摘要
We present an open-source homemade Stokes polarimeter for real-time determination of the state and degree of polarization of a light beam by measuring all four Stokes parameters. Our design uses standard, commercially available optical components, mounted on a 3D printed base. Our user-friendly software allows visualization of the measured polarization state on the Poincar & eacute; sphere. We model the polarimeter using realistic Mueller matrices and present a detailed methodology to characterize every optical component and calibrate the instrument. The accuracy of the polarimeter largely depends on the quality of the chosen optical components, which in turn affects its price. Depending on this choice, the instrument can be used for applications ranging from simple didactic purposes to more accurate, research-grade measurements. In our case, we estimate an accuracy better than +/- 2.2(degrees) in the determination of the state of polarization and +/- 6.0% in the degree of polarization. This makes our instrument useful for the correct alignment of polarization-maintaining fibers, as it allows one to quickly and efficiently match the polarization axis of the light to the polarization axes of the fiber. The high sampling rate of 60 k samples/s opens up the possibility to analyze the polarization of rapidly fluctuating signals such as light pulses. Moreover, the flexibility of our design permits to analyze light signals in a broad range of wavelengths, from deep UV to infrared, after opportune change of the optical components, representing an alternative low cost to more expensive commercial instruments. (c) 2024 Optica Publishing Group. All rights, including for text and data mining (TDM), Artificial Intelligence (AI) training, and similar technologies, are reserved.
引用
收藏
页码:7177 / 7187
页数:11
相关论文
共 50 条
  • [1] Error analysis and optimization for a full-Stokes division-of-space polarimeter
    Wang, Luo
    Zhang, Haiyang
    Zhao, Changming
    Luo, Pingping
    APPLIED OPTICS, 2023, 62 (25) : 6816 - 6825
  • [2] Liquid crystal polymer full-stokes division of focal plane polarimeter
    Myhre, Graham
    Hsu, Wei-Liang
    Peinado, Alba
    LaCasse, Charles
    Brock, Neal
    Chipman, Russell A.
    Pau, Stanley
    OPTICS EXPRESS, 2012, 20 (25): : 27393 - 27409
  • [3] Design and calibration for a Full-Stokes imaging polarimeter
    Montes Gonzalez, Ivan
    Bruce, Neil C.
    UNCONVENTIONAL OPTICAL IMAGING, 2018, 10677
  • [4] Snapshot compressive imaging full-Stokes polarimeter
    Xu, Zhilong
    Meng, Jianwen
    Zhao, Man
    Yang, Ting
    Wu, Dan
    Zhang, Rui
    Xie, Yingge
    Ren, Wenyi
    OPTICS COMMUNICATIONS, 2022, 509
  • [5] Optimum angles for a full-Stokes scanning polarimeter
    Hu, Qiang
    Qiu, Zhenwei
    Cui, Weihua
    Hong, Jin
    APPLIED OPTICS, 2019, 58 (16) : 4277 - 4282
  • [6] Chip-scale, full-Stokes polarimeter
    Lin, Zhongjin
    Rusch, Leslie
    Chen, Yuxuan
    Shi, Wei
    OPTICS EXPRESS, 2019, 27 (04) : 4867 - 4877
  • [7] Method to calibrate a full-Stokes polarimeter based on variable retarders
    Montes-Gonzalez, Ivan
    Bruce, Neil C.
    Rodriguez-Herrera, Oscar G.
    Rodriguez Nunez, Omar
    APPLIED OPTICS, 2019, 58 (22) : 5952 - 5957
  • [8] Full-Stokes Polychromatic Polarimeter Design for Arago
    Pertenais, Martin
    Neiner, Coralie
    Petit, Pascal
    SPACE TELESCOPES AND INSTRUMENTATION 2016: ULTRAVIOLET TO GAMMA RAY, 2016, 9905
  • [9] Simplified Stokes polarimeter based on division-of-amplitude
    Negara, Christian
    Li, Zheng
    Laengle, Thomas
    Beyerer, Juergen
    PHOTONICS AND EDUCATION IN MEASUREMENT SCIENCE, 2019, 11144
  • [10] Division of focal plane red-green-blue full-Stokes imaging polarimeter
    Tu, Xingzhou
    McEldowney, Scott
    Zou, Yang
    Smith, Matthew
    Guido, Christopher
    Brock, Neal
    Miller, Sawyer
    Jiang, Linan
    Pau, Stanley
    APPLIED OPTICS, 2020, 59 (22) : G33 - G40