Development of an automated blood sampling system for use in humans

被引:77
作者
Henley, D.E. [1 ,2 ]
Leendertz, J.A. [1 ,3 ]
Russell, G.M. [1 ]
Wood, S.A. [1 ]
Taheri, S. [1 ,5 ]
Woltersdorf, W.W. [1 ,4 ]
Lightman, S.L. [1 ]
机构
[1] Henry Wellcome Laboratories for Integrative Neuroscience and Endocrinology, University of Bristol, Bristol
[2] Faculty of Medicine, Dentistry Health Sciences, University of Western Australia, Perth, WA
[3] Department of Electrical and Electronic Engineering, University of Bristol, Bristol
[4] Department of Clinical Biochemistry, Bristol Royal Infirmary, Bristol
[5] Division of Medical Sciences, University of Birmingham and Heart of England NHS Foundation Trust, Birmingham
关键词
Automated blood sampling; Cortisol; Hypothalamic-pituitary- adrenal axis; Stress; Ultradian;
D O I
10.1080/03091900802185970
中图分类号
学科分类号
摘要
Many hormones are released in a pulsatile or burst-like pattern resulting in fluctuating blood levels that can undergo rapid modulation by physiological and pathological signals. To accurately measure these changes in hormone concentration requires frequent blood sampling, often over extended periods as the overall rhythmicity may vary over 24 hours. The aim of this study was to develop a computerized, automated blood sampling system which allows repeated stress-free blood sample collection from humans over an extended period under basal or test conditions. The system incorporates a peristaltic pump, fraction collector and standard infusion pump together with a custom built electronic control unit linked to a personal computer. Disposable tubing prevents cross-contamination between study participants. The computer programme is modifiable to adjust for the number of specimen tubes and volume of blood collected per sampling cycle. Patency of the collecting line is maintained with 0.9% saline, without the need for heparinization. To validate the system, 10-minute samples for cortisol were collected over 24 hours from five healthy volunteers, of whom two had additional concomitant ACTH sampling. Deconvolution analysis revealed an expected number of hormone secretory episodes and a non-pathological degree of orderliness within the data. There was high concordance between ACTH and cortisol secretory events. The ability of the system to allow multiple measurements and of the software program to link with other physiological monitoring equipment provides a powerful tool to study physiologic/pathophysiologic change in relation to blood hormone and other biomarker levels.
引用
收藏
页码:199 / 208
页数:9
相关论文
共 34 条
[1]  
Veldhuis J.D., Carlson M.L., Johnson M.L., The pituitary gland secretes in bursts: Appraising the nature of glandular secretory impulses by simultaneous multiple-parameter deconvolution of plasma hormone concentrations, Proceedings of the National Academy of Sciences USA, 84, pp. 7686-7690, (1987)
[2]  
Lightman S.L., Windle R.J., Julian M.D., Harbuz M.S., Shanks N., Wood S.A., Kershaw Y.M., Ingram C.D., Significance of pulsatility in the HPA axis, Novartis Foundation Symposium, 227, pp. 244-260, (2000)
[3]  
Young E.A., Abelson J., Lightman S.L., Cortisol pulsatility and its role in stress regulation and health, Frontiers in Neuroendocrinology, 25, pp. 69-76, (2004)
[4]  
Conway-Campbell B.L., McKenna M.A., Wiles C.C., Atkinson H.C., de Kloet E.R., Lightman S.L., Proteosome-dependent down-regulation of activated nuclear hippocampal glucocorticoid receptors determines dynamic responses to corti-costerone, Endocrinology, 148, pp. 5470-5477, (2007)
[5]  
Young E.A., Carlson N.E., Brown M.B., Twenty-four-hour ACTH and cortisol pulsatility in depressed women, Neuropsychopharmacology, 25, pp. 267-276, (2001)
[6]  
Young E.A., Veldhuis J.D., Disordered adrenocorticotro-pin secretion in women with major depression, Journal of Clinical Endocrinology & Metabolism, 91, pp. 1924-1928, (2006)
[7]  
Matthews D.R., Naylor B.A., Jones R.G., Ward G.M., Turner R.C., Pulsatile insulin has greater hypoglycemic effect than continuous delivery, Diabetes, 32, pp. 617-621, (1983)
[8]  
Matthews D.R., Lang D.A., Burnett M.A., Turner R.C., Control of pulsatile insulin secretion in man, Diabetologia, 24, pp. 231-237, (1983)
[9]  
O'Rahilly S., Turner R.C., Matthews D.R., Impaired pulsatile secretion of insulin in relatives of patients with non-insulin-dependent diabetes, New England Journal of Medicine, 318, pp. 1255-1230, (1988)
[10]  
Polonsky K.S., Given B.D., Hirsch L.J., Tillil H., Shapiro E.T., Beebe C., Frank B.H., Galloway J.A., Van Cauter E., Abnormal patterns of insulin secretion in non-insulin-dependent diabetes mellitus, New England Journal of Medicine, 318, pp. 1231-1239, (1988)