INTRAMEDULLARY NAIL OR LOCKING COMPRESSION PLATE FOR FIXING A FRACTURED DISTAL TIBIA: FINITE ELEMENT ANALYSIS ALONG WITH AN ADAPTATION MODEL

被引:1
作者
Moslehi, Amir Abbas [1 ]
Rouhi, Gholamreza [1 ]
机构
[1] Amirkabir Univ Technol, Dept Biomed Engn, Tehran, Iran
关键词
distal tibial fractures; intramedullary nailing; plating; finite element method; bone density distribution; bone remodeling theory; von Mises stress; COMPUTED-TOMOGRAPHY; BONE LOSS; FIXATION; STRESS; OSTEOSYNTHESIS; PROTECTION; DENSITY;
D O I
10.1615/IntJMultCompEng.2023048159
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Intramedullary (IM) nailing and plating are two major options for the treatment of distal tibial fractures; each method has its own pros and cons. Bone loss due to IM nailing or plating can be a serious concern, which may lead to the premature failure of the implants. Thus, investigating bone remodeling in response to orthopedic implantation is of paramount importance for the purpose of designing more efficient prostheses. The main goal of this study was to predict the effects of nailing and plating on the bone apparent density distribution in tibia, through use of the finite element method (FEM) along with a bone remodeling model. To this end, three-dimensional geometry of tibia, with properties derived from computer tomography (CT) imaging data, IM nail and locking compression plate (LCP) were created. Then, a bone remodeling model and FEM, with physiological -like loading, were combined and implemented using Python programming language. The results showed that both nailing and plating methods cause a reduction in bone apparent density, but the plate affects just the distal part of the tibia, where the implant is present; whereas the IM nail affects almost the whole tibia. Moreover, in agreement with clinical observations, it was found here that the stress shielding effects of the LCP are less pronounced than that of IM nail, in the long term. Stress analysis revealed that the load, to most extent, transferred to IM nail; whereas it transferred to LCP just in the distal part of tibia. On the basis of the results of this work, it might be concluded that plating is a better choice than nailing for distal tibial fracture, from the bone remodeling point of view.
引用
收藏
页码:1 / 16
页数:16
相关论文
共 55 条
[41]   Prediction of Stress Shielding Around Orthopedic Screws: Time-Dependent Bone Remodeling Analysis Using Finite Element Approach [J].
Rouhi, Gholamreza ;
Tahani, Masoud ;
Haghighi, Behnoud ;
Herzog, Walter .
JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 35 (04) :545-554
[42]   Intramedullary nailing versus percutaneous locked plating of distal extra-articular tibial fractures: A retrospective study [J].
Seyhan M. ;
Unay K. ;
Sener N. .
European Journal of Orthopaedic Surgery & Traumatology, 2013, 23 (5) :595-601
[43]   Numerical Simulation of Mandible Bone Remodeling under Tooth Loading: A Parametric Study [J].
Su, Kangning ;
Yuan, Li ;
Yang, Jie ;
Du, Jing .
SCIENTIFIC REPORTS, 2019, 9 (1)
[44]   Fracture on removal of the ACE tibial nail [J].
Takakuwa, M ;
Funakoshi, M ;
Ishizaki, K ;
Aono, T ;
Hamaguchi, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (03) :444-445
[45]   BONE ATROPHY AFTER PLATE FIXATION - COMPUTED-TOMOGRAPHY OF FEMORAL-SHAFT FRACTURES [J].
TERJESEN, T ;
NORDBY, A ;
ARNULF, V .
ACTA ORTHOPAEDICA SCANDINAVICA, 1985, 56 (05) :416-418
[46]  
TERJESEN T, 1986, CLIN ORTHOP RELAT R, P108
[47]   THE STRESS-PROTECTING EFFECT OF METAL PLATES ON THE INTACT RABBIT TIBIA [J].
TERJESEN, T ;
BENUM, P .
ACTA ORTHOPAEDICA SCANDINAVICA, 1983, 54 (06) :810-818
[48]   PROTECTION FROM STRESS IN BONE AND ITS EFFECTS - EXPERIMENTS WITH STAINLESS-STEEL AND PLASTIC PLATES IN DOGS [J].
TONINO, AJ ;
DAVIDSON, CL ;
KLOPPER, PJ ;
LINCLAU, LA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1976, 58 (01) :107-113
[49]  
Tytler R., 2002, Australian Science Teachers' Journal, V48, P30
[50]   2 PROCESSES OF BONE REMODELING IN PLATED INTACT FEMORA - AN EXPERIMENTAL-STUDY IN DOGS [J].
UHTHOFF, HK ;
FOUX, A ;
YEADON, A ;
MCAULEY, J ;
BLACK, RC .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1993, 11 (01) :78-91