Validation of Monte Carlo Model for Dose Evaluation outside the Treatment Field for Siemens 6MV Beam

被引:0
作者
Sinousy D.M. [1 ]
Attalla E.M. [2 ]
Fathy I.H. [3 ,4 ,5 ]
Elhussiny F.A. [4 ]
Elmekawy A.F. [4 ]
机构
[1] National Cancer Institute, Cairo University
[2] Faculty of Science, Physics Department, Cairo University
来源
Sinousy, Dina Mohamed | 1600年 / Mashhad University of Medical Sciences卷 / 17期
关键词
Linear Accelerator; Monte Carlo Method; Out-of-Field Dose; Radiotherapy;
D O I
10.22038/ijmp.2019.42881.1646
中图分类号
学科分类号
摘要
Introduction: There has been a concern about the unintended doses to critical structures outside the treatment field due to the increased risk of radiation-induced second cancer following radiotherapy — treatments. Today, Monte Carlo (MC) simulation is considered the most accurate method for dose calculations in different domains of medical physics. Material and Methods: The Geant4 Application for Tomographic Emission (GATE) code was used to create an MC model of 6MV Siemens Primus linac. Measurements were taken in a water phantom using an ion — chamber to validate the MC model. Dose profiles outside the treatment field at 1.5 (dmax), 5.0 andl0.0 cm depths for field sizes from 5&multipul; 5 to 20&multipul; 20 cm2 were measured in the present study. Out-of-field percentage depth dose (PDD) curves at 0.0, 5.0, and 7.5 cm off axis for field size 10&multipul; 10 cm2 were investigated for both measurements and simulation. However out-of-field PDDs from 10 to 15 cm off axis for field size 10&multipul; 10 cm2 were studied only by simulation. Results: The comparisons showed agreement between the measured and simulated doses for the out-of-field profiles along the in-plane direction for all considered field sizes and depths, as well as for the PDDs at 0.0 and 5.0 cm off axis, but with less agreement at 7.5 cm off axis. All the simulated out-of-field PDDs at distances > 10 cm off axis had similar trend shapes. Conclusion: The developed MC model is considered a good representation of 6 MV Siemens Primus linac for the out-of-field dose calculation in lieu of measurements. The authors would like to thank "Mit Ghamr Oncology Center" for its generous cooperation with this study. The authors would also like to extend their gratitude to Eng. Mohamed Osama, a biomedical engineer and president of On-medical Company for his great efforts in this project. © 2020. All Rights Reserved.
引用
收藏
页码:410 / 420
页数:10
相关论文
共 50 条
[11]   Monte Carlo modeling of a 6 and 18 MV Varian Clinac medical accelerator for in-field and out-of-field dose calculations: development and validation [J].
Bednarz, Bryan ;
Xu, X. George .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04) :N43-N57
[12]   Elaboration and experimental validation of a Monte Carlo source model for linac 6 MV photon beams with and without Flattening Filter [J].
Aboulbanine, Zakaria ;
Bahhous, Karim .
RADIATION PHYSICS AND CHEMISTRY, 2022, 201
[13]   A comparative Monte Carlo study on 6MV photon beam characteristics of Varian 21EX and Elekta SL-25 linacs [J].
Mesbahi, A. ;
Mehnati, P. ;
Keshtkar, A. .
IRANIAN JOURNAL OF RADIATION RESEARCH, 2007, 5 (01) :23-30
[14]   Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam [J].
Borges, C. ;
Zarza-Moreno, M. ;
Heath, E. ;
Teixeira, N. ;
Vaz, P. .
MEDICAL PHYSICS, 2012, 39 (01) :415-423
[15]   Monte Carlo evaluation of out-of-field dose in 18 MV pelvic radiotherapy using a simplified female MIRD phantom [J].
Geraily, Ghazale ;
Elmtalab, Soheil ;
Mohammadi, Najmeh ;
Alirezaei, Zahra ;
Martinez-Ovalle, S. A. ;
Jabbari, Iraj ;
Vega-Carrillo, Hector Rene ;
Karimi, Amir Hossein .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (01)
[16]   Determination of the small-field output factor for 6 MV photon beam using EGSnrc Monte Carlo [J].
Chuah, K. W. ;
Aziz, Abdul M. Z. ;
Jayamani, J. .
JOURNAL OF MEDICAL PHYSICS, 2022, 47 (03) :301-308
[17]   Verification measurements and clinical evaluation of the iPlan RT Monte Carlo dose algorithm for 6 MV photon energy [J].
Petoukhova, A. L. ;
van Wingerden, K. ;
Wiggenraad, R. G. J. ;
van de Vaart, P. J. M. ;
van Egmond, J. ;
Franken, E. M. ;
van Santvoort, J. P. C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16) :4601-4614
[18]   Monte Carlo dosimetry of a cell culture irradiation model using a 6 MV X-ray beam [J].
Hernandez-Bojorquez, Mariana ;
Trejo-Solis, Cristina ;
Manuel Larraga-Gutierrez, Jose ;
Martinez-Davalos, Arnulfo .
RADIATION PHYSICS AND CHEMISTRY, 2021, 180
[19]   A Monte Carlo study on dose perturbation due to dental restorations in a 15 MV photon beam [J].
Azizi, Mona ;
Mowlavi, Ali Asghar ;
Ghorbani, Mehdi ;
Knuap, Courtney ;
Behmadi, Marziyeh .
JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (03) :491-497
[20]   An investigation of the depth dose in the build-up region, and surface dose for a 6-MV therapeutic photon beam: Monte Carlo simulation and measurements [J].
Apipunyasopon, Lukkana ;
Srisatit, Somyot ;
Phaisangittisakul, Nakorn .
JOURNAL OF RADIATION RESEARCH, 2013, 54 (02) :374-382