3D-printed activated charcoal inlet filters for oxygen concentrators: A circular economy approach

被引:9
作者
Williams E. [1 ]
Piaggio D. [1 ]
Andellini M. [1 ]
Pecchia L. [1 ]
机构
[1] School of Engineering, University of Warwick, CV7AL, Coventry
基金
英国工程与自然科学研究理事会;
关键词
Circular economy; Clinical engineering; Filter; Oxygen concentrator; Supply chain; Sustainable development;
D O I
10.1016/j.deveng.2022.100094
中图分类号
学科分类号
摘要
As of May 2021, the current COVID-19 pandemic is still plaguing the world, challenging all the countries and their health systems, globally. In this context, conditions typical of low-resource settings surfaced also in high-resource ones (e.g., the lack of essential medical equipment, of resources etc.), while exacerbating in the already resource-scarce settings, because of COVID-19. This is the case of oxygen concentrators that are one of the first-line medical devices for treating COVID-19 patients. Since the beginning of 2020, their demand has been rapidly growing worldwide, aggravating the situation for low-resource settings, where the availability of devices providing oxygen-enriched air was already scarce. In fact, due to their delicacy, the lack of spare parts and of an appropriate health technology management system, oxygen concentrators can often be found broken or not working properly in these settings. The underlying problems have deep roots. The current regulatory frameworks and standards, which are set by high-income countries, are too stringent, and do not take into account the limited resources of poorer settings. Thus, they are often inapplicable in such settings. One of the main issues affecting the oxygen concentrators, is that related to the filters, which are designed to filter out dust, particles, bacteria, and to be used in medical locations complying with international standards (e.g., the air filtration level in a surgical theatre in Italy is at 99.97%). When used in low-resource settings, which do not comply with these standards and face several challenges (e.g., dust), these filters have a much-reduced lifespan. For these reasons, this paper aims to present the redesign of the inlet filter of an oxygen concentrator, which is used to prevent gross particles to enter the device. The redesign is based on a reverse engineering approach, and on the use of 3D-printing along with activated charcoal. After testing the filtration efficiency with a particle counter, the filter design has been refined through several iterations. The final prototype performs particularly well when filtering particles above 1 μm (with a filtration efficiency of 64.2%), and still has a satisfactory performance with any particle size over 0.3 μm (with a filtration efficiency of 38.8%). Following the United Nations Sustainable Development Goals, this project aims to empower local communities, and start a positive trend of self-sustained supply chain of simple spare parts for medical devices, leveraging on frugal engineering, 3D-printing, locally produced activated charcoal, and circular economy. © 2022
引用
收藏
相关论文
共 29 条
  • [1] Andellini M., De Santis S., Nocchi F., Bassanelli E., Pecchia L., Ritrovato M., Clinical needs and technical requirements for ventilators for COVID-19 treatment critical patients: an evidence-based comparison for adult and pediatric age, Health Technol., 10, 6, pp. 1403-1411, (2020)
  • [2] Arasaratnam A., Humphreys G., Emerging economies drive frugal innovation, World Health Organization. Bulletin of the World Health Organization, 91, 1, (2013)
  • [3] Beroza A., 3D printing IN low resource healthcare settings: analysis OF potential implementations, (2019)
  • [4] Bhisey R., Portable oxygen concentrators market: covid-19 impact on global market demand, opportunities, and analysis & forecast to 2026, (2020)
  • [5] Bradley B., Light J., Ebonyi A., N'Jai P., Ideh R., Ebruke B., Nyassi E., Peel D., Howie S., Implementation and 8-year follow-up of an uninterrupted oxygen supply system in a hospital in the Gambia, Int. J. Tubercul. Lung Dis., 20, 8, pp. 1130-1134, (2016)
  • [6] Bradley B., Light J., Singhateh B., Koli D., Peel D., Cheng Y.-L., Howie S., Temperature and humidity trends in a health centre in the Gambia: implications for back up battery-life in tropical settings, 7th International Conference on Appropriate Healthcare Technologies for Developing Countries, (2012)
  • [7] Bradley B.D., Chow S., Cheng Y.-L., Nyassi E., Howie S.R., Peel D., The Use of Oxygen Concentrators in the Gambia: A Study of over 5 Years of Experience in, (2013)
  • [8] Bradley B.D., Chow S., Nyassi E., Cheng Y.-L., Peel D., Howie S.R., A retrospective analysis of oxygen concentrator maintenance needs and costs in a low-resource setting: experience from the Gambia, Health Technol., 4, 4, pp. 319-328, (2015)
  • [9] Chawla S., Kurani S., Wren S.M., Stewart B., Burnham G., Kushner A., McIntyre T., Electricity and generator availability in LMIC hospitals: improving access to safe surgery, J. Surg. Res., 223, pp. 136-141, (2018)
  • [10] Di Pietro L., Piaggio D., Oronti I., Maccaro A., Houessouvo R.C., Medenou D., De Maria C., Pecchia L., Ahluwalia A., A framework for assessing healthcare facilities in low-resource settings: field studies in Benin and Uganda, J. Med. Biol. Eng., pp. 1-9, (2020)