Accuracy of dynamic computer-assisted implant surgery in fully edentulous patients: An in vitro study

被引:1
|
作者
Ruiz-Romero, Victor [1 ]
Jorba-Garcia, Adria [1 ]
Camps-Font, Octavi [2 ,3 ]
Figueiredo, Rui [2 ,3 ]
Valmaseda-Castellon, Eduard [2 ,3 ]
机构
[1] Univ Barcelona, Fac Med & Hlth Sci, Oral Surg & Implantol, Barcelona, Spain
[2] Univ Barcelona, Fac Med & Hlth Sci, Oral Surg, Barcelona, Spain
[3] IDIBELL Inst, Barcelona, Spain
关键词
Computer-assisted surgery; Dental Implants; Surgical navigation systems; Edentulous jaw; Implant-supported dental prosthesis; Dimensional measurement accuracy; COMPLETE DENTAL PROSTHESES; SURVIVAL RATES; NAVIGATION; PLACEMENT; COMPLICATIONS;
D O I
10.1016/j.jdent.2024.105290
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To compare miniscrew versus bone tracing registration methods on dental implant placement accuracy and time efficiency in edentulous jaws using a dynamic computer-assisted implant surgery (d-CAIS) system. Methods: Twelve fully edentulous maxillary models were allocated into two groups: miniscrew tracing (MST) group, where registration was performed by tracing four miniscrews; and bone tracing (BT) group, where registration was conducted by tracing maxillary bone fiducial landmarks. Six implants were placed on each model using the X-Guide (R) d-CAIS system. Pre- and postoperative cone-beam computed tomography (CBCT) scans were superimposed to evaluate implant placement accuracy. The time required for registration and the overall surgery time were also recorded. Results: Thirty-six implants were placed in each group. The MST group showed significantly lower mean angulation deviations (mean difference (MD): -3.33 degrees; 95 % confidence interval (CI): -6.56 to -0.09); p = 0.044), 3D platform deviations (MD: -1.01 mm; 95 % CI: -1.74 to -0.29; p = 0.006), 2D platform deviations (MD: -0.97 mm; 95 % CI: -1.71 to -0.23; p = 0.010), and 3D apex deviations (MD: -1.18 mm; 95 % CI: -1.92 to -0.44; p = 0.002) versus the BT group. The overall surgery time was similar for both groups (MD: 6.10 min.; 95 % CI: -0.31 to 12.51; p = 0.06), though bone tracing required significantly more time compared with miniscrew registration (MD: 4.79 min.; 95 % CI: 2.96 to 6.62; p < 0.05). Conclusions: Registration with MST increases the accuracy of implant placement with a d-CAIS system in edentulous jaws compared with the BT method, and slightly reduces the overall surgery time.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Are Sleeves Necessary in Static Computer-Assisted Implant Surgery? A Comparative In Vitro Analysis
    Hang, Jenna
    Guentsch, Arndt
    CLINICAL ORAL IMPLANTS RESEARCH, 2025, 36 (01) : 117 - 126
  • [42] Accuracy of autonomous robotic surgery for dental implant placement in fully edentulous patients: A retrospective case series study
    Li, Ping
    Chen, Jiahao
    Li, An
    Luo, Ke
    Xu, Shulan
    Yang, Shuo
    CLINICAL ORAL IMPLANTS RESEARCH, 2023, 34 (12) : 1428 - 1437
  • [43] Accuracy of computer-assisted implant placement with insertion templates
    Naziri, Eleni
    Schramm, Alexander
    Wilde, Frank
    GMS INTERDISCIPLINARY PLASTIC AND RECONSTRUCTIVE SURGERY DGPW, 2016, 5
  • [44] Accuracy of immediate anterior implantation using static and robotic computer-assisted implant surgery: A retrospective study
    Li, Jun
    Dai, Meng
    Wang, Siwei
    Zhang, Xitao
    Fan, Qin
    Chen, Lin
    JOURNAL OF DENTISTRY, 2024, 148
  • [45] Accuracy of implant placement with a combined use of static and dynamic computer-assisted implant surgery in single tooth space: A randomized controlled trial
    Yotpibulwong, Thanida
    Arunjaroensuk, Sirida
    Kaboosaya, Boosana
    Sinpitaksakul, Phonkit
    Arksornnukit, Mansuang
    Mattheos, Nikos
    Pimkhaokham, Atiphan
    CLINICAL ORAL IMPLANTS RESEARCH, 2023, 34 (04) : 330 - 341
  • [46] Accuracy of computer-assisted, template-guided implant placement compared with conventional implant placement by hand-An in vitro study
    Schneider, David
    Sax, Caroline
    Sancho-Puchades, Manuel
    Hammerle, Christoph H. F.
    Jung, Ronald Ernst
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (09) : 1052 - 1060
  • [47] Computer-assisted surgery
    Gebhard, F
    Kinzl, L
    Arand, M
    UNFALLCHIRURG, 2000, 103 (08): : 612 - 617
  • [48] A Prospective Study on Accuracy of Computer-Based Fully Guided Versus Pilot-Guided Implant Surgery
    Kim, Hyo Joon
    Kim, Hee Jin
    Moon, Seong Yong
    APPLIED SCIENCES-BASEL, 2020, 10 (06):
  • [49] Effect of the number and distribution of fiducial markers on the accuracy of robot-guided implant surgery in edentulous mandibular arches: An in vitro study
    Zhao, Yaoyu
    Liao, Yilin
    Wu, Xinyu
    Zhang, Yufeng
    Shi, Bin
    Yan, Qi
    JOURNAL OF DENTISTRY, 2023, 134
  • [50] Computer-Assisted Surgery in the Edentulous Jaw Based on 3 Fixed Intraoral Reference Points
    Widmann, Gerlig
    Keiler, Martin
    Zangerl, Antoniette
    Stoffner, Rudolf
    Longato, Stefano
    Bale, Reto
    Puelacher, Wolfgang
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2010, 68 (05) : 1140 - 1147