Assessment of 4D flow MRI for quantification of left-to-right shunt in pediatric patients with ventricular septal defect: comparison with right heart catheterization

被引:2
作者
Zamani-Aliabadi, Seyed Mohammad [1 ]
Qanadli, Salah D. [2 ]
Fatehi-Feyzabad, Seyed Hasan [3 ]
Ghasemnezhad, Mohsen [4 ]
Ghaemi, Hamidreza [4 ]
Azarine, Arshid [5 ]
Mohammadzadeh, Ali [3 ]
Bitarafan-Rajabi, Ahmad [1 ]
Mortezaeian, Hojjat [4 ]
Rezaei-Kalantari, Kiara [3 ]
机构
[1] Iran Univ Med Sci, Sch Med, Dept Med Phys, Tehran, Iran
[2] Lausanne Univ Hosp, Dept Diagnost & Intervent Radiol, Cardiothorac & Vasc Div, Lausanne, Switzerland
[3] Iran Univ Med Sci, Rajaie Cardiovasc & Med Res Ctr, Dept Radiol, Tehran, Iran
[4] Iran Univ Med Sci, Rajaie Cardiovasc Med & Res Ctr, Dept Pediat Cardiol, Tehran, Iran
[5] Univ Paris Saclay, Hop Marie Lannelongue, Dept Radiol, Grp Hosp Paris St Joseph, Saclay, Paris, France
来源
FRONTIERS IN CARDIOVASCULAR MEDICINE | 2024年 / 11卷
关键词
right heart catheterization; cardiac magnetic resonance imaging; four-dimensional (4D) flow; two-dimensional (2D) flow; shunt fraction; ventricular septal defect (VSD); congenital heart disease (CHD); MAGNETIC-RESONANCE; CARDIAC-CATHETERIZATION; AMERICAN SOCIETY; BLOOD-FLOW; ECHOCARDIOGRAPHY; DISEASE; ADULT; LIMITATIONS; GUIDELINES;
D O I
10.3389/fcvm.2024.1399110
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives The percentage of shunt fraction significantly impacts the management of patients with congenital shunts, influencing strategic choices such as surgical or interventional procedures. This study compared the estimated shunt fraction (the ratio of pulmonary-to-systemic flow, Qp/Qs) for quantifying the left-to-right shunt in children with ventricular septal defect (VSD) using heart catheterization, four-dimensional (4D) flow, and two-dimensional (2D) flow magnetic resonance imaging (MRI). The goal was to establish a non-invasive and reliable measurement ratio between pulmonary and systemic blood flow in these patients.Methods Between July 2022 and June 2023, patients scheduled to undergo invasive right heart catheterization were included in this study. MRI was performed one hour before the catheterization procedure. The correlation of shunt fraction was assessed between all methods after calculating the Qp/Qs ratio from 2D and 4D flow MRI and catheterization.Results A total of 24 patients (aged 3-15 years, eight females) were ultimately included in the study. The Qp/Qs ratios obtained from 4D flow had a robust correlation (correlation coefficient r = 0.962) compared to those obtained during catheterization. Cardiac catheterization recorded the mean shunt fraction at 1.499 +/- 0.396, while 4D flow measured it at 1.403 +/- 0.344, with no significant difference between the two techniques. Moreover, there was a reasonable correlation (r = 0.894) between 2D flow measurements of Qp/Qs and the results obtained from catheterization, with a mean shunt fraction of 1.326 +/- 0.283.Conclusion 4D flow MRI has the potential to be a non-invasive method for accurately measuring the left-to-right shunt in children with VSD.
引用
收藏
页数:8
相关论文
共 38 条
[31]   Characterization and improved quantification of left ventricular inflow using streamline visualization with 4DFlow MRI in healthy controls and patients after atrioventricular septal defect correction [J].
Calkoen, Emmeline E. ;
Roest, Arno A. W. ;
Kroft, Lucia J. M. ;
van der Geest, Rob J. ;
Jongbloed, Monique R. M. ;
van den Boogaard, Pieter J. ;
Blom, Nico A. ;
Hazekamp, Mark G. ;
de Roos, Albert ;
Westenberg, Jos J. M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (06) :1512-1520
[32]   Scan-rescan reproducibility of diastolic left ventricular kinetic energy, viscous energy loss and vorticity assessment using 4D flow MRI: analysis in healthy subjects [J].
Kamphuis, Vivian P. ;
Westenberg, J. M. ;
van der Palen, Roel L. F. ;
van den Boogaard, Pieter J. ;
van der Geest, Rob J. ;
de Roos, Albert ;
Blom, Nico A. ;
Roest, Arno A. W. ;
Elbaz, Mohammed S. M. .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2018, 34 (06) :905-920
[33]   Use of Hand Carried Ultrasound, B-type Natriuretic Peptide, and Clinical Assessment in Identifying Abnormal Left Ventricular Filling Pressures in Patients Referred for Right Heart Catheterization [J].
Goonewardena, Sascha N. ;
Blair, John E. A. ;
Manuchehry, Amin ;
Brennan, J. Matthew ;
Keller, Michael ;
Reeves, Ryan ;
Price, Adam ;
Spencer, Kirk T. ;
Puthumana, Jyothy ;
Gheorghiade, Mihai .
JOURNAL OF CARDIAC FAILURE, 2010, 16 (01) :69-75
[34]   Hemodynamic forces in the left and right ventricles of the human heart using 4D flow magnetic resonance imaging: Phantom validation, reproducibility, sensitivity to respiratory gating and free analysis software [J].
Toger, Johannes ;
Arvidsson, Per M. ;
Bock, Jelena ;
Kanski, Mikael ;
Pedrizzetti, Gianni ;
Carlsson, Marcus ;
Arheden, Hakan ;
Heiberg, Einar .
PLOS ONE, 2018, 13 (04)
[35]   Gated F-18 FDG PET for Assessment of Left Ventricular Volumes and Ejection Fraction Using QGS and 4D-MSPECT in Patients with Heart Failure: A Comparison with Cardiac MRI [J].
Li, Yan ;
Wang, Li ;
Zhao, Shi-Hua ;
He, Zuo-Xiang ;
Wang, Dao-Yu ;
Guo, Feng ;
Fang, Wei ;
Yang, Min-Fu .
PLOS ONE, 2014, 9 (01)
[36]   Novel application of 4D sonography with B-flow imaging and spatio-temporal image correlation (STIC) in the assessment of the anatomy of pulmonary arteries in fetuses with pulmonary atresia and ventricular septal defect [J].
Volpe, P. ;
Campobasso, G. ;
Stanziano, A. ;
De Robertis, V. ;
Di Paolo, S. ;
Caruso, G. ;
Volpe, N. ;
Gentile, M. .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2006, 28 (01) :40-46
[37]   Left ventricular global longitudinal strain in bicupsid aortic valve patients: head-to-head comparison between computed tomography, 4D flow cardiovascular magnetic resonance and speckle-tracking echocardiography [J].
van den Hoven, Allard T. ;
Yilmazer, Sultan ;
Chelu, Raluca G. ;
van Grootel, Roderick W. J. ;
Minderhoud, Savine C. S. ;
Bons, Lidia R. ;
van Berendoncks, An M. ;
Duijnhouwer, Anthonie L. ;
Siebelink, Hans-Marc J. ;
van den Bosch, Annemien E. ;
Budde, Ricardo P. J. ;
Roos-Hesselink, Jolien W. ;
Hirsch, Alexander .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2020, 36 (09) :1771-1780
[38]   Left ventricular global longitudinal strain in bicupsid aortic valve patients: head-to-head comparison between computed tomography, 4D flow cardiovascular magnetic resonance and speckle-tracking echocardiography [J].
Allard T. van den Hoven ;
Sultan Yilmazer ;
Raluca G. Chelu ;
Roderick W. J. van Grootel ;
Savine C. S. Minderhoud ;
Lidia R. Bons ;
An M. van Berendoncks ;
Anthonie L. Duijnhouwer ;
Hans-Marc J. Siebelink ;
Annemien E. van den Bosch ;
Ricardo P. J. Budde ;
Jolien W. Roos-Hesselink ;
Alexander Hirsch .
The International Journal of Cardiovascular Imaging, 2020, 36 :1771-1780