Adaptive tracking of orthopedic force for bone external fixation robot based on impedance control

被引:0
作者
Su P. [1 ]
Xie S. [1 ]
Liu L. [1 ]
Zhao W. [2 ]
Zhang L. [2 ]
机构
[1] School of Mechanical and Electrical Engineering, Beijing Information Science and Technology University, Beijing
[2] Rehabilitation Hospital, National Research Center for Rehabilitation Technical Aids, Beijing
来源
Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument | 2023年 / 44卷 / 11期
关键词
adaptive impedance control; bone external fixation robot; force tracking control; impedance parameters; RCM mechanism;
D O I
10.19650/j.cnki.cjsi.J2311359
中图分类号
学科分类号
摘要
To address the problem that the orthopedic force of the traditional bone external fixator cannot be accurately controlled and has poor safety, the article establishes a knee bone external fixation robot based on the RCM configuration and proposes an orthopedic force tracking control method with indirect adaptive impedance. Due to the existence of mechanical friction and external uncertainty, to address the problem of poor robustness of impedance control, the force error signal is adopted as the driving force of the target impedance, a new impedance model is formulated to adapt to the changes in the orthopedic environment. According to the changes in the contact force, the adaptive law is designed to regulate the impedance parameters in real time, which compensates for the uncertainty of the environmental dynamics on the line. Therefore, the system′s force tracking error is zero. Comparative simulations and prototype experiments of the proposed control algorithm for tibial orthopedic force control are implemented. The results show that the impedance control simulation results of the weighted average of the performance indicators are 2. 12, 8. 58 and 13. 2, and the maximum error of the force tracking experiment is 20 N. Compared with impedance control, the weighted average of the performance indicators of adaptive impedance control simulation is only 0. 36, 0. 18 and 0. 61, and the force tracking fluctuation error is controlled to be within ±3N, which has better robust and adaptive ability. © 2023 Science Press. All rights reserved.
引用
收藏
页码:99 / 108
页数:9
相关论文
共 25 条
  • [1] WEI M T, CHEN J W, GUO Y, Et al., The compute raided parallel external fixator for complex lower limb deformity correction [J], International Journal of Computer Assisted Radiology & Surgery, 2, 12, pp. 2107-2117, (2017)
  • [2] YAO Y F, YANG Y L, GUO J L, Et al., A review of postoperative knee rehabilitation robot research, Journal of Mechanical Engineering, 57, 5, pp. 1-18, (2021)
  • [3] MENG L, DONG H T, HOU J, Et al., Soft exoskeleton robot facing to lower-limb rehabilitation: A narrative review, Chinese Journal of Scientific Instrument, 42, 4, pp. 206-217, (2021)
  • [4] SU P, LAI Y L, ZHANG L, Et al., Design and analysis of centering orthopedic bone external fixation robot, Chinese Journal of Scientific Instrument, 43, 11, pp. 262-273, (2022)
  • [5] ZHANG L, QIN S H, SHAO J J, Et al., Innovative application of orthoses combined with external fixator in the correcting deformity of lower limbs, Orthopedic Journal of China, 27, 7, pp. 655-658, (2019)
  • [6] CUI R, CHEN D SH, SU P, Et al., Research progress on trajectory planning of fracture reduction and deformity correction robot, Journal of Mechanical Engineering, 58, 13, pp. 1-21, (2022)
  • [7] NI Z Q, WANG T M, LIU D, Et al., Survey on medical robotics, Journal of Mechanical Engineering, 51, 13, pp. 45-52, (2015)
  • [8] ZUO S P, DONG M J, LI J F, Et al., Configuration design and correction ability evaluation of a novel external fixator for foot and ankle deformity treated by U osteotomy, Medical & Biological Engineering & Computing, 58, 3, pp. 541-558, (2020)
  • [9] HAN H ZH, Research and development of bone external fixation system, (2016)
  • [10] DI J., Researches on key technologies of simulation of Taylor external fixator based on OpenGL, (2015)