On the performance of a Coriolis Mass Flowmeter (CMF): Experimental measurement and FSI simulation

被引:0
作者
Gace D.A. [1 ]
机构
[1] School of Science and Technology, University of Rwanda, Kigali
关键词
Coriolis flowmeters performance; Flow measurement; Fluid structure interactions; Iterative two-way coupling;
D O I
10.1051/ijmqe/2022002
中图分类号
学科分类号
摘要
Computational methods that make use of single one-way Fluid Structure Interaction (FSI) for modeling the Coriolis Mass Flowmeters' (CMFs) operations are prone to inaccuracies. These errors are due to their limitations in describing a fully coupled fluid structure interaction. The aim of this study is to produce a CFD model of a CMF that uses an iterative two-way coupling of fluid structure interaction to accurately study its performance. The computational findings are benchmarked against accurate experimental measurements of the U-shape CFM. The deviation between the computed results and experimental measurements remains about 0.1% which is deemed acceptable. This reduction of uncertainties is largely attributed to the capability of the model to describe the effects of tube vibrations on the meter's operation. © D.A. Gace, Published by EDP Sciences, 2022.
引用
收藏
相关论文
共 33 条
  • [1] Coriolis G.G., Mémoire sur les Équations du Mouvement Relatif des Systèmes de Corps, (1835)
  • [2] Henry M., Tombs M., Duta M., Zhou F., Mercado R., Kenyery F., Shen J., Morles M., Garcia C., Langansan R., Two-phase flow metering of heavy oil using a coriolis mass flow meter: A case study, Flow Measur. Instr., 17, pp. 399-413, (2006)
  • [3] Luo R., Wu J., Wan S., Fluid-structure coupling analysis and simulation of viscosity effect on Coriolis mass flowmeter, APCOM & ISCM, (2013)
  • [4] Skea A.F., Hall A.R.W., Effects of gas leaks in oil flow on single-phase flowmeters, Flow Measur. Instr., 10, pp. 145-150, (1999)
  • [5] Cheesewright R., Clark C., Bisset D., The identification of external factors which influence the calibration of Coriolis massflow meters, Flow Measur. Instrum., 11, pp. 1-10, (2000)
  • [6] Vetter G., Notzon S., Effect of pulsating flow on Coriolis mass flowmeters, Flow Measur. Instr., 5, pp. 263-273, (1994)
  • [7] Svete A., Kutin J., Bobovnik G., Bajsi I., Theoretical and experimental investigations of flow pulsation effects in Coriolis mass flowmeters, J. Sound Vibr., 352, pp. 30-45, (2015)
  • [8] Belhadj A., Cheesewright R., Clark C., The simulation of coriolis meter response to pulsating flow using a general purpose F.E. code, J. Fluids Struct., 14, pp. 613-634, (2000)
  • [9] Cheesewright R., Belhadj A., Clark C., Effect of mechanical vibrations on coriolis mass flow meters, J. Dyn. Syst. Measur. Control Trans. ASME, 125, pp. 103-113, (2003)
  • [10] Cheesewright R., Clark C., Hou Y.Y., The response of Coriolis flowmeters to pulsating flows, Flow Measur. Instru., 15, pp. 59-67, (2004)