Purpose To investigate the surgical outcomes of intrascleral intraocular lens (IOL) fixation using a modified extraocular forceps-guided technique. Study design Retrospective case series. Methods Overall, 81 eyes of 78 patients who underwent intrascleral IOL fixation using the modified extraocular forceps-guided technique were included. The procedure entailed creating 2 scleral half-layer T-shaped incisions perpendicular to the main incision and forming a scleral tunnel. A 25-gauge trocar was inserted at the lower end of the T-shaped incision to perform vitrectomy. A 27-gauge needle was inserted from the left-hand port, and the leading haptic was inserted into the needle lumen. After removal of the right-hand trocar, a 90 degrees-curved intrascleral fixation forceps was inserted into the eye, exposing the tip at the main incision, thus allowing the tip of the extraocular trailing haptic to be gripped and both haptics to be pulled out. The left-hand trocar was removed, and the haptics were buried in the scleral tunnel. The surgical outcomes of this technique were retrospectively evaluated on the basis of the medical records. Results The induction of haptics was successful in all cases. The preoperative best-corrected visual acuity improved from 0.35 +/- 0.68 to 0.12 +/- 0.36 logMAR postoperatively (P<0.01). The refractive error was -0.27 +/- 0.87 D; IOL decentration, 0.39 +/- 0.18 mm; IOL tilt, 5.97 +/- 2.65 degrees; IOL astigmatism, 0.35 +/- 0.36 D; and corneal endothelial cell loss, 10.3 +/- 12.7%. There were no serious complications related to the surgical technique. Conclusion The modified extraocular forceps-guided technique allows for safe and straightforward induction of the trailing haptics and enables the performance of intrascleral IOL fixation with minimal scleral incisions.
引用
收藏
页码:538 / 547
页数:10
相关论文
共 31 条
  • [11] Kim DB., 2017, CATARACT REFRACT SUR, V18, P27
  • [12] Glued intraocular lens in eyes with deficient capsules: retrospective analysis of long-term effects
    Kumar, Dhivya Ashok
    Agarwal, Amar
    Dhawan, Atul
    Thambusamy, Vijay Anand
    Sivangnanam, Soundari
    Venktesh, Triveni
    Chandrasekar, Radhika
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (04) : 496 - 503
  • [13] Extraocular Technique of Intrascleral Intraocular Lens Fixation Using a Pair of the Shaft-Bended 27-Gauge Needles
    Maruko, Ichiro
    Koizumi, Hideki
    Kogure-Katakura, Akiko
    Iida, Tomohiro
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2017, 37 (01): : 191 - 193
  • [14] Influence of sclerotomy size on intraocular lens tilt after intrascleral intraocular lens fixation
    Matsumura, Takehiro
    Takamura, Yoshihiro
    Makita, Jun
    Kobori, Akira
    Inatani, Masaru
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (10) : 1446 - 1451
  • [15] Clinical Outcomes of Endoscope-Assisted 30-Gauge Single-Needle Technique for Intrascleral Intraocular Lens Fixation
    Mitamura, Hayato
    Kaga, Tatsushi
    Kojima, Takashi
    Yoshida, Norihiko
    Sato, Hiroyuki
    Matsuda, Taisuke
    Yokoyama, Sho
    Mori, Toshio
    Ichikawa, Kazuo
    [J]. OPHTHALMIC RESEARCH, 2021, 64 (02) : 253 - 260
  • [16] A Modified Intrascleral Intraocular Lens Fixation Technique Using 27-Gauge Blunted Needles with Fewer Intraocular Manipulations
    Morino, Kazuya
    Iida, Yuto
    Akimoto, Masayuki
    [J]. JOURNAL OF OPHTHALMOLOGY, 2021, 2021
  • [17] Relationship between Postoperative Anterior Chamber Depth and Refraction Based on the Haptic Fix Position in Intraocular Lens Intrascleral Fixation
    Noguchi, Santaro
    Nakakura, Shunsuke
    Noguchi, Asuka
    Tabuchi, Hitoshi
    [J]. JOURNAL OF CLINICAL MEDICINE, 2023, 12 (05)
  • [18] Ohta T., 2015, JPN J CATARACT REFRA, V29, P162
  • [19] Simplified and safe method of sutureless intrascleral posterior chamber intraocular lens fixation: Y-fixation technique
    Ohta, Toshihiko
    Toshida, Hiroshi
    Murakami, Akira
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01) : 2 - 7
  • [20] Haptic Erosion Following Sutureless Scleral-fixated Intraocular Lens Placement
    Pakravan, Parastou
    Patel, Veshesh
    Chau, Viet
    Rohowetz, Landon
    Lai, James
    Fan, Kenneth C.
    Al-khersan, Hasenin
    Melo, Isabela M.
    Muni, Rajeev H.
    Tsao, Sean W.
    Kaplan, Richard
    Jung, Jesse J.
    Hoyek, Sandra
    Patel, Nimesh A.
    Kuriyan, Ajay E.
    Laura, Diana M.
    Mantopoulos, Dimosthenis
    Syed, Zeba A.
    Yannuzzi, Nicolas A.
    [J]. OPHTHALMOLOGY RETINA, 2023, 7 (04): : 333 - 337