Evaluation of the conformity of intensity-modulated radiation therapy and volumetric modulated arc therapy using AAPM TG 119 protocol

被引:0
|
作者
Tam, Dang Thi Minh [1 ]
Ho, Phan Long [2 ,3 ,4 ]
Uy, Phan Quoc [1 ,2 ,3 ]
Hieu, Nguyen Trung [1 ]
Linh, Vo Tan [1 ]
Hoa, Nguyen Thi [1 ]
Lam, Nguyen Thi The [1 ]
Nga, Bui Thi Thuy [1 ]
Thanh, Truong Huu [1 ]
Thanh, Tran Thien [2 ,3 ]
Tao, Chau Van [2 ,3 ]
机构
[1] Ho Chi Minh Oncol Hosp, Dept Radiol Technol, Ho Chi Minh, Vietnam
[2] Univ Sci, Fac Phys & Engn Phys, Dept Nucl Phys, 227,Nguyen Van Cu St,Dist 5, Ho Chi Minh, Vietnam
[3] Vietnam Natl Univ, Ho Chi Minh, Vietnam
[4] Inst Publ Hlth Ho Chi Minh City, Ho Chi Minh, Vietnam
关键词
IMRT; VMAT; TG; 119; protocol; TrueBeam linac; Semiflex ion chamber; EPID; Quality assurance; IMRT; VERIFICATION; RADIOTHERAPY; IRRADIATION; CARCINOMA; SYSTEM; VMAT;
D O I
10.1007/s00411-024-01091-2
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The aim of this work was to evaluate the conformity of intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT), and verify the accuracy of the planning and delivery system used in this work based on the AAPM TG-119 protocol. The Eclipse 13.6 treatment planning system (TPS) was used to plan the TG-119 test suite, which included four test cases: MultiTarget, Prostate, Head/Neck, and C-Shape for IMRT and VMAT techniques with 6 MV and 10 MV acceleration voltages. The results were assessed and discussed in terms of the TG-119 protocol and the results of previous studies. In addition, point dose and planar dose measurements were done using a semiflex ion chamber and an electronic portal imaging device (EPID), respectively. The planned doses of all test cases met the criteria of the TG-119 protocol, except those for the spinal cord of the C-Shape hard case. There were no significant differences between the treatment planning doses and the doses given in the TG-119 report, with p-values ranging from 0.974 to 1 (p > 0.05). Doses to the target volumes were similar in the IMRT and VMAT plans, but the organs at risk (OARs) doses were different depending on the test case. The planning results showed that IMRT is more conformal than VMAT in certain cases. For the point dose measurements, the confidence limit (CLpoint) of 0.030 and 0.021 were better than the corresponding values of 0.045 and 0.047 given in the TG-119 report for high-dose and low-dose areas, respectively. Regarding the planar dose measurements, the CLplanar value of 0.38 obtained in this work was lower than that given in the TG-119 report (12.4). It is concluded that the dosimetry measurements performed in this study showed better confidence limits than those provided in the TG 119 report. IMRT remains more conformal in certain circumstances than the more progressive VMAT. When selecting the method of delivering a dose to the patient, several factors must be considered, including the radiotherapy technique, energy, treatment site, and tumour geometry.
引用
收藏
页码:557 / 571
页数:15
相关论文
共 50 条
  • [21] Dosimetric effects of weight loss or gain during volumetric modulated arc therapy and intensity-modulated radiation therapy for prostate cancer
    Pair, Matthew L.
    Du, Weiliang
    Rojas, Hector D.
    Kanke, James E.
    McGuire, Sean E.
    Lee, Andrew K.
    Kuban, Deborah A.
    Kudchadker, Rajat J.
    MEDICAL DOSIMETRY, 2013, 38 (03) : 251 - 254
  • [22] Impact of machines on plan quality: volumetric modulated arc therapy and intensity modulated radiation therapy
    S. Clemente
    M. Cozzolino
    C. Oliviero
    A. Fiorentino
    C. Chiumento
    V. Fusco
    Clinical and Translational Oncology, 2014, 16 : 141 - 146
  • [23] Quality of treatment plans and accuracy of in vivo portal dosimetry in hybrid intensity-modulated radiation therapy and volumetric modulated arc therapy for prostate cancer
    Bedford, James L.
    Smyth, Gregory
    Hanson, Ian M.
    Tree, Alison C.
    Dearnaley, David P.
    Hansen, Vibeke N.
    RADIOTHERAPY AND ONCOLOGY, 2016, 120 (02) : 320 - 326
  • [24] Feasibility of a unified approach to intensity-modulated radiation therapy and volume-modulated arc therapy optimization and delivery
    Hoover, Douglas A.
    MacFarlane, Michael
    Wong, Eugene
    Battista, Jerry J.
    Chen, Jeff Z.
    MEDICAL PHYSICS, 2015, 42 (02) : 726 - 734
  • [25] Volumetric modulated arc therapy versus intensity-modulated proton therapy in the postoperative irradiation of thymoma
    Davide Franceschini
    Luca Cozzi
    Mauro Loi
    Ciro Franzese
    Giacomo Reggiori
    Pietro Mancosu
    Alessandro Clivio
    Antonella Fogliata
    Marta Scorsetti
    Journal of Cancer Research and Clinical Oncology, 2020, 146 : 2267 - 2276
  • [26] Volumetric Modulated Arc Therapy and Conventional Intensity-modulated Radiotherapy for Simultaneous Maximal Intraprostatic Boost: a Planning Comparison Study
    Shaffer, R.
    Morris, W. J.
    Moiseenko, V.
    Welsh, M.
    Crumley, C.
    Nakano, S.
    Schmuland, M.
    Pickles, T.
    Otto, K.
    CLINICAL ONCOLOGY, 2009, 21 (05) : 401 - 407
  • [27] Dosimetric comparison of hybrid volumetric-modulated arc therapy, volumetric-modulated arc therapy, and intensity-modulated radiation therapy for left-sided early breast cancer
    Lin, Jia-Fu
    Yeh, Dah-Cherng
    Yeh, Hui-Ling
    Chang, Chen-Fa
    Lin, Jin-Ching
    MEDICAL DOSIMETRY, 2015, 40 (03) : 262 - 267
  • [28] Treatment Efficiency of Volumetric Modulated Arc Therapy in Comparison With Intensity-Modulated Radiotherapy in the Treatment of Prostate Cancer
    Hall, William A.
    Fox, Timothy H.
    Jiang, Xiaojun
    Prabhu, Roshan S.
    Rossi, Peter J.
    Godette, Karen
    Jani, Ashesh B.
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2013, 10 (02) : 128 - 134
  • [29] Comparison of three-dimensional conformal radiation therapy, intensity-modulated radiation therapy, and volumetric-modulated arc therapy in the treatment of cervical esophageal carcinoma
    Yang, Hao
    Feng, Cong
    Cai, Bo-Ning
    Yang, Jun
    Liu, Hai-Xia
    Ma, Lin
    DISEASES OF THE ESOPHAGUS, 2017, 30 (02):
  • [30] Predictive factors for acute radiation pneumonitis in postoperative intensity modulated radiation therapy and volumetric modulated arc therapy of esophageal cancer
    Zhao, Yaqin
    Chen, Lu
    Zhang, Shu
    Wu, Qiang
    Jiang, Xiaoqin
    Zhu, Hong
    Wang, Jin
    Li, Zhiping
    Xu, Yong
    Zhang, Ying Jie
    Bai, Sen
    Xu, Feng
    THORACIC CANCER, 2015, 6 (01) : 49 - 57