Biomechanical Comparison of Fixation Methods for Posterior Wall Fractures of the Acetabulum: Conventional Reconstruction Plate vs. Spring Plate vs. Variable Angle Locking Compression Plate

被引:0
|
作者
Chung, Hoejeong [1 ]
Sohn, Hoon-Sang [1 ]
Oh, Jong-Keon [2 ]
Lee, Sangho [1 ]
Kim, Doosup [1 ]
机构
[1] Yonsei Univ, Wonju Severance Christian Hosp, Wonju Coll Med, Dept Orthoped Surg, 20 Ilsan Ro, Wonju 26426, South Korea
[2] Korea Univ, Korea Univ Med, Guro Hosp, Dept Orthopaed Surg, 148 Gurodong Ro, Seoul 08308, South Korea
来源
MEDICINA-LITHUANIA | 2024年 / 60卷 / 06期
基金
新加坡国家研究基金会;
关键词
biomechanics; acetabulum; posterior wall fracture; variable angle plate; INTERNAL-FIXATION; HIP;
D O I
10.3390/medicina60060882
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives: Acetabular fractures, though infrequent, present considerable challenges in treatment due to their association with high-energy trauma and poor prognoses. Posterior wall fractures, the most common type among them, typically have a more favorable prognosis compared to other types. Anatomical reduction and stable fixation of the posterior wall are crucial for optimal treatment outcomes. This study aimed to biomechanically compare three commonly used fixation methods for posterior wall fractures of the acetabulum-a conventional reconstruction plate, a spring plate, and a 2.7 mm variable angle locking compression plate (VA-LCP). Materials and Methods: The study utilized 6 fresh-frozen cadavers, yielding 12 hemipelvises free from prior trauma or surgery. Three fixation methods were compared using a simple acetabulum posterior wall fracture model. Fixation was performed by an orthopedic specialist, with prebending of plates to minimize errors. Hemipelvises were subjected to quasi-static and cyclic loading tests, measuring fracture gap, stiffness, and displacement under load. Results: It showed no significant differences in fracture gap among the three fixation methods under cyclic loading conditions simulating walking. However, the conventional reconstruction plate exhibited a greater stiffness compared to the spring and variable angle plates. Fatigue analysis revealed no significant differences among the plates, indicating a similar stability throughout cyclic loading. Despite differences in stiffness, all three fixation methods demonstrated adequate stability under loading conditions. Conclusions: While the conventional reconstruction plate demonstrated a superior stiffness, all three fixation methods provided sufficient stability under cyclic loading conditions similar to walking. This suggests that postoperative limitations are unlikely with any of the three methods, provided excessive activities are avoided. Furthermore, the variable angle plate-like the spring plate-offers an appropriate stability for fragment-specific fixation, supporting its use in surgical applications. These findings contribute to understanding the biomechanical performance of different fixation methods for acetabular fractures, facilitating improved surgical outcomes in challenging cases.
引用
收藏
页数:12
相关论文
共 45 条
  • [1] Precontoured buttress plate vs reconstruction plate for acetabulum posterior wall fractures: A biomechanical study
    Altun, Guray
    Saka, Gursel
    Demir, Teyfik
    Elibol, Fatma Kubra Erbay
    Polat, Mehmet Orcun
    WORLD JOURNAL OF ORTHOPEDICS, 2019, 10 (05): : 219 - 227
  • [2] Comparison of Fixation between Dynamic Compression Plate Vs Locking Compression Plate in Lower Limb Diaphyseal Fractures by Bridge Plating Technique
    Aziz, Shaheryar
    Ahmed, Faheem
    Aslam, Muhammad Usman
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (04): : 1011 - 1014
  • [3] Efficacy of Hemiarthroplasty vs. Locking Plate Fixation for Proximal Humerus Fractures: A Meta-Analysis
    Deng, Jiali
    Zhang, Shuai
    Yu, Yuanyuan
    Zhang, Li
    Zhang, Li
    Jiang, Wen
    Yang, Kai
    Xi, Xiaoyan
    FRONTIERS IN SURGERY, 2021, 8
  • [4] Which implant is better for the fixation of posterior wall acetabular fractures: A conventional reconstruction plate or a brand-new calcaneal plate?
    Tosyali, Hakan Koray
    Elibol, Fatma Kubra Erbay
    Hancioglu, Sertan
    Kacmaz, Suat Emre
    Caliskan, Ozde Irem
    Tolunay, Tolga
    Demir, Teyfik
    Oklu, Guvenir
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2024, 55 (04):
  • [5] Locking Compression Pilon Plate for Fixation of Comminuted Posterior Wall Acetabular Fractures: A Novel Technique
    Lee, Adam K.
    Wagner, Benjamin R.
    McPhillips, Kristin
    Horwitz, Daniel S.
    Widmaier, James C., Jr.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (01) : E32 - E36
  • [6] Polyaxial locking plate fixation in distal femur fractures: A biomechanical comparison
    Wilkens, Kenneth J.
    Curtiss, Shane
    Lee, Mark A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (09) : 624 - 628
  • [7] Single column plate plus other column lag screw fixation vs. both column plate fixation for anterior column with posterior hemitransverse acetabular fractures - a biomechanical analysis using different loading protocols
    Quang, Huy Le
    Schmoelz, Werner
    Lindtner, Richard A.
    Dammerer, Dietmar
    Schwendinger, Peter
    Krappinger, Dietmar
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2021, 52 (04): : 699 - 704
  • [8] Comparative Study of W-Shaped Angular Plate and Reconstruction Plate in Treating Posterior Wall Fractures of the Acetabulum
    Zhang, Qi
    Chen, Wei
    Wu, Xiaobo
    Su, Yanling
    Hou, Zhiyong
    Zhang, Yingze
    PLOS ONE, 2014, 9 (03):
  • [9] A Biomechanical Comparison of a Locking Plate, a Nail, and a 95° Angled Blade Plate for Fixation of Subtrochanteric Femoral Fractures
    Forward, Daren P.
    Doro, Christopher J.
    O'Toole, Robert V.
    Kim, Hyunchul
    Floyd, John C. P.
    Sciadini, Marcus F.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (06) : 334 - 340
  • [10] A Biomechanical Comparison of Locking Versus Conventional Plate Fixation for Distal Fibula Fractures in Trimalleolar Ankle Injuries
    Nguyentat, Annie
    Camisa, William
    Patel, Sandeep
    Lagaay, Pieter
    JOURNAL OF FOOT & ANKLE SURGERY, 2016, 55 (01) : 132 - 135