A phantom for 2D dose measurements in the vicinity of metal implants for photon and proton beams

被引:0
作者
Lechner, Wolfgang [1 ,2 ,3 ]
Knaeusl, Barbara [1 ,2 ,3 ]
Brunner, Jacob [1 ,3 ]
Georg, Dietmar [1 ,2 ,3 ]
Kuess, Peter [1 ,2 ,3 ]
机构
[1] Med Univ Vienna, Dept Radiat Oncol, Div Med Phys, Vienna, Austria
[2] Med Univ Vienna, Dept Radiat Oncol, Christian Doppler Lab Image & Knowledge Driven Pre, Vienna, Austria
[3] MedAustron Ion Therapy Ctr, Med Phys, Wiener Neustadt, Austria
来源
FRONTIERS IN PHYSICS | 2024年 / 12卷
基金
欧盟地平线“2020”;
关键词
proton and photon radiotherapy; radiation oncology; dosimetry; phantom design; metal implants; dose calculation accuracy; ARTIFACT REDUCTION; RADIOTHERAPY; THERAPY; HETEROGENEITY; VALIDATION; AUDIT; FILM;
D O I
10.3389/fphy.2024.1433208
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
This work aimed to characterize a dedicated phantom for assessing the dose near metal implants for radiotherapy with photons and protons. A dosimetry audit phantom was redesigned to position a Gafchromic EBT-3 film within a bisected titanium pedicle screw (6.5 mm diameter). The mass density and the water equivalent thickness (WET) of the phantom material were determined. The phantom was irradiated using a photon arc and a horizontal proton beam in combination with a couch rotation of 20 degrees, with three repeated measurements each. Treatment plans utilizing a single field covering the screw and the EBT-3 film were optimized to deliver a physical dose of 2 Gy using a collapsed cone and Monte Carlo dose engine for photons and protons, respectively. The mass density and the WET of the phantom were determined as (1.033 +/- 0.010) g cm(-3) and (1.022 +/- 0.013), respectively. Ionisation chamber measurements agreed within 1% (photons) and 0.5% (protons) with the calculated dose values. Relative photon dosimetry measurements using EBT-3 films revealed an agreement between measured and calculated horizontal profiles within the confidence interval for areas beyond 5 mm from the center. For photon plans, significant deviations of more than 10% were found at the interfaces between phantom material and screw. The proton measurements showed a gradual decrease of 3% across both profiles. In contrast to photon plans, no dose increase was measured within the screw, but significant dose fluctuations ( > 5%) in the beam's exit region. This study showed that the behavior of dose engine is affected by metal implants and thus dosimetric measurements are highly recommended. The presented phantom can serve as foundation for dedicated end-2-end phantoms.
引用
收藏
页数:12
相关论文
共 39 条
  • [1] [Anonymous], 2017, Dosimetry of Small Static Fields Used in Radiotherapy: An International Code of Practice for Reference and Relative Dose Determination Technical Report Series 484
  • [2] Dosimetric characteristics of 3D-printed and epoxy-based materials for particle therapy phantoms
    Brunner, Jacob
    Langgartner, Lorenz
    Danhel, Hannah
    Birkfellner, Wolfgang
    Richter, Christian
    Wagenaar, Dirk
    Stock, Markus
    Georg, Dietmar
    Knaeusl, Barbara
    [J]. FRONTIERS IN PHYSICS, 2024, 12
  • [3] Clinical validation of a GPU-based Monte Carlo dose engine of a commercial treatment planning system for pencil beam scanning proton therapy
    Fracchiolla, Francesco
    Engwall, Erik
    Janson, Martin
    Tamm, Fredrik
    Lorentini, Stefano
    Fellin, Francesco
    Bertolini, Mattia
    Algranati, Carlo
    Righetto, Roberto
    Farace, Paolo
    Amichetti, Maurizio
    Schwarz, Marco
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 226 - 234
  • [4] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08) : R49 - R80
  • [5] Effects of metal implants and validation of four treatment planning methods used for radiotherapy dose calculation
    Gnanasambandam, Amirtharaj
    Raj, N. Arunai Nambi
    Sollinselvan, K.
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (05) : 821 - 831
  • [6] Investigation of the Bragg peak degradation caused by homogeneous and heterogeneous lung tissue substitutes: proton beam experiments and comparison to current clinical dose calculation
    Hranek, A.
    Resch, A. F.
    Georg, D.
    Knaeusl, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (24)
  • [7] Approaches to reducing photon dose calculation errors near metal implants
    Huang, Jessie Y.
    Followill, David S.
    Howell, Rebecca M.
    Liu, Xinming
    Mirkovic, Dragan
    Stingo, Francesco C.
    Kry, Stephen F.
    [J]. MEDICAL PHYSICS, 2016, 43 (09) : 5117 - 5130
  • [8] IAEA, 2024, Absorbed dose determination in external beam radiotherapy, DOI [10.61092/iaea.ve7q-y94k, DOI 10.61092/IAEA.VE7Q-Y94K]
  • [9] Global availability of dosimetry audits in radiotherapy: The IAEA dosimetry audit networks database
    Izewska, Joanna
    Lechner, Wolfgang
    Wesolowska, Paulina
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 5 : 1 - 4
  • [10] Testing the methodology for dosimetry audit of heterogeneity corrections and small MLC-shaped fields: Results of IAEA multi-center studies
    Izewska, Joanna
    Wesolowska, Paulina
    Azangwe, Godfrey
    Followill, David S.
    Thwaites, David I.
    Arib, Mehenna
    Stefanic, Amalia
    Viegas, Claudio
    Luo Suming
    Ekendahl, Daniela
    Bulski, Wojciech
    Georg, Dietmar
    [J]. ACTA ONCOLOGICA, 2016, 55 (07) : 909 - 916