The impact of the Impeller's hub design on the performance and blood damage in a microaxial mechanical circulatory support device – A numerical study

被引:0
作者
Gabso Y. [1 ,2 ]
Rosenfeld M. [1 ]
Avrahami I. [2 ]
机构
[1] School of Mechanical Engineering, Tel Aviv University, Tel Aviv
[2] Department of Mechanical Engineering and Mechatronics, Ariel University
关键词
Computational fluid dynamics; Hemolysis; Impeller design; Mechanical circulatory support device; Platelets activation;
D O I
10.1016/j.compbiomed.2024.108695
中图分类号
学科分类号
摘要
This study uses CFD methods to investigate the effects of the impeller's geometry on the hemodynamic characteristics, pump performance, and blood damage parameters, in a percutaneous microaxial Mechanical Circulatory Support (MCS) device. The numerical simulations employ the steady state Reynolds-Averaged Navier-Stokes approximation using the SST k−ω turbulent model. Three different impeller models are examined with different hub conversion angles (α = 0○, 3○ and 5○). The analysis includes 23 cases for different pressure heads (Δp = 60–80 mmHg) and angular velocities (ω = 30–52 kRPM). The obtained flow rate is compared between the cases to assess the effect of the impeller's design and working conditions on the pump performance. The comparative risk of shear-induced platelet activation is estimated using the statistical median of the stress-accumulation values calculated along streamlines. The risk of hemolysis is estimated using the average exposure time to shear stress above a threshold (τ > 425 Pa). The results reveal that the shape of the impeller's hub has a great impact on the flow patterns, performance, and risk of blood damage, as well as the angular velocity. The highest flow rate (Q = 3.7 L/min) and efficiency (η = 11.3 %) were achieved using a straight hub (α = 0○). Similarly, for the same condition of flow and pressure, the straight hub impeller has the lowest blood damage risk parameters. This study shed light on the effect of pump design on the performance and risk of blood damage, indicating the roles of the hub shape and angular velocity as dominant parameters. © 2024 Elsevier Ltd
引用
收藏
相关论文
共 41 条
[1]  
Dhruva S.S., Ross J.S., Mortazavi B.J., Hurley N.C., Krumholz H.M., Curtis J.P., Berkowitz A.P., Masoudi F.A., Messenger J.C., Parzynski C.S., Ngufor C.G., Girotra S., Amin A.P., Shah N.D., Desai N.R., Use of mechanical circulatory support devices among patients with acute myocardial infarction complicated by cardiogenic shock, JAMA Netw. Open, 4, 2, (2021)
[2]  
Amin A.P., Spertus J.A., Curtis J.P., Desai N., Masoudi F.A., Bach R.G., McNeely C., Al-Badarin F., House J.A., Kulkarni H., Rao S.V., The evolving landscape of impella use in the United States among patients undergoing percutaneous cornary intervention with mechanical circulatory support, Circulation, pp. 273-284, (2020)
[3]  
Rihal C.S., Naidu S.S., Givertz M.M., Szeto W.Y., Burke J.A., Kapur N.K., Kern M., Garratt K.N., Goldstein J.A., Dimas V., Tu T., 2015 SCAI/ACC/HFSA/STS clinical expert consensus statement on the use of percutaneous mechanical circulatory support devices in cardiovascular care: endorsed by the American heart assocation, the cardiological society of India, and sociedad latino americ, J. Am. Coll. Cardiol., 65, 19, pp. e7-e26, (2015)
[4]  
Stretch R., Sauer C.M., Yuh D.D., Bonde P., National trends in the utilization of short-term mechanical circulatory support: incidence, outcomes, and cost analysis, J. Am. Coll. Cardiol., 64, 14, pp. 1407-1415, (2014)
[5]  
Cormican D.S., Madden C., Rodrigue M.F., Mechanical circulatory support: complications, outcomes, and future directions, Int. Anesthesiol. Clin., 60, 4, pp. 72-80, (2022)
[6]  
Li Y., Wang H., Xi Y., Sun A., Deng X., Chen Z., Fan Y., Multi-indicator analysis of mechanical blood damage with five clinical ventricular assist devices, Comput. Biol. Med., 151, (2022)
[7]  
Marcuschamer I.A., Abelow A., Ran K., The Evolving Role of Percutaneous Ventricular Assist Devices in High-Risk Cardiac Patients, (2018)
[8]  
Yun Z., Yao J., Wang L., Tang X., Feng Y., The design and evaluation of the outflow structures of an interventional microaxial blood pump, Front. Physiol., 14, (2023)
[9]  
Letzen B., Park J., Tuzun Z., Bonde P., Design and development of a miniaturized percutaneously deployable wireless left ventricular assist device : early prototypes and feasibility testing, ASAIO (January), pp. 147-152, (2018)
[10]  
Apel J., Neudel F., Reul H., Computational fluid dynamics and experimental validation of a microaxial blood pump, Am. Soc. Artif. Intern. Organs Trans., (2001)