Tidal breathing patterns derived from structured light plethysmography in COPD patients compared with healthy subjects

被引:25
作者
Motamedi-Fakhr S. [1 ]
Wilson R.C. [1 ]
Iles R. [2 ]
机构
[1] PneumaCare Ltd, Ely
[2] Cambridge University Hospitals NHS Foundation Trust, Cambridge
关键词
Chronic obstructive pulmonary disease; IE50; Structured light plethysmography; Thoraco–abdominal asynchrony; Tidal breathing;
D O I
10.2147/MDER.S119868
中图分类号
R96 [药理学]; R3 [基础医学]; R4 [临床医学];
学科分类号
1001 ; 1002 ; 100602 ; 100706 ;
摘要
Purpose: Differences in tidal breathing patterns have been reported between patients with chronic obstructive pulmonary disease (COPD) and healthy individuals using traditional measurement techniques. This feasibility study examined whether structured light plethysmography (SLP) – a noncontact, light-based technique – could also detect differences in tidal breathing patterns between patients with COPD and healthy subjects. Patients and methods: A 5 min period of tidal (quiet) breathing was recorded in each patient with COPD (n=31) and each healthy subject (n=31), matched for age, body mass index, and sex. For every participant, the median and interquartile range (IQR; denoting within-subject variability) of 12 tidal breathing parameters were calculated. Individual data were then combined by cohort and summarized by its median and IQR. Results: After correction for multiple comparisons, inspiratory time (median tI) and its variability (IQR of tI) were lower in patients with COPD (p<0.001 and p<0.01, respectively) as were ratios derived from tI (tI/tE and tI/tTot, both p<0.01) and their variability (p<0.01 and p<0.05, respectively). IE50SLP(the ratio of inspiratory to expiratory flow at 50% tidal volume calculated from the SLP signal) was higher (p<0.001) in COPD while SLP-derived time to reach peak tidal expiratory flow over expiratory time (median tPTEFSLP/tE) was shorter (p<0.01) and considerably less variable (p<0.001). Thoraco–abdominal asynchrony was increased (p<0.05) in COPD. Conclusion: These early observations suggest that, like traditional techniques, SLP is able to detect different breathing patterns in COPD patients compared with subjects with no respiratory disease. This provides support for further investigation into the potential uses of SLP in assessing clinical conditions and interventions. © 2017 Motamedi-Fakhr et al.
引用
收藏
页码:1 / 9
页数:8
相关论文
共 30 条
[1]  
Ghezzi M., Tenero L., Piazza M., Bodini A., Piacentini G., Structured light plethysmography: New method to evaluate expiratory flow limitation in asthmatic children, Eur Respir J, 46, (2015)
[2]  
Hmeidi H., Chadwick E., Lenney W., Et al., IE50 measured by structured light plethysmography (SLP) can differentiate between children with and without asthma, and can detect response to a bronchodilator, Am J Respir Crit Care Med, (2016)
[3]  
Hmeidi H., Chadwick E., Lenney W., Et al., Structured light plethysmography (SLP) can quantify abnormal breathing in children aged 2-12 admitted with acute asthma, Am J Respir Crit Care Med, (2016)
[4]  
Elshafie G., Kumar P., Motamedi-Fakhr S., Iles R., Wilson R.C., Naidu B., Measuring changes in chest wall motion after lung resection using structured light plethysmography: A feasibility study, Interact Cardiovasc Thorac Surg, 23, 4, pp. 544-547, (2016)
[5]  
Chen Y., Xin Z., Qin C., Analysis of tidal breathing flow-volume curves in stable COPD patients, Chinese J Prac Intern Med, 11, pp. 978-980, (2005)
[6]  
Kostianev S., Hristova A., Iluchev D., Characteristics of tidal expiratory flow pattern in healthy people and patient with chronic obstructive pulmonary disease, Folia Med (Plovdiv), 41, 3, pp. 18-25, (1999)
[7]  
Morris M.J., Williams E.M., Madgwick R., Banerjee R., Phillips E., Changes in lung function and tidal airflow patterns after increasing extrathoracic airway resistance, Respirology, 9, 4, pp. 474-480, (2004)
[8]  
Wilkens H., Weingard B., Lo Mauro A., Et al., Breathing pattern and chest wall volumes during exercise in patients with cystic fibrosis, pulmonary fibrosis and COPD before and after lung transplantation, Thorax, 65, 9, pp. 808-814, (2010)
[9]  
Williams E.M., Madgwick R.G., Morris M.J., Tidal expired airflow patterns in adults with airway obstruction, Eur Respir J, 12, 5, pp. 1118-1123, (1998)
[10]  
Williams E.M., Powell T., Eriksen M., Neill P., Colasanti R., A pilot study quantifying the shape of tidal breathing waveforms using centroids in health and COPD, J Clin Monit Comput, 28, 1, pp. 67-74, (2014)