise ratio (CNR) and signal to noise ratio (SNR). Results: As evidenced by obtained results at 1.5 Tesla, Fast Spin Echo (FSE) and Fast Fluid Attenuated Inversion Recovery (FLAIR) were most comparable in SNR although, acquisition time of FSE is 62%, 9 %, and 15% less than FLAIR at different values of 4000ms, 4200ms and 4600ms of TR. CNR of Conventional Spin Echo (CSE) was 143% and 93% better than FSE and FLAIR respectively. The time difference between CSE and FSE was 6 min and 34 sec while this difference was 6 min and 43 sec between CSE and FLAIR. Conclusion: FSE and FLAIR produced optimal image quality for many tissues. Their reduced acquisition time could make them perfect option for patients who cannot tolerate longer imaging time. Nonetheless long acquisition time cannot undervalue importance of CSE since it has yielded significantly higher contrast and SNR in T2-weighted images among other pulse sequences of MRI. © 2020. All Rights Reserved.
引用
收藏
页码:350 / 358
页数:8
相关论文
共 26 条
[1]  
Katarzyna K, Monika BF., Artifacts in magnetic resonance imaging, Pol J Radiol, 80, pp. 93-106, (2015)
[2]  
Morelli JN, Runge VM, Ai F, Attenberger U, Vu L, Schmeets SH, Et al., An. image-based approach to understanding the physics of MR artifacts, Radiographics, 31, pp. 849-866, (2011)
[3]  
Honda E, Sasaki T, Simms FC, Maruyama K., An optimized fast protocol for magnetic resonance imaging of the temporomandibular joint, Dentomaxillofacial. Radiology, 30, (2001)
[4]  
Tsao J., Ultrafast imaging: principles, pitfalls, solutions, and applications, J MagnReson Imaging, 32, 2, pp. 252-266, (2010)
[5]  
Meng Belinda, Umi N, Bee K, Ibrahim S., A framework of MRI fat suppressed imaging fusion system for femur abnormality analysis, Procedia Computer Science, 60, pp. 808-817, (2015)
[6]  
Mekle R, Laine AF, Wu EX., Combined MR data acquisition of multicontrast images using variable acquisition parameters and K-space data sharing, IEEE Trans Med Imaging, 22, 7, pp. 806-823, (2003)
[7]  
Augui J, Vignaux O, Argaud C, Coste J, Gouya H, Legmann P., Liver: T2-weighted MR Imaging with breath-hold fast-recovery optimized fast spin-Echo compared with breath-hold half-fourier and nonbreath-hold respiratory-triggered fast spin-echo pulse sequences, Radiology, 223, pp. 853-859, (2002)
[8]  
Naima A, Afzal M, Yousaf M, Arshad J., Comparison amongst pulse sequences and imaging parameters for enhanced CNR in T1, T2-weighted study of MRI, J Pak Med Assoc, 67, pp. 225-232, (2017)
[9]  
Mokhtar M, Zeineb T, Mouna C, Fethi L., Comparison of 3D MR imaging sequences in knee articular cartilage at 1.5 T, Biomedical Research, 29, 14, pp. 2963-2970, (2018)
[10]  
Westbrook C, Kaut C, Talbot J., MRI in Practice, (2011)