Exploratory development and clinical research of a mixed reality guided radiotherapy positioning system

被引:0
作者
Zhang, Sai [1 ,2 ,3 ,4 ]
Zhang, Heng [1 ,2 ,3 ,4 ]
Li, Qixuan [1 ,2 ,3 ,4 ]
Wang, Ziyi [1 ,2 ,3 ,4 ]
Cao, Nannan [1 ,2 ,3 ,4 ]
Sun, Jiawei [1 ,2 ,3 ,4 ]
Xie, Kai [1 ,2 ,3 ,4 ]
Gao, Liugang [1 ,2 ,3 ,4 ]
Ni, Xinye [1 ,2 ,3 ,4 ]
机构
[1] Nanjing Med Univ, Affiliated Changzhou Peoples Hosp 2, Dept Radiotherapy, Changzhou 213003, Peoples R China
[2] Jiangsu Prov Engn Res Ctr Med Phys, Changzhou 213003, Peoples R China
[3] Nanjing Med Univ, Ctr Med Phys, Changzhou 213003, Peoples R China
[4] Key Lab Med Phys Changzhou, Changzhou 213003, Peoples R China
基金
中国国家自然科学基金;
关键词
Mixed reality; HoloLens2; Radiotherapy positioning; Virtual-real relation; AUGMENTED REALITY; RADIATION-THERAPY; IMMOBILIZATION; LOCALIZATION; ACCURACY; GUIDANCE; MOTION;
D O I
10.1016/j.jrras.2024.101095
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: This study aims to develop an application software deployed on HoloLens2 for guiding patient radiotherapy setup and assessing its feasibility and accuracy in radiotherapy. Methods: The developed system features an application software deployed on the HoloLens2 device, enabling automatic tracking and registration of a virtual cube to the location of a real calibration cube. This functionality is achieved through the utilization of the Vuforia SDK, which determines the location point linking the virtual and real spaces. The system then generates a reference hologram based on simulated positioning computed tomography (CT) data sets and anchors this reference hologram at the determined location point. In the process of treatment positioning, the real human body in the scene of HoloLens2 is matched with the surface of the reference hologram to complete the positioning. This paper explores the factors that affect the accuracy of the system, namely the viewing distance and Angle, to determine the optimal conditions for using the system. Then, based on the optimal conditions, the error between the cone beam computed tomography (CBCT) data of patients (head, neck and chest and abdomen) obtained by the system and conventional positioning were compared by Ttest method to evaluate the performance of the system. Results: The observation distance had a minor impact on the system's accuracy, as there was no significant difference between the average six-dimensional error and the gold standard (all P > 0.05). A significant impact on the positioning error of the system was observed when the observation angle and isocentric deviation were both set at 45 degrees (the vertical error was notably different from the gold standard, P < 0.05). The observation angle (0 degrees-30 degrees) exhibited a slight influence on the accuracy of the system (all P > 0.05). Patients with head and neck cancer exhibited a significant difference in positioning error between this system and the conventional method in the vertical direction (1.0 f 0.9 mm vs. 1.8 f 0.8 mm, P < 0.05), while no statistical significance was noted in other directions. Patients with chest and abdominal cancer showed significant discrepancies in the transformation between the two systems in lateral direction (1.9 f 1.0 mm vs. 2.5 f 0.7 mm, P < 0.05), vertical direction (1.4 f 1.2 mm vs. 2.2 f 1.1 mm, P < 0.05), and longitudinal direction (2.0 f 1.2 mm vs. 3.1 f 1.4 mm, P < 0.05), with no statistical significance found in rotation. Conclusion: In this study, a radiotherapy treatment positioning system utilizing mixed reality guidance is introduced. This system demonstrates the capability to notably decrease positioning errors in the transformation of patients with chest and abdominal cancer, as well as in the vertical direction for patients with head and neck cancer. Furthermore, it shows enhancements in the overall accuracy of radiotherapy procedures.
引用
收藏
页数:9
相关论文
共 30 条
  • [1] Dose calculation and treatment plan optimization including imaging dose from kilovoltage cone beam computed tomography
    Alaei, Parham
    Spezi, Emiliano
    Reynolds, Margaret
    [J]. ACTA ONCOLOGICA, 2014, 53 (06) : 839 - 844
  • [2] Surface guided radiation therapy: An international survey on current clinical practice
    Batista, V.
    Gober, M.
    Moura, F.
    Webster, A.
    Oellers, M.
    Ramtohul, M.
    Kuegele, M.
    Freislederer, P.
    Buschmann, M.
    Anastasi, G.
    Steiner, E.
    Al-Hallaq, H.
    Lehmann, J.
    [J]. TECHNICAL INNOVATIONS & PATIENT SUPPORT IN RADIATION ONCOLOGY, 2022, 22 : 1 - 8
  • [3] Influence of daily imaging on plan quality and normal tissue toxicity for prostate cancer radiotherapy
    Bell, Katharina
    Heitfeld, Marina
    Licht, Norbert
    Ruebe, Christian
    Dzierma, Yvonne
    [J]. RADIATION ONCOLOGY, 2017, 12
  • [4] Past, Present, and Future of Simultaneous Localization and Mapping: Toward the Robust-Perception Age
    Cadena, Cesar
    Carlone, Luca
    Carrillo, Henry
    Latif, Yasir
    Scaramuzza, Davide
    Neira, Jose
    Reid, Ian
    Leonard, John J.
    [J]. IEEE TRANSACTIONS ON ROBOTICS, 2016, 32 (06) : 1309 - 1332
  • [5] Visual Coherence in Mixed Reality: A Systematic Enquiry
    Collins, Jonny
    Regenbrecht, Holger
    Langlotz, Tobias
    [J]. PRESENCE-TELEOPERATORS AND VIRTUAL ENVIRONMENTS, 2017, 26 (01) : 16 - 41
  • [6] RAD-AR: RADiotherapy - Augmented Reality An augmented reality tool for radiotherapy implemented on consumer electronics devices
    Cosentino, Francesco
    John, Nigel W.
    Vaarkamp, Jaap
    [J]. 2017 INTERNATIONAL CONFERENCE ON CYBERWORLDS (CW), 2017, : 226 - 228
  • [7] Augmented reality visualisation for orthopaedic surgical guidance with pre- and intra-operative multimodal image data fusion
    El-Hariri, Houssam
    Pandey, Prashant
    Hodgson, Antony J.
    Garbi, Rafeef
    [J]. HEALTHCARE TECHNOLOGY LETTERS, 2018, 5 (05): : 189 - 193
  • [8] Augmented Reality in Neurosurgery: A Review of Current Concepts and Emerging Applications
    Guha, Daipayan
    Alotaibi, Naif M.
    Nhu Nguyen
    Gupta, Shaurya
    McFaul, Christopher
    Yang, Victor X. D.
    [J]. CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 2017, 44 (03) : 235 - 245
  • [9] Use of surface-guided radiation therapy in combination with IGRT for setup and intrafraction motion monitoring during stereotactic body radiation therapy treatments of the lung and abdomen
    Heinzerling, John H.
    Hampton, Carnell J.
    Robinson, Myra
    Bright, Megan
    Moeller, Benjamin J.
    Ruiz, Justin
    Prabhu, Roshan
    Burri, Stuart H.
    Foster, Ryan D.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (05): : 48 - 55
  • [10] Haptic Augmented Reality: Taxonomy and an Example of Stiffness Modulation
    Jeon, Seokhee
    Choi, Seungmoon
    [J]. PRESENCE-TELEOPERATORS AND VIRTUAL ENVIRONMENTS, 2009, 18 (05) : 387 - 408