Exploratory development and clinical research of a mixed reality guided radiotherapy positioning system

被引:0
作者
Zhang, Sai [1 ,2 ,3 ,4 ]
Zhang, Heng [1 ,2 ,3 ,4 ]
Li, Qixuan [1 ,2 ,3 ,4 ]
Wang, Ziyi [1 ,2 ,3 ,4 ]
Cao, Nannan [1 ,2 ,3 ,4 ]
Sun, Jiawei [1 ,2 ,3 ,4 ]
Xie, Kai [1 ,2 ,3 ,4 ]
Gao, Liugang [1 ,2 ,3 ,4 ]
Ni, Xinye [1 ,2 ,3 ,4 ]
机构
[1] Nanjing Med Univ, Affiliated Changzhou Peoples Hosp 2, Dept Radiotherapy, Changzhou 213003, Peoples R China
[2] Jiangsu Prov Engn Res Ctr Med Phys, Changzhou 213003, Peoples R China
[3] Nanjing Med Univ, Ctr Med Phys, Changzhou 213003, Peoples R China
[4] Key Lab Med Phys Changzhou, Changzhou 213003, Peoples R China
基金
中国国家自然科学基金;
关键词
Mixed reality; HoloLens2; Radiotherapy positioning; Virtual-real relation; AUGMENTED REALITY; RADIATION-THERAPY; IMMOBILIZATION; LOCALIZATION; ACCURACY; GUIDANCE; MOTION;
D O I
10.1016/j.jrras.2024.101095
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: This study aims to develop an application software deployed on HoloLens2 for guiding patient radiotherapy setup and assessing its feasibility and accuracy in radiotherapy. Methods: The developed system features an application software deployed on the HoloLens2 device, enabling automatic tracking and registration of a virtual cube to the location of a real calibration cube. This functionality is achieved through the utilization of the Vuforia SDK, which determines the location point linking the virtual and real spaces. The system then generates a reference hologram based on simulated positioning computed tomography (CT) data sets and anchors this reference hologram at the determined location point. In the process of treatment positioning, the real human body in the scene of HoloLens2 is matched with the surface of the reference hologram to complete the positioning. This paper explores the factors that affect the accuracy of the system, namely the viewing distance and Angle, to determine the optimal conditions for using the system. Then, based on the optimal conditions, the error between the cone beam computed tomography (CBCT) data of patients (head, neck and chest and abdomen) obtained by the system and conventional positioning were compared by Ttest method to evaluate the performance of the system. Results: The observation distance had a minor impact on the system's accuracy, as there was no significant difference between the average six-dimensional error and the gold standard (all P > 0.05). A significant impact on the positioning error of the system was observed when the observation angle and isocentric deviation were both set at 45 degrees (the vertical error was notably different from the gold standard, P < 0.05). The observation angle (0 degrees-30 degrees) exhibited a slight influence on the accuracy of the system (all P > 0.05). Patients with head and neck cancer exhibited a significant difference in positioning error between this system and the conventional method in the vertical direction (1.0 f 0.9 mm vs. 1.8 f 0.8 mm, P < 0.05), while no statistical significance was noted in other directions. Patients with chest and abdominal cancer showed significant discrepancies in the transformation between the two systems in lateral direction (1.9 f 1.0 mm vs. 2.5 f 0.7 mm, P < 0.05), vertical direction (1.4 f 1.2 mm vs. 2.2 f 1.1 mm, P < 0.05), and longitudinal direction (2.0 f 1.2 mm vs. 3.1 f 1.4 mm, P < 0.05), with no statistical significance found in rotation. Conclusion: In this study, a radiotherapy treatment positioning system utilizing mixed reality guidance is introduced. This system demonstrates the capability to notably decrease positioning errors in the transformation of patients with chest and abdominal cancer, as well as in the vertical direction for patients with head and neck cancer. Furthermore, it shows enhancements in the overall accuracy of radiotherapy procedures.
引用
收藏
页数:9
相关论文
共 30 条
  • [21] Augmented and Mixed Reality: Technologies for Enhancing the Future of IR
    Park, Brian J.
    Hunt, Stephen J.
    Martin, Charles, III
    Nadolski, Gregory J.
    Wood, Bradford J.
    Gade, Terence P.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 31 (07) : 1074 - 1082
  • [22] Evaluation of a surface imaging system's isocenter calibration methods
    Paxton, Adam B.
    Manger, Ryan P.
    Pawlicki, Todd
    Kim, Gwe-Ya
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02): : 85 - 91
  • [23] Sá AC, 2018, REP PRACT ONCOL RADI, V23, P61, DOI [10.1016/j.rpor.2017.12.007, 10.1016/j.rpor.7017.12.007]
  • [24] 3D Reconstruction and alignment by consumer RGB-D sensors and fiducial planar markers for patient positioning in radiation therapy
    Sarmadi, Hamid
    Munoz-Salinas, Rafael
    Alvaro Berbis, M.
    Luna, Antonio
    Medina-Carnicer, R.
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2019, 180
  • [25] A method for patient set-up guidance in radiotherapy using augmented reality
    Talbot, J.
    Meyer, J.
    Watts, R.
    Grasset, R.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (04) : 203 - 211
  • [26] Hologram stability evaluation for Microsoft® HoloLens™
    Vassallo, Reid
    Rankin, Adam
    Chen, Elvis C. S.
    Peters, Terry M.
    [J]. MEDICAL IMAGING 2017: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2017, 10136
  • [27] Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (Catalyst™)
    Walter, F.
    Freislederer, P.
    Belka, C.
    Heinz, C.
    Soehn, M.
    Roeder, F.
    [J]. RADIATION ONCOLOGY, 2016, 11
  • [28] Quality assurance for nonradiographic radiotherapy localization and positioning systems: Report of Task Group 147
    Willoughby, Twyla
    Lehmann, Joerg
    Bencomo, Jose A.
    Jani, Shirish K.
    Santanam, Lakshmi
    Sethi, Anil
    Solberg, Timothy D.
    Tome, Wolfgang A.
    Waldron, Timothy J.
    [J]. MEDICAL PHYSICS, 2012, 39 (04) : 1728 - 1747
  • [29] Accuracy Evaluation Trial of Mixed Reality-Guided Spinal Puncture Technology
    Wu, Jiajun
    Gao, Lei
    Shi, Qiao
    Qin, Chunhui
    Xu, Kai
    Jiang, Zhaoshun
    Zhang, Xixue
    Li, Ming
    Qiu, Jianjian
    Gu, Weidong
    [J]. THERAPEUTICS AND CLINICAL RISK MANAGEMENT, 2023, 19 : 599 - 609
  • [30] Initial clinical experience of surface guided stereotactic radiation therapy with open-face mask immobilization for improving setup accuracy: a retrospective study
    Zhou, Shun
    Li, Junyu
    Zhu, Xianggao
    Du, Yi
    Yu, Songmao
    Wang, Meijiao
    Yao, Kaining
    Wu, Hao
    Yue, Haizhen
    [J]. RADIATION ONCOLOGY, 2022, 17 (01)