X marks the spot: catheter aspiration using the Inari FlowTriever device to debulk defibrillator lead vegetations prior to transvenous lead extraction-a case report

被引:0
作者
Clark, James [1 ]
Zaidi, Abbas [1 ]
O'Callaghan, Peter [1 ]
von Oppell, Ulrich [2 ]
Sharp, Andrew S. P. [1 ,3 ]
机构
[1] Univ Hosp Wales, Dept Cardiol, Cardiff CF14 4XW, Wales
[2] Univ Hosp Wales, Dept Cardiothorac Surg, Cardiff CF14 4XW, Wales
[3] Cardiff Univ, Cardiff CF14 4XW, Wales
关键词
Vegetation; Aspiration; CIED; Defibrillator; Endocarditis; Case report; INFECTIVE ENDOCARDITIS; CLINICAL PRESENTATION; PACEMAKER; MANAGEMENT; SIZE;
D O I
10.1093/ehjcr/ytae332
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background When cardiac implantable electronic device infection occurs, standard therapy is usually total system extraction. Transvenous lead extraction is preferable to open heart surgical extraction, unless contraindicated because of the presence of very large vegetations on the intravenous leads according to the European Society of Cardiology guidelines. Extraction of transvenous leads with vegetations risks distal embolism resulting in obstruction and/or infection in the pulmonary arteries. Catheter aspiration of vegetations or thrombi has been performed prior to transvenous lead extraction using a partial veno-venous extracorporeal bypass circuit. We report the use of a single-access aspiration system using the Inari FlowTriever 24 French system to debulk a defibrillator lead before percutaneous extraction.Case summary A 79-year-old male presented with fever 18 years after his first implantable cardioverter defibrillator implant and 9 years after his most recent pulse generator change. Two large vegetations were identified on his transvenous defibrillator lead on the atrial aspect, near the tricuspid annulus, which were aspirated using the Inari Medical 24Fr FlowTriever aspiration catheter. We describe anatomical considerations during the approach and a technique to localize the vegetations based on a combination of fluoroscopy and transoesophageal echocardiogram guidance.Discussion This case demonstrates the safe and effective use of the Inari Medical 24Fr FlowTriever aspiration catheter in debulking a defibrillator lead before transvenous lead extraction. This method uses a single venous puncture and is not dependent on extracorporeal bypass. Apart from reducing complexity, this technique may be advantageous in patients where anticoagulation needs to be minimised.
引用
收藏
页数:5
相关论文
共 14 条
[1]  
Delgado V, 2023, EUR HEART J, V44, P3948, DOI 10.1093/eurheartj/ehad193
[2]   Endocarditis in patients with a permanent pacemaker: A 1-year epidemiological survey on infective endocarditis due to valvular and/or pacemaker infection [J].
Duval, X ;
Selton-Suty, C ;
Alla, F ;
Salvador-Mazenq, M ;
Bernard, Y ;
Weber, M ;
Lacassin, F ;
Nazeyrolas, P ;
Chidiac, C ;
Hoen, B ;
Leport, C .
CLINICAL INFECTIOUS DISEASES, 2004, 39 (01) :68-74
[3]   Transcatheter vacuum aspiration of valvular and lead related infective endocarditis [J].
Gill, Gauravpal S. ;
Chakrala, Teja ;
Kanmanthareddy, Arun ;
Alla, Venkata Mahesh .
CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2023, 57 :8-15
[4]   Percutaneous Pacemaker and Implantable Cardioverter-Defibrillator Lead Extraction in 100 Patients With Intracardiac Vegetations Defined by Transesophageal Echocardiogram [J].
Grammes, Jon A. ;
Schulze, Christopher M. ;
Al-Bataineh, Mohammad ;
Yesenosky, George A. ;
Saari, Christine S. ;
Vrabel, Michelle J. ;
Horrow, Jay ;
Chowdhury, Mashiul ;
Fontaine, John M. ;
Kutalek, Steven P. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (09) :886-894
[5]   Influence of Vegetation Size on the Clinical Presentation and Outcome of Lead-Associated Endocarditis Results From the MEDIC Registry [J].
Greenspon, Arnold J. ;
Le, Katherine Y. ;
Prutkin, Jordan M. ;
Sohail, M. Rizwan ;
Vikram, Holenarasipur R. ;
Baddour, Larry M. ;
Danik, Stephan B. ;
Peacock, James ;
Falces, Carlos ;
Miro, Jose M. ;
Naber, Christoph ;
Carrillo, Roger G. ;
Tseng, Chi-Hong ;
Uslan, Daniel Z. .
JACC-CARDIOVASCULAR IMAGING, 2014, 7 (06) :541-549
[6]  
Issa Ziad F, 2016, HeartRhythm Case Rep, V2, P17, DOI 10.1016/j.hrcr.2015.08.008
[7]   Systemic infection related to endocarditis on pacemaker leads - Clinical presentation and management [J].
Klug, D ;
Lacroix, D ;
Savoye, C ;
Goullard, L ;
Grandmougin, D ;
Hennequin, JL ;
Kacet, S ;
Lekieffre, J .
CIRCULATION, 1997, 95 (08) :2098-2107
[8]  
Krishnan Balaji, 2021, HeartRhythm Case Rep, V7, P170, DOI 10.1016/j.hrcr.2020.12.005
[9]   2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction [J].
Kusumoto, Fred M. ;
Schoenfeld, Mark H. ;
Wilkoff, Bruce L. ;
Berul, Charles I. ;
Birgersdotter-Green, Ulrika M. ;
Carrillo, Roger ;
Cha, Yong-Mei ;
Clancy, Jude ;
Deharo, Jean-Claude ;
Ellenbogen, Kenneth A. ;
Exner, Derek ;
Hussein, Ayman A. ;
Kennergren, Charles ;
Krahn, Andrew ;
Lee, Richard ;
Love, Charles J. ;
Madden, Ruth A. ;
Mazzetti, Hector Alfredo ;
Moore, JoEllyn Carol ;
Parsonnet, Jeffrey ;
Patton, Kristen K. ;
Rozner, Marc A. ;
Selzman, Kimberly A. ;
Shoda, Morio ;
Srivathsan, Komandoor ;
Strathmore, Neil F. ;
Swerdlow, Charles D. ;
Tompkins, Christine ;
Wazni, Oussama .
HEART RHYTHM, 2017, 14 (12) :E503-+
[10]  
Misra SK., 2021, Heart Rhythm, V18, pS2