Accuracy of implant placement using a mixed reality-based dynamic navigation system versus static computer-assisted and freehand surgery: An in Vitro study

被引:5
|
作者
Shusterman, Ariel
Nashef, Rizan [1 ]
Tecco, Simona [2 ]
Mangano, Carlo [2 ]
Lerner, Henriette [3 ]
Mangano, Francesco Guido [4 ]
机构
[1] Shaare Zedek Med Ctr, Oral & Maxillofacial Surg Unit, Jerusalem, Israel
[2] San Raffaele Univ, Dept Dent Sci, Milan, Italy
[3] Goethe Univ Frankfurt, Acad Teaching & Res Inst, Frankfurt, Germany
[4] IM Sechenov First State Med Univ, Dept Pediat Prevent Dent & Orthodont, Moscow, Russia
关键词
Guided implant surgery; Static guided implant surgery; Freehand implant placement; Mixed reality; Holograms; accuracy;
D O I
10.1016/j.jdent.2024.105052
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This in vitro study aimed to compare the accuracy of dental implant placement in partially edentulous maxillary models using a mixed reality-based dynamic navigation (MR-DN) system to conventional static computer-assisted implant surgery (s-CAIS) and a freehand (FH) method. Methods: Forty-five partially edentulous models (with teeth missing in positions #15, #16 and #25) were assigned to three groups (15 per group). The same experienced operator performed the model surgeries using an MR-DN system (group 1), s-CAIS (group 2) and FH (group 3). In total, 135 dental implants were placed (45 per group). The primary outcomes were the linear coronal deviation (entry error; En), apical deviation (apex error; Ap), XY and Z deviations, and angular deviation (An) between the planned and actual (post-surgery) position of the implants in the models. These deviations were computed as the distances between the stereolithographic (STL) files for the planned implants and placed implants captured with an intraoral scanner. Results: Across the three implant sites, the MR-DN system was significantly more accurate than the FH method (in XY, Z, En, Ap and An) and s-CAIS (in Z, Ap and An), respectively. However, S-CAIS was more accurate than MRDN in XY, and no difference was found between MR-DN and s-CAIS in En. Conclusions: Within the limits of this study (in vitro design, only partially edentulous models), implant placement accuracy with MR-DN was superior to that of FH and similar to that of s-CAIS. Statement of Clinical Relevance: In vitro, MR-DN showed greater accuracy in implant positioning than FH, and similar accuracy to s-CAIS: it could, therefore, represent a new option for the surgeon. However, clinical studies are needed to determine the feasibility of MR-DN.
引用
收藏
页数:10
相关论文
共 42 条
  • [1] Implant placement using mixed reality-based dynamic navigation: A proof of concept
    Shusterman, Ariel
    Nashef, Rizan
    Tecco, Simona
    Mangano, Carlo
    Mangano, Francesco
    JOURNAL OF DENTISTRY, 2024, 149
  • [2] Comparison of the accuracy of dental implant placement using static and dynamic computer-assisted systems: an in vitro study
    Zhou, Miao
    Zhou, Hui
    Li, Shu-Yi
    Zhu, Yi-Bo
    Geng, Yuan-Ming
    JOURNAL OF STOMATOLOGY ORAL AND MAXILLOFACIAL SURGERY, 2021, 122 (04) : 343 - 348
  • [3] Accuracy of implant placement via dynamic navigation and autonomous robotic computer-assisted implant surgery methods: A retrospective study
    Zhang, Sihui
    Cai, Qin
    Chen, Weiyi
    Lin, Yuxuan
    Gao, Yan
    Wu, Dong
    Chen, Jiang
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (02) : 220 - 229
  • [4] Accuracy of Different Surgical Guide Designs for Static Computer-Assisted Implant Surgery: An In Vitro Study
    Wu, Yu Tsung
    Papaspyridakos, Panos
    Kang, Kiho
    Finkelman, Matthew
    Kudara, Yukio
    De Souza, Andre B.
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (05) : 351 - 357
  • [5] Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study
    Yuzhang Feng
    Zhenya Su
    Anchun Mo
    Xingmei Yang
    International Journal of Implant Dentistry, 8
  • [6] Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study
    Feng, Yuzhang
    Su, Zhenya
    Mo, Anchun
    Yang, Xingmei
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2022, 8 (01)
  • [7] The accuracy of implant placement using a combination of static and dynamic computer-assisted implant surgery in fully edentulous arches: A prospective controlled clinical study
    Lorwicheanrung, Jarungvit
    Mahardawi, Basel
    Arunjaroensuk, Sirida
    Kaboosaya, Boosana
    Mattheos, Nikos
    Pimkhaokham, Atiphan
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (08) : 841 - 853
  • [8] Accuracy of freehand surgery, static and dynamic computer assisted surgery on zygomatic implant placement: A systematic review and meta-analyses
    Traboulsi-Garet, Bassel
    Jorba-Garcia, Adria
    Bara-Casaus, Javier
    Camps-Font, Octavi
    Valmaseda-Castellon, Eduard
    Figueiredo, Rui
    Sanchez-Garces, Angels
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2025, 53 (04) : 301 - 311
  • [9] Comparison of the accuracy of implant placement using different drilling systems for static computer-assisted implant surgery: A simulation-based experimental study
    Sittikornpaiboon, Paknisa
    Arunjaroensuk, Sirida
    Kaboosaya, Boosana
    Subbalekha, Keskanya
    Mattheos, Nikos
    Pimkhaokham, Atiphan
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2021, 23 (04) : 635 - 643
  • [10] Accuracy of robot-assisted implant surgery versus freehand placement: a retrospective clinical study
    Younis, Hamza
    Xu, Boya
    Acharya, Kiran
    He, Longlong
    Du, Liangzhi
    Elayah, Sadam Ahmed
    Chang, Xiaofeng
    Lv, Chengpeng
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2025, 11 (01)