Task-based assessment for radiation dose optimization in CT abdominal examinations: A phantom study

被引:0
作者
Halim, Esnu [1 ]
Kechik, Mohd Mustafa Awang [1 ]
Ibahim, Mohamad Johari [2 ]
Harun, Hanif Haspi [1 ]
Shaffiq, Said Mohd [3 ]
Yusof, Aimi Adibah [1 ]
Karim, Muhammad Khalis Abdul [1 ]
机构
[1] Univ Putra Malaysia, Fac Sci, Dept Phys, Serdang 43400, Selangor, Malaysia
[2] Univ Teknol MARA, Fac Med, Sungai Buloh, Selangor, Malaysia
[3] Natl Canc Inst, Putrajaya 62250, Malaysia
关键词
Abdominal CT; Optimization; Radiation dose; Signal-to-noise ratio; Figure-of-merit; IMAGE QUALITY; COMBINATION;
D O I
10.1016/j.radphyschem.2024.111966
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Recent advancements in Computed Tomography (CT) imaging technology have significantly enhanced healthcare delivery. However, public concern regarding the safety and health risks associated with increased exposure to ionizing radiation has escalated. In response, this study performs a quantitative analysis comparing the image quality of the standard CT abdomen protocol (P1 protocols) with alternative optimization techniques. Adjustments to standard CT scan parameters, including tube potential (P2 protocols), effective mAs value (P3), and pitch factor (P4, P5, P6 protocols), were made to evaluate the effectiveness of these optimization strategies. A 64slice CT scanner (Canon, Japan) and a polymethylmethacrylate (PMMA) textural water phantom was utilized in this study. CT dose descriptors, volume weighted CT Dose Index (CTDI vol ) and dose-length product (DLP) alongside evaluation parameters such as Hounsfield Unit (HU) number, noise levels, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured and analysed. Furthermore, this research introduced a Figure of Merit (FOM) based on the ratio of image quality parameters (SNR and CNR) to the CTDI vol , establishing a quantitative measure for optimization. A notable reduction in radiation dose -34.85% below the standard protocol (P1) -was observed in protocol P6, attributed to an increased pitch factor. Conversely, increase the effective mAs value has improved SNR and CNR values. The findings reveal that protocols P1 and P2 deliver the highest FOM for soft tissue and fat imaging, respectively. Protocol P1, with its higher attenuation factor, is optimal for abdominal examinations targeting soft tissues. Meanwhile, protocol P2 's increase effective mAs value enhances beam intensity, improving CT numbers for fat tissues with their high-density composition. The study also found a strong correlation between FOM based on SNR and CNR values. In conclusion, adjustments to primary CT parameters significantly influence image quality. The FOM serves as an effective tool for measuring the overall performance of image quality, facilitating post-optimization assessment and contributing to the development of safer, more efficient CT imaging protocols.
引用
收藏
页数:7
相关论文
共 26 条
  • [21] Paquet F., 2017, Annals of the ICRP, V46, P1, DOI [10.1177/0146645317734963, 10.1016/j.icrp.2013.01.001, 10.1016/j.icrp.2008.08.001]
  • [22] Ramadhan M.M., 2020, Fast and Accurate Detection of Covid-19-Related Pneumonia from Chest X-Ray Images with Novel Deep Learning Model.
  • [23] Radiation exposure of the interventional radiologist during percutaneous biopsy using a multiaxis interventional C-arm CT system with 3D laser guidance: a phantom study
    Rathmann, Nils
    Kostrzewa, Michael
    Kara, Kerim
    Bartling, Soenke
    Haubenreisser, Holger
    Schoenberg, Stefan O.
    Diehl, Steffen J.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1055)
  • [24] Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms
    Richard, Samuel
    Husarik, Daniela B.
    Yadava, Girijesh
    Murphy, Simon N.
    Samei, Ehsan
    [J]. MEDICAL PHYSICS, 2012, 39 (07) : 4115 - 4122
  • [25] Virtual monochromatic imaging in dual-source dual-energy CT: Radiation dose and image quality
    Yu, Lifeng
    Christner, Jodie A.
    Leng, Shuai
    Wang, Jia
    Fletcher, Joel G.
    McCollough, Cynthia H.
    [J]. MEDICAL PHYSICS, 2011, 38 (12) : 6371 - 6379
  • [26] A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography
    Zhao, Yong-Xia
    Zuo, Zi-Wei
    Suo, Hong-Na
    Wang, Jia-Ning
    Chang, Jin
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2016, 40 (05) : 784 - 790